Digital Image Postprocessing (Chapter 31)

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/36

flashcard set

Earn XP

Description and Tags

Comprehensive vocabulary flashcards covering digital image postprocessing concepts, including spatial, intensity, and frequency domains, gradation processing, dynamic range compression, filtering methods, and exposure indicator systems.

37 Terms

1
New cards

Spatial Domain

The image domain in which pixels are sorted and arranged strictly according to their physical locations in the image matrix.

2
New cards

Intensity Domain

The image domain in which pixels are sorted according to their pixel values or gray levels, resulting in a histogram.

3
New cards

Frequency Domain

The image domain in which objects and details within an image are sorted according to their physical size.

4
New cards

Point Processing Operations

Spatial domain operations that perform a specific mathematical algorithm on each individual pixel in sequence, pixel by pixel (e.g., image subtraction).

5
New cards

Area Processing Operations

Spatial domain operations that execute a mathematical function on a local group of pixels or a specific subsection of the image (e.g., magnification or zoom).

6
New cards

Global Operations

Spatial domain operations that carry out a massive function across the entire area of the image (e.g., rotate, invert, or translate).

7
New cards

Kernel

A smaller submatrix of core values passed over the original pixels of an image row by row and column by column, calculating and assigning new pixel values through interpolation.

8
New cards

Rescaling (Normalization)

The initial processing stage that remaps input pixel values to consistent output values, standardizing image brightness and roughly matching the SS range to the QQ range to standardize contrast.

9
New cards

Gradation Processing

Postprocessing performed in the intensity domain that applies an intensity transformation curve to customize image brightness (indicated by the left-to-right position of the curve) and contrast (represented by the steepness of the curve).

10
New cards

Anatomical LUT

A manufacturer-provided look-up table determined by the operator's exam selection at the console that customizes the gray scale and brightness curve specifically for the anatomy being demonstrated.

11
New cards

Gradient Curve Slope

The angle of the body of a gradient function curve; a slope steeper than 45o45^\text{o} increases contrast, while a slope shallower than 45o45^\text{o} decreases contrast.

12
New cards

Fuji GT Curve M

A specific Fuji gradient type curve that reverses image contrast from white-on-black to black-on-white, producing the 'black bone' feature.

13
New cards

GC (Gradation Center)

A Fuji gradation parameter representing the center point (average brightness) around which the gradient curve rotates.

14
New cards

GA (Gradation Amount)

A Fuji gradation parameter representing the rotation amount or slope change of the function curve, which controls image contrast.

15
New cards

GS (Gradation Shift)

A Fuji gradation parameter that shifts the function curve left or right (speed or sensitivity), adjusting overall brightness such that shifting to the right darkens the image.

16
New cards

Data Clipping

The truncation or cutting off of extreme high or low pixel values when adjustments to brightness or contrast exceed the bit depth of the hardware or dynamic range of the software.

17
New cards

Bit Depth

The absolute range of different gray (brightness) levels that a computer, LCD monitor, or other hardware device is capable of reproducing.

18
New cards

Dynamic Range

The range of different gray (brightness) levels made available by a computer imaging system, including its installed software.

19
New cards

Gray Scale

The actual range of different gray (brightness) levels present in a displayed digital image.

20
New cards

Dynamic Range Compression (DRC)

Also called tissue equalization or contrast equalization; a process that finds the midpoint of the gray scale curve, darkening pixels above it and lightening pixels below it to eliminate the darkest and lightest densities (gray scale truncation).

21
New cards

Detail Processing

A postprocessing operation that treats fine details as a separate component of the image, decoupling local contrast from global contrast without changing the overall brightness or contrast of the whole image.

22
New cards
<p>Kernel Mathematical Operations</p>

Kernel Mathematical Operations

Calculations applied across kernel cells: (A) smoothing occurs when all core values are equal and greater than zero; (B) edge enhancement occurs when positive and negative values sum to 11; (C) background suppression occurs when positive and negative values sum to 00.

23
New cards

"Unsharp" Mask Filtering

A spatial domain edge enhancement method where a low-pass blurred mask containing only gross structures is reversed into a positive and superimposed over the original image to subtract blurry broad areas and preserve sharp fine details.

24
New cards

Low-Pass Filtering

A frequency filtering operation (smoothing) that removes high-frequency components (small details) to soften edges and suppress image noise or quantum mottle.

25
New cards

High-Pass Filtering

A frequency filtering operation (edge enhancement) that removes low-frequency components (large, background structures) to make small details more visible, at the cost of increasing visible noise.

26
New cards

Fourier Transformation

A mathematical algorithm that breaks down a complex spatial waveform representing a row of pixel values into its individual constituent sine waves and frequencies.

27
New cards

Inverse Fourier Transformation

A mathematical algorithm that recomposes an image by summing all component sine waves back into a single complex waveform that controls monitor display voltages.

28
New cards

Multiscale Processing

A process that decomposes a digital image into eight or more separate frequency bands, allowing each band to be filtered and amplified independently before image reconstruction (e.g., Philips' UNIQUE, Agfa's MUSICA, Fuji's MFP).

29
New cards

Pyramidal Decomposition

The repeated splitting of an original image into high-frequency and low-frequency components, generating a 3-D stack of image layers in which lower layers have progressively smaller matrix sizes.

30
New cards

Periodic Noise

Artifacts of roughly consistent size that occur in a regular, repeating pattern (such as electronic 'snow'), which are best eliminated using frequency domain processing.

31
New cards

Random Noise

Artifacts of variable size that occur in an irregular, chaotic distribution across multiple frequencies (such as quantum mottle), which are best suppressed using spatial kernel operations.

32
New cards

Exposure Indicator (EI)

A numerical value calculated from the histogram midpoint between SminS_{\text{min}} and SmaxS_{\text{max}} that provides feedback to the radiographer regarding detector exposure, independent of displayed image brightness.

33
New cards

Logarithmic Exposure Scale

An exposure indicator scale where a change of 0.30.3 bels (or 300300 mbels) represents a change in actual detector exposure by a factor of 22 (e.g., Agfa LgM, Carestream EI).

34
New cards

Inversely Proportional Exposure Scale

An exposure indicator scale (such as Fuji's S Value) where the numerical indicator moves opposite to detector exposure; an indicator value halved represents double the exposure, and an indicator value doubled represents half the exposure.

35
New cards

Target Exposure Index (EITEI_T)

The ideal detector exposure level for a specific anatomical projection, determined by the image acquisition system, the kVp range, and the digital processing speed class.

36
New cards

Deviation Index (DI)

A standardized metric calculated as DI=10×log10(EI/EIT)DI = 10\times\text{log}_{10}(EI / EI_T), which changes by +1.0+1.0 for every +25%+25\% exposure increase and −1.0-1.0 for every −20%-20\% exposure decrease.

37
New cards

Saturation

Severe detector overexposure—typically requiring 8 to 108\text{ to }10 times normal technique (EI=3000EI = 3000 for Carestream, S=25S = 25 for Fuji)—that overwhelms detection elements and causes irreversible loss of data, presenting as flat black areas.