Fluid Bag:
Macro IV line (macrodrip set)
An IV fluid line that delivers 15 drops/mL
Allows for a HIGH VOLUME to be delivered in a SHORT TIME
Used for patients weighing MORE than 7 kg
Micro IV line (microdrop set)
An IV fluid line that delivers 60 drops/mL
Allows for GREATER precision
Used for patients weighing LESS than 7 kg


IV Burette set
Attached to a 1L fluid bag
Contains 100 mL
Allows for the MOST accurate measurement for fluids administered
Used on SMALL patients

IV lines have to be PRIME - running fluid to the line - to get RID of AIR
IV fluid clamps = CONTROL the flow of fluid through the IV line
OPEN clamp —→ FLUID FLOWS
CLOSE clamp —→ FLUID STOPS
OVERFILLING IV chamber - may NOT see drops & may NOT know if fluid is RUNNING
Fluid Pump:
ALWAYS in mL/hr
RATE = Fluids given PER HOUR
FIRST HOUR = 5 mL/kg/hr
AFTER FIRST HOUR DECREASE TO = 3 mL/kg/hr
Ex: 10 kg rate is 50 mL/hr for the FIRST hour (10 kg × 5 mL/kg/hr = 50 mL/hr)
VTBI = Volume to be Infused
AMOUNT of fluid the pump is SET to GIVE
VI = Volume Infused
How much the animal have RECEIVED
Should be CLEARED when inputting the rate & VTBI in the pump

Bolus
LARGE amount of fluid given in a SHORT amount of time
Given for DEHYDRATED animal & LOW BP
Practice Problems for Bolus Fluid Rate :
Want to give 100 mL bolus in 15 minutes. What will the RATE be?
60/15 = 4 —→ 100 × 4 = 400 mL/hr
VTBI = 100 mL
BOLUS RATE = 400 mL/hr
Want to give 80 mL bolus in 15 minutes. What will the RATE be?
60/15 = 4 —→ 80 × 4 = 320 mL/hr
VTBI = 80 mL
BOLUS RATE = 320 mL/ hr
Want to give 60 mL bolus in 20 minutes. What will the RATE be?
60/20 = 3 —→ 60 × 3 = 180 mL/hr
VTBI = 60 mL
BOLUS RATE = 180 mL/hr
Want to give 80 mL bolus in 10 minutes. What will the RATE be?
60/10 = 6 —→ 80 × 6 = 480 mL/hr
VTBI = 80 mL
BOLUS RATE = 480 mL/hr

Practice Problem for Bolus Fluid Rate & Maintenance Fluid Rate:
Dog Surgical Rate = 5 mL/kg/hr
Cat Surgical Rate = 3 mL/kg/hr
10 kg dog
Bolus rate: 120 mL in 15 minutes
60/15 = 4 —→ 120 × 4 = 480 mL/hr
BOLUS RATE = 480 mL/hr
VTBI = 120 mL
TIME = 15 minutes
Maintenance Rate = 5 mL/kg/hr
10 kg x 5 mL/kg/hr = 50 mL/hr
MAINTENANCE RATE = 50 mL/hr
VTBI = 50 mL
Time = 1 hour
5 kg cat
Bolus rate: 60 mL in 20 minutes
60/20 = 3 —→ 60 × 3 = 180 mL/hr
BOLUS RATE = 180 mL/hr
VTBI = 60 mL
TIME = 20 minutes
Maintenance rate: 3 mL/kg/hr
5 kg x 3 mL/kg/hr = 15 mL/hr
MAINTENANCE RATE = 15 mL/hr
VTBI = 15 mL
TIME = 1 hour
30 kg dog
Bolus rate: 100 mL in 10 minutes
60/10 = 6 —→ 100 × 6 = 600 mL/hr
BOLUS RATE = 600 mL/hr
VTBI = 100 mL
TIME = 10 minutes
Maintenance rate: 5 mL/kg/hr
30 kg x 5 mL/kg/hr = 150 mL/hr
MAINTENANCE RATE = 150 mL/hr
VTBI = 150 mL
TIME = 1 hour

Practice Problem for Bolus Fluid Rate, Maintenance Fluid Rate & VTBI:
5 kg dog
a) Surgical rate = 5 mL/kg/hr
5 kg x 5 mL/kg/hr = 25 mL/hr
b) 1 hr bolus in 10 minutes
25 mL/hr x 1 hr = 25 mL
60/10 = 6 —→ 25 × 6 = 150 mL/hr
Maintenance rate = 25 mL/hr
Maintenance time = 1 hour
VTBI = 25 mL
Bolus rate = 150 mL/hr
Bolus time = 10 minutes
20 kg dog
a) Surgical rate = 5 mL/kg/hr
20 kg x 5 mL/kg/hr = 100 mL/hr
b) 1 hr bolus in 15 minutes
100 mL/hr x 1 hr = 100 mL
60/15 = 4 —→ 100 × 4 = 400 mL/hr
Maintenance rate = 100 mL/hr
Maintenance time = 1 hour
VTBI = 100 mL
Bolus rate = 400 mL/hr
Bolus time = 15 minutes
3 kg cat
a) Surgical rate = 3 mL/kg/hr
3 kg x 3 mL/kg/hr = 9 mL/hr
b) 1 hr bolus in 20 minutes
9 mL/hr x 1 hr = 9 mL
60/20 = 3 —→ 9 × 3 = 27 mL/hr
Maintenance rate = 9 mL/hr
Maintenance time = 1 hour
VTBI = 9 mL
Bolus rate = 27 mL/hr
Bolus time = 20 minutes

IV Catheter Supplies & Set-Up:
IV catheter

Flush (sterile SALINE)
T-Port
FLUSH w/ sterile saline to get rid of AIR (CLAMP SHOULD be OPEN) & leave syringe attached
Connects to IV fluids


Male adapter
Open package & LOOSEN green plug
END of the male adapter is the INJECTION PART

4 pieces of tape
2 fat ones & 2 skinny ones w/ tabs

Vet wrap
Should be RED to indicate patient has an IV catheter
Needs a V-cut

Clippers
USE No.40 blade - required for hair removal from the catheter site
Scrub & Rinse
Scrub = Chlorhexidine
Rinse = 70% isopropyl alcohol
3 times in circular motion
Parts of an IV Catheter:


Filter NEEDS to be STERILE
Catheter (plastic tube) is INSERTED inside the patient
IV Catheter - Step-By-Step Placement:
Step 1:
AFTER you have shaved & scrubbed your insertion site (PALPATE BEFORE you SCRUB to find VEIN), LOOSEN the CATHETER & FILTER
Restrainer HOLDS OFF VEIN

Step 2:
HOLD the STYLET (needle) w/ the THUMB, MIDDLE, & RING finger - this will drive the stylet FORWARD
USE INDEX FINGER to hold the CATHETER on the stylet

Step 3:
Once you see a FLASH of BLOOD in the HUB (blue part) of the stylet - ADVANCE the stylet a tiny bit to ensure the CATHETER is also in the VESSEL
USE INDEX FINGER to advance the CATHETER OFF of the STYLET & INTO the VESSEL
DO NOT ADVANCE or REMOVE STYLET

Step 4:
Once the catheter is ADVANCED COMPLETELY - REMOVE STYLET & PLUG FILTER at the end of the catheter hub
USE THUMB to HOLD the FILTER & CATHETER while applying the FIRST tape

Step 5:
USE SKINNY TAPE w/ STICKY SIDE UP & place it UNDER the catheter
Bring the LOOSE end over the TOP of the catheter & WRAP it AROUND the LEG
ONLY SECURE CATHETER - NOT THE FILTER

Step 6:
USE FAT TAPE w/ NOTCH w/ STICKY SIDE DOWN & place it UNDER the catheter
Bring the LOOSE end UNDER the leg - then OVER the TOP of the catheter to WRAP it AROUND the LEG
ONLY SECURE CATHETER - NOT THE FILTER

Step 7:
REPLACE the FILTER w/ T-PORT & firmly secure to catheter
Restrainer HOLDS OFF VEIN

Step 8:
FLUSH CATHETER to ensure PATENCY & if INSIDE VEIN - CLAMP is OPEN
ASPIRATE FIRST then FLUSH

Step 9:
USE SKINNY TAPE & TWIST the CENTER w/ STICKY SIDE UP - place it UNDER the T-PORT INJECTION HUB
CRISS-CROSS the tape OVER the T-PORT HUB

Step 10:
Place NOTCHED RED VETRAP UNDER the T-PORT
WARP AROUND the ENTIRE TAPED AREA

Step 11:
Place LAST FAT TAPE w/ STICKY SIDE UP UNDER the END of the T-PORT LINE
FOLD the END of the TAPE - the area where the hard plastic female end & flexible line meet
Then wrap the tape AROUND the LEG - SECURING the T-PORT
The CLAMP should be in the MIDDLE of the LINE at the TOP
FLUSH the catheter one LAST TIME to check for PATENCY then CLOSE the CLAMP

Step 12:
SWAP FLUSH syringe to MALE ADAPTER
Attach GREEN PLUG from the MALE ADAPTER to the FLUSH syringe

FLUID = CLAMP OPEN
INJECTION = CLAMP OFF