1/25
Comprehensive vocabulary flashcards covering the principles, equipment, indications, procedural steps, probe manipulations, standard views, and Doppler/3D imaging assessments in Transesophageal Echocardiography (TEE).

Transesophageal Echocardiography (TEE)
An echocardiographic imaging modality where an ultrasound probe is positioned within the esophagus to provide high-resolution images of cardiac structures, particularly posterior anatomy, without interference from intervening lung or bone tissue.
TEE Transducer Frequencies
Higher-frequency transducers, specifically 5 MHz and 7 MHz, made possible by decreased image depth and improved signal-to-noise ratio compared to transthoracic imaging.
Repositioning (TEE Probe)
Movement of the TEE probe up and down along the esophagus.
Rotation (TEE Probe)
Rotating the ultrasound imaging plane from 0⊥ to 180⊥ electronically using the multiplane control knob.
Turning (TEE Probe)
Moving the entire transducer shaft in a rotational fashion within the esophagus to achieve a mediolateral change in the imaging plane.
Angulation (TEE Probe)
Bending and extending the probe tip so that the image plane is directed superiorly or inferiorly at an angle to the original image plane.
Tilt (TEE Probe)
Lateral motion of the transducer tip to image different structures within the same plane.

Four Standard TEE Transducer Depth Levels
The standard anatomical positions within the gastrointestinal tract: Upper Esophageal (UE), Mid Esophageal (ME), Transgastric (TG), and Deep Transgastric (DTG).
NPO Guidelines for TEE
Patient preparation requirement to fast for at least 4 hours before the procedure and avoid eating or drinking for 1 hour after the procedure.
Vegetation Size Threshold for TEE vs. TTE
TEE is indicated for evaluating suspected endocarditis when vegetations are ≤2 mm, whereas vegetations >2 mm can usually be visualized via transthoracic echocardiography (TTE).

Mid-Esophageal Four-Chamber View (ME 4-Chamber View)
A standard TEE view obtained at a 0⊥–10⊥ angle displaying the left atrium, right atrium, left ventricle, right ventricle, interatrial septum, interventricular septum, mitral valve leaflets, and tricuspid valve.

Fossa Ovalis
The thin central region of the interatrial septum visualized between the arrows in a mid-esophageal view, often evaluated in biplane imaging at 90⊥.

Mid-Esophageal Two-Chamber View (ME 2-Chamber View)
A view obtained at an 80⊥–100⊥ transducer angle displaying the left atrium, left ventricle, left atrial appendage (LAA), coronary sinus, and anterior and posterior mitral valve leaflets.

Mid-Esophageal Long-Axis View (ME Long-Axis View)
A view acquired at approximately 120⊥–140⊥ multiplane rotation displaying the left atrium, left ventricle, LVOT, aortic valve, proximal ascending aorta, and mitral valve leaflets (P2 and A2).

Mid-Esophageal RV Outflow Tract View
A mid-esophageal view obtained at approximately 90⊥ (or 50⊥–70⊥) showing the right ventricular outflow tract (RVOT), pulmonic valve (PV), and main pulmonary artery (PA).

Mid-Esophageal Bicaval View
A view acquired at 90⊥–110⊥ with the probe turned toward the patient's right side, imaging the right atrium, superior vena cava (SVC), inferior vena cava (IVC), interatrial septum, and often a Eustachian valve.

Mid-Esophageal Aortic Valve Short-Axis View (ME AV SAX View)
A view obtained at 25⊥–45⊥ displaying a cross-section of the aortic valve cusps, right atrium, left atrium, superior interatrial septum, RVOT, and pulmonary valve.

Transgastric Short-Axis Mid-LV View
A cross-sectional view obtained by retroflexing the probe in the stomach at 0⊥, widely used during intraoperative monitoring to evaluate left ventricular size and regional wall motion at the papillary muscle level.

Transgastric Short-Axis Mitral Valve View
A view obtained from the transgastric position by withdrawing the probe slightly toward the gastroesophageal junction, defining the anterior (AMVL) and posterior (PMVL) mitral valve leaflets in short axis.

Descending Thoracic Aorta View
An imaging view obtained by turning the transducer posteriorly toward the left of the spine, displaying the aorta in short axis at 0⊥ and in long axis at 90⊥.

Left Ventricular Inflow Doppler on TEE
Pulsed Doppler interrogation at the mitral leaflet tips in an anteriorly angulated four-chamber view, producing inverted E and A waves because diastolic flow travels away from the esophageal probe.

Pulmonary Vein Doppler Flow Profile
Pulsed Doppler tracing of pulmonary venous inflow demonstrating systolic (S) and diastolic (D) forward flow into the left atrium, followed by a small atrial reversal (a) wave.

Biplane Imaging
A 3D TEE capability allowing simultaneous visualization of two orthogonal planes (e.g., four-chamber and two-chamber views at 90⊥ to each other) from a single probe position.

3D Left Ventricular Ejection Fraction (LVEF) Measurement
Semiautomated border detection on 3D full-volume datasets used to quantify end-diastolic volume (EDV), end-systolic volume (ESV), and EF along with a frame-by-frame time--volume curve.
Mid-Esophageal Mitral Commissural View
A mid-esophageal plane obtained at a 50⊥–70⊥ angle imaging the left atrium, left ventricle, coronary sinus, papillary muscles, chordae tendineae, and mitral scallops (P3,A3,A2,A1,P1).
Eustachian Valve
A prominent remnant valve or flap often identified at the junction of the inferior vena cava (IVC) and the right atrium (RA) in the bicaval view.