proteome, metabolome, and lipidome can all be measured by

mass spectrometry

2
New cards

basic principle of mass spectrometry

" generate ions, separate by mass-to-charge ratio (m/z), detect qualitatively and quantitatively by m/z and abundance"

3
New cards

"Parts of Mass Spectrometer and their functions"

ion source - ionization
mass analyzer - ion separation
detector - ion detection

4
New cards

what are the two main forms of ionization in MS

MALDI (Matrix Assisted Laser Desorption Ionization) and ESI (Electro-Spray Ionization)

5
New cards

What is the major difference between MALDI and ESI?

MALDI converts solid particles into gas, ESI converts liquid particles into gas

6
New cards

how do MS spectra resulting from MALDI and ESI differ and why?

ESI has many MS peaks over a large m/z scale (multiple charges) due to many proton transfers caused by explosions and collisions in final step

7
New cards

MALDI process

"

  1. Analyte (peptides) imbedded into organic acid matrix 
  2. Illumination by laser impulse which is absorbed by matrix 
  3. Matrix heating and expansion (explosion into gas phase), carrying analytes 
"

8
New cards

ESI process

"

  1. Steel capillary as anode (+), MS entry as cathode (-) 
  2. Liquid with ions drawn into tailored cone 
  3. Droplets containing solvent and positive ions discharged from steel capillary 
  4. Solvent molecules evaporate, shrinking droplets and causing increased electrostatic repulsion in comparison to surface tension, explosion into droplets into gas 
"

9
New cards

What causes the different heights of peaks when reading peptide mass fingerprint?

"some peptides are more efficiently ionized than others"

10
New cards

What do the charges on an MS peptide peak represent?

"isotopes, seen as smaller equidistant peaks with a distance of (1/z)"

11
New cards

"Monoisotopic mass"

"sum of all exact isotopic masses of the most abundant isotope of the elements that make up a molecule "

12
New cards

"Average Mass formula"

"(sum of peak intensities x exact masses of peaks) / (sum of peak intensities) "

13
New cards

"relation of monoisotopic mass and average mass values"

"average mass always higher than monoisotopic mass "

14
New cards

mass resolution formula

"(mass of isotope peak) / (delta mass), with delta mass being the width of the peak at half intensity "

15
New cards

mass accuracy formula

"(experimental mass – true mass) / (true mass), in ppm"

16
New cards

Why is tandem mass spectrometry used?

"peptide mass fingerprint is not sufficient for unambiguous identification "

17
New cards

tandem MS principle

"Select individual molecule ions (precursor ion), induce fragmentation, analyze fragment spectra (product) "

18
New cards

parts of tandem mass spectrometer and functions

"

Ion source - ionization
Mass analyzer - ion separation 
Collision Cell – Fragmentation 
Mass analyzer - ion separation 
Detector - ion detection 
"

19
New cards

What are the three defined modes of fragmentation in TMS and their associated letters?

between alpha carbon and carboxyl - a,x
within peptide bond - b,y
between amino group and alpha carbon - c,z

20
New cards

What mode of peptide fragmentation is most common in collision induced/associated dissociation?

within peptide bond - b/y

21
New cards

"Product Ion Scan (TMS)"

"Select single m/z ion from sample, induce collision in cell, m/z scan products "

22
New cards

"Precursor Ion Scan (TMS)"

"Select one peptide after the other, induce collision induced dissociation, mass analyzer filters and lets past indicator ion with defined m/z "

23
New cards

"Neutral Loss Scan (TMS)"

"Both analyzers set to scanning mode with offset (as opposed to filter), see which peptides give rise to lost amount "

24
New cards

Which TMS scan(s) is(are) used to detect PTMs?

"Precursor Ion Scan, Neutral Loss Scan "

25
New cards

What are the two types of particle dissociation?

collision induced/associated and electron transfer

26
New cards

electron transfer dissociation

"chemical is charged with electron, collision causes transfer along backbone until cleavage occurs "

27
New cards

Why would you use electron transfer dissociation as opposed to collision induced dissociation?

not all peptides are efficiently cleaved by CID, such as long ones and those with PTMs

28
New cards

what are the steps of protein identification?

"Extraction, separation (gels), digestion (trypsin), peptide separation, sample ionization, MS, data analysis "

29
New cards

what information is matched up to identify proteins?

"information obtained from MS and information contained in protein and/or DNAseq databases "

30
New cards

probability based Mowse score formula

"-10Log(P)"

31
New cards

what must the Mowse score P value be greater than to be statistically significant? (non-random match)

70

32
New cards

what should the deviation of MS and calculated data masses be?

within instrument mass accuracy

33
New cards

what factors could cause a low probability score or peptide ambiguities?

"

Sample is not yet sequenced 
Imperfect sample separation 
Isoenzymes (closely related) 
PTMs (changing ion mass, reads as negative hit) 
Miscleavages 
Contamination with trypsin and keratin 
Mass accuracy low 
MS calibration 
Complexity of spectra increased due to salt adduct or fragmentation 
"

34
New cards

what are the two ways to use MS/MS to improve probability score?

autocorrelation and probability based matching

35
New cards

autocorrelation (MS/MS)

"comparison between theoretical MS/MS spectra of all peptides in database and experimental data "

36
New cards

probability based matching (MS/MS)

"comparison between theoretical MS/MS spectra of all peptides in database and experimental data, starting with most intense b and y peaks in top down method (y then b) "

37
New cards

what are some difficulties in detecting PTMs?

"

Transient 
Small fraction of peptides modified 
Multiple modifications within same protein/peptide 
In vivo degradation 
Labile during extract preparation 
Lost during sample workup 
Labile in mass spectrometer 
Not all peptides detected 
Increased search space 
"

38
New cards

what are some strategies for finding PTMs?

"

Compare modified and non-modified sample (measure MS sample peaks for before and after treating)
Search for indicator ions 
Data dependent neutral loss MS/MS/MS scan 
Enrichment of modified peptides 
"

39
New cards

two forms of labeling for quantitative proteomics

"SILAC (stable isotope labeling with amino acids in cell culture) and dimethyl labeling"

40
New cards

SILAC process

"

  1. Two cell states (generally one normal and one supplied with heavy arg or lys isotopes) 
  2. Combine lysates and purify proteins 
  3. Digest and mass spec 
  4. Peptides differ in mass, quantify levels from peak ratio 
"

41
New cards

dimethyl labeling process

"

  1. Treat cells, lyse, and digest separately 
  2. Separately label with different reagents (formaldehyde variants) 
  3. Mass spec, Quantify levels from peak ratio 
"

42
New cards

what two tags are used for protein quantification using MS/MS?

"ITRAQ (isobaric tag for relative and absolute quantification) and TMT (tandem mass tag)"

43
New cards

how does MS/MS quantification work? (tags)

"both ITRAQ and TMT tags contain indicator ion and balance group
intial MS scan forms one peak of all samples
MS/MS scan shows indicator ion
"

44
New cards

advantages of MS quantification

"

Label introduction early in sample preparation 
Quantification with several data points during elution 
"

45
New cards

disadvantages of MS quantification

"

Expensive 
Not possible with all samples 
Increased sample complexity limits multiplexing 
All peaks to be compared have to be sequenced by MS/MS 
"

46
New cards

disadvantages of MS/MS quantification

"Expensive

Introduction of label late during sample preparation 
Quantification only in single MS/MS spectra not during several elution points 
"

47
New cards

advantages of MS/MS quantification

"

Peptide ID and quantification within same spectrum 
High multiplex capacity 
No increase in sample complexity in MS 
"