Who invented the fluoroscope, and when?
Thomas Edison, in 1896.
What is the primary purpose of fluoroscopy?
Dynamic, real-time imaging for functional studies and active diagnosis.
What did early fluoroscopes consist of?
An X-ray tube and hand-held fluorescent viewing screen.
Which body areas received high exposures with early hand-held fluoroscopes?
Eyes, neck, and hands.
When were image intensification tubes developed?
1948.
What replaced mirror optics for image-intensifier viewing?
Closed-circuit television systems.
By approximately what year did digital fluoroscopy predominate?
2005.
What matrix technology is associated with digital fluoroscopy?
Thin-film transistor (TFT) matrices.
Name four clinical uses of fluoroscopy.
GI studies, angiograms, line placement, and orthopedic surgery.
Why should fluoroscopy not be used to preview routine radiographic positioning?
It unnecessarily increases patient radiation exposure and is considered unethical by the ASRT.
What operating mA range is listed for fluoroscopic X-ray tubes?
0.5-5.0 mA.
What mA range is given for standard radiographic X-ray tubes?
50-1200 mA.
What is the minimum SOD for fixed fluoroscopic equipment?
15 inches (38 cm).
How is fluoroscopic exposure commonly activated?
Foot switch or handle switch.
How is the fluoroscopic field collimated?
Electronically.
What are the main components of an image-intensifier tube?
Input screen, photocathode, electrostatic lenses, magnification tubes, anode, and output screen.
What material forms the image-intensifier input phosphor on surface
Sodium-activated cesium iodide (CsI).
What is the input screen (phosphor) made of.
Glass, titanium, steel, or aluminum
What is the thickness of the input phosphor?
0.1-0.2 mm.
What is the input screen's primary conversion?
X-ray photons into light photons.
Why is the input screen curved?
So electrons from every point travel the same distance to the output screen
What is the traitor for the input screen being curved?
Causes pin cushion distortion (See image quality)
What is the photocathode's function?
It emits photoelectrons when illuminated by light from the input phosphor.
What metals are used in the photocathode?
Photoemissive metals: Cesium and antimony.
What is photoemission?
Release of electrons from the photocathode in response to light.
What do electrostatic lenses do?
Accelerate and focus electrons toward the anode/output screen.
What happens to image orientation in the image intensifier from the electrostatic lenses
It is inverted and reversed.
Electrostatic lenses
Focusing the electrons onto the small output screen is what makes ______ Possible
Minification gain (primary source of brightness gain)
What do the magnification tubes do?
Ad voltage
What does increasing voltage to electrostatic lenses do?
Increases electron acceleration and shifts the focal point away from the anode.
Magnification tubes are described according to what
According to the diameter of the input screen used
What type of tubes are used according to the diameter of the input screen used?
Dual focus
23/15 cm (9”/6”)
Multi field intensifier are popular
Triple field (12”/9”/6”)
Quad Field (14”/12”/9”/6”)
What happens to spatial resolution in image-intensifier magnification mode?
It increases (good!)
What happens to patient dose in image-intensifier magnification mode?
It increases. (Bad!)
What is the image-intensifier magnification formula?
Magnification = input screen diameter / diameter of input screen used during magnification
practice
what is the magnification for an image viewed with an I.I tube with an input screen diameter of 9 inches that is using a 5 inch diameter area during magnification
9/5 = 1.8 times.
What charge does the image-intensifier anode have?
Positive.
What approximate anode voltage is listed?
25 kVp.
Why does the anode have a central opening? (Hole in center)
To allow electrons to pass to the output phosphor.
The output screen is a ______ Screen
Glass fluorescent
What phosphor material is listed for the output screen?
Zinc-cadmium sulfide.
What does the output screen emit when struck by electrons?
Green light.
some I.I tubes coat the output phosphor onto a ______ Instead of glass, which carries light directly to the camera
Fiber optic disc
What can replace the conventional output screen in some newer image intensifiers?
A fiber-optic disc.
Trace the image-intensifier conversion chain.
X-rays -> input phosphor/light -> photocathode/electrons -> focused electrons -> output phosphor/light.
What is total brightness gain?
A measurement of the increase of image intensity achieved by a I.I tube
Brightness gain of I.I tubes tend to be -
5000 to 30,000
What is the brightness gain formula?
Brightness gain = Minification gain x flux gain.
What typical brightness gain range is listed?
Approximately 5,000-30,000.
What causes minification gain?
Focusing electrons from a large input phosphor area onto a smaller output phosphor area.
input (6”/9”/12”) —→ output (1”)
What is the minification gain formula?
(Input screen diameter / output screen diameter)^2.
What is minification gain for a 6-inch input and 1-inch output?
(6/1)^2 = 36.
What is flux gain?
A measurement of conversion efficiency of output screen
Electrons in; Light out
If 50 light photons emerge per electron, what is flux gain?
50.
What is total brightness gain if minification gain is 36 and flux gain is 60?
36 x 60 = 2,160.
Maintaining fluoroscopic, image, brightness, and contrast does what
It maintains brightness of image by automatically changing exposure factors according to subject density/thickness
Maintaining fluoroscopic, image, brightness, and contrast monitors what
It monitors current flowing between the cathode and the anode of I.I tube Or intensity of output screen
The primary beam is changed when what happens
When current falls below established level (vary mA, kV & pulse time)
Maintaining fluoroscopic, image, brightness, and contrast
This control has a wide variety of systems and names:
automatic brightness control (ABC)
Automatic brightness stabilization (ABS)
AUTOMATIC DOSE CONTROL (ADC)
Automatic gain control (AGS)
The control and maintaining fluoroscopic image brightness and contrast is very important in maintaining what?
Image quality
Image quality
contrast: is controlled by:
Amplitude of video signal
Image quality
contrast is affected by:
scattered ionizing radiation
Penumbral light scatter
Image quality
resolution; varies depending on
SID, OID, minification gain
Image quality
distortion is caused by
The curved shape of the image intensifier input screen
Image quality
distortion: what is the pincushion effect
Image is stretched and warped toward the edges
Image quality
distortion: what is vignetting
Image brightness falls off toward the edges
Image quality
quantum mottle is caused by what
It is blotchy grainy appearance caused by two little exposure to input Phosphor
Image quality
Quantum mottle: what are the options to improve it?
Increase MA, Reduced distance between patient and Imogen intensifier, Increase KVP
What is the purpose of video viewing systems?
To convert the light from the output of the image intensifier tube to digital signal
What are the most common video viewing systems?
semiconductor (CCD or CMOS) Attached to the output, Phosphor
Display monitor for viewing
What are the components of CCD? (Charged coupled device)
CCD array- convert light to electrical signal incense to computer
ADC- Converts e- to digital signal
What are the components of CMOS (Complementary metal oxide semiconductor)
CMOS sensor - Converts light to e- And send to computer
ADC- Convert e- To digital Value
What is flat panel fluoroscopy.
Replace the image intensifier with a digital flat panel TFT detector