1/67
Looks like no tags are added yet.
Sample
contains solvent, analyte, interferents, and more
Analyte
compound that you are testing for within a sample
Homogeneous sample
sample composition is the same everywhere (pure chocolate bar)
Heterogeneous sample
material composition differs from region to region (chocolate bar with nuts)
Interferent
species other than the analyte that increases or decreases the response of the analytical method & must be removed when prepping sample
Chromatography
separates analytes based on their affinity for the solvent vs particulate and measures quantity of each compound
In chromatography if a compound passes through the column fast then it has a ____ affinity for the solute
high
In chromatography of a compound passes through the column slowly then it has a ____ affinity for the solute
low
Calibration curve
measures detector response for known concentrations of each analyte & graphs peak height vs concentration
Must be linear & unknown must fall within concentration range
electric balance
uses electromagnetic force to compensate for a load placed on the pan
readability
smallest increment of mass that can be indicated on an electronic balance
hygroscopic reagents
absorb moisture from the air which will affect the weight, so it must be weighed using a capped bottle
ambient temperature
temperature of surroundings
what your sample must be at when weighing
micropipet
delivers small volumes of liquid in a controlled manner
often 1 - 1000 micro liters
1 μ (micro) liter = _____ liters
1 × 10-6
experimental error
difference between the true (expected) value and measured value
Systematic (determinate) error
occurs due to a flaw in equipment or experiment design & will occur again if you reproduced the experiment
can be discovered by:
analyze a known sample
use a different method to measure amounts
have different labs analyze identical samples
analyze blank sample
Can be corrected by: calibration of glassware and instruments
Ex: serial dilution done improperly
formula for serial dilutions
C1 x V1 = C2 x V2
Random (indeterminate) error
caused by uncontrolled variables in the measurements and can be reduced but NEVER eliminated
equal chance of being positive or negative
always present
Ex: subjective reading of a burette
Precision
describes the reproducibility of a result

Accuracy
how close a measured value is to the ‘true’ value

Uncertainty
variability within a set of measurements, can be absolute or relative
absolute uncertainty
expresses the margin of uncertainty associated with any measurement, typically expressed as ±
however, it cannot determine if it is a high or low margian of error, this is why you use relative uncertainty
Relative uncertainty
compares the size of the absolute uncertainty with the size of its associated measurement

propagation of uncertainty


absolute uncertainty


Gaussian Distibution
experiment results tend to cluster symmetrically around an average value

symbol for mean
x̄
symbol for standard deviation
s
equation for standard deviation (for sample)


equation for relative standard deviation (RSD)


variance equation


standard deviation definition
measures how closely data are clustered about the mean
as precision increases, standard deviation ______ & the gaussian curve becomes more _____
decreases, narrow
degrees of freedom equation
n - 1
You can decrease uncertainty by…
I) a factor of 2 by making ____ many measurements
II) a factor of 4 by making ____ many measurements
I) 4 times II) 16 times
Standard deviation of the mean ux equation


F-Test definition
compares two variances (S2) to determine if the null hypothesis should be accepted or rejected
Equation for F calculated


null hypothesis
two sets of data are drawn from populations with the same properties (statistically similar); will be rejected if there is <5% probability of observing experimental results from two populations with the same value
If F calculated < F table
accept null hypothesis
confidence interval equation


t test
determines if there is a statistical difference between mean 1 and mean 2
if t calculated > t table
reject null hypothesis
Grubbs Test
tests to see if a variable is an outlier, can only be used with one outlier value
Grubbs test equation


is G calculated > G table
questionable point is an outlier and should be discarded
standards
known quantities that are used to compare when finding unknown concentrations
method of least squares
used to draw the ‘best’ straight line through experimental data points that contain some scatter
equation y = mx + b
when using method of least squares you should assume…
uncertainty in y values is much greater than uncertainty in x values & uncertainties of all y values are similar
standard solutions
contain known concentrations of analyte
blank solutions
contain all reagents and solvents used in the analysis but contain NO deliberately added analyte
linear range
concentration range over which calibration curve is linear
dynamic range
concentration range over which there is measurable response
when calculating an unknown concentration use the
linear range
detection limit definition
smallest quantity of analyte that is significantly different from the blank
quantitation limit definition
smallest quantity of analyte that can be measured with reasonable accuracy
false positives
when a blank falsely produces a signal above the detection limit and none is present
false negatives
when an analyte fails to produce a signal above the detection limit which causes it to appear there is no analyte present
signal detection limit

calibration line equation

detection limit equation

Quantitation limit equation

Reporting limit
concentration below which regulations say an analyte is ‘not detected’
does NOT mean analyte is not observed, just below prescribed level
set 5 to 20 times higher than detection limit
Matrix effect
change in analytical sensitivity caused by something in the sample other than analyte & makes traditional calibration curves unreliable
standard addition definition
known quantities of the analyte added to the unknown (standard in different volumes and IunknownI stays the same)
mainly used when sample composition is unknown, complex, or affects the analytical signal
Standard addition equation


Equation for (Vs)0
-b / m