Imaging CT, MRI, Portal, and Ultrasound

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/96

encourage image

There's no tags or description

Looks like no tags are added yet.

97 Terms

1
New cards

What are the layers of an EPID panel?

Cover

air gap

Copper plate

Scintillator (CsI)

Photodiode array

TFT array

2
New cards

What is a direct conversion flat panel?

Photoconductor converts x-ray energy into electrical charge that is recorded in the TFT array

3
New cards

What is an indirect conversion flat panel?

X-rays are converted into visible light

That light is then converted into an electrical signal

Needs a scintillator

4
New cards

How does an EPID panel work?

x-rays interact with a copper layer where Compton scattering interaction take place

Electrons from the copper interact with a scintillator to produce visible light

The light reaches an array of photodiodes which convert light into electrical charge

Charge is stored for each pixel and is proportional to the detected signal

Thin film transistors (TFT) sequentially read out the charge from each pixel

The electrical signals are digitized and reconstructed into a 2D image

5
New cards

What are EPID panels used for?

position verification

Treatment field verification

dosimetric measurements

6
New cards

What is the difference between MV and kV EPID panels

MV panels: lower contrast from Compton scattering, need copper plate to improve response

kV panels: great photoelectric contribution, higher contrast

7
New cards

Advantages and limitations of indirect EPID

widely used technology

good sensitivity

light spread can reduce resolution

scintillator afterglow can casue lag

8
New cards

Advantages and limitations of direct EPIDs

avoids scintillator light spread

potentially better spatial resolution

charge trapping can affect signal

requires high voltage across photoconductor

9
New cards

What corrections are needed for EPIDs?

Flood field maps: correct for detector response non-uniformity


10
New cards

What are flex maps?

used to correct for changes in EPID response casued by mechanical flexing of the detector as its position or orientation changes

Gravity can cause mechanical deformation of the panel and support arm leading to changes in the detector’s geometry/pixel position relative to the beam

11
New cards
<p></p>


Dead pixel map

represents the locations of pixels that don’t respond correctly

Acquired using a flood image

Values can be replaced with the average value of surrounding pixels

12
New cards
term image

dark image correction

Some pixel values have background values greater than zero (noise)

Dark image is acquired without radiation and is subtracted out

13
New cards
term image

Gain image correction (detector sensitivity)

Accounts for differences in detector sensitivity

Gain image is acquired by averaging a series of homogenous fields and is also corrected using the dark image

14
New cards

How does CBCT work?

Planar projection images are acquired as the x-ray source and panel move around the patient

3D volumetric images are reconstructed from the projections using filtered backprojection

15
New cards

Why is CT image quality better than CBCT?

Geometry

Fan beam used in CT has less scatter radiation that reaches the detectors

Longitudinal divergence of the beam in CBCT causes geometric artifacts

16
New cards

What is T1 relaxation?

Spin-lattice interaction

Result of dissipation of absorbed energy into the molecular lattice

17
New cards

What is T2 relaxation?

Spin-spin interactions

Comes from loss of phase coherence of individual spins caused by magnetic inhomogeneities of surrounding spins

18
New cards

What are the TR and TE for a T1 weighted image?

Short TR (500 ms): maximizes differences in longitudinal magnetization

Short TE (8 ms): suppresses T2

19
New cards

What are the TR and TE for a T2 weighted image?

Long TR (3000 ms): reduces T1 effects

Long TE (100 ms): maximizes T2 differences

20
New cards

What are the TR and TE for a proton density images?

Long TR (2400 ms)

Short TE (30 ms)

21
New cards

Parameters for a head CT

120 kVp

350 mAs

0.5 pitch

50 mGy

22
New cards

Parameters for an abdomen CT

120 kVp

150 mAs

0.5 pitch

15 mGy

23
New cards

What is the impact of increasing peak potential (kVP0?

decreased patient dose

decreased soft tissue contrast

24
New cards

What is the impact of increasing beam current (mA)

increased patient dose

reduced image noise

25
New cards

What is the impact of increasing pitch?

decreased scan time

decreased patient dose

decreased image resolution

increased image noise

26
New cards

What contrast agents are used in CT?

Iodine - vascular studies

Barium - gastrointestinal studies

27
New cards

What are the CT parameters for a 4DCT

120 kVP

800 mAs

0.1 pitch

80 mGy

28
New cards

High vs low compressibility in ultrasound

High compressibility = slow speed

greater stiffness = high speed

29
New cards

What is the speed of sound for air, tissue, and water?

Air = 330 m/s

tissue = 1540 m/s

water = 1480 m/s

30
New cards

What is the range of medical ultrasound

1-20 MHz

50+ for specialized applications

31
New cards

What impacts the speed of sound in different tissues

acoustic impedance - product of density and speed of sound

basis for echo formation

reflection and refraction occur at tissue boundaries that have different acoustic impedance

32
New cards

What is the impact of higher frequencies in ultrasound

provides better resolution

higher attenuation

limited depth of penetration

33
New cards

How are ultrasound pulses created

Pulses are created using a short duration voltage spike in the compression mode

The piezoelectric vibrates at a natural frequency in the thickness mode of l/2

Crystal thickness determines frequency

34
New cards

Why is gel used with ultrasound?

used to improve contact and eliminate air pockets that impact the speed of sound

35
New cards

What are the types of resolution in ultrasound

Axial: ability to distinguish closely spaced objects in the direction of the beam (1/2 spatial pulse length)

Lateral: perpendicular to the beam (beamwidth is determining factor)

Elevational: slice thickness resolution (depth dependent)

36
New cards

How does ultrasound work

ultrasound waves are transmitted into the patient

Partial reflections from tissue boundaries create echoes that return to the transducer as a function of travel time and depth (acoustic impedance)

Receiver detects echoes to form images

37
New cards

TG-66

CT simulator QA

38
New cards

TG-233

CT sim performance

39
New cards

TG-291

Multi energy CT

40
New cards

TG-126

PET/CT acceptance and QA

41
New cards

TG-284

MRI sim

42
New cards

TG-128 and TG-154

Ultrasound

43
New cards

TG-58 and TG-307

EPIDs

44
New cards

TG-104

In room kV imaging

45
New cards

TG-75 and TG-180

image guidance doses

46
New cards

TG-76

Respiratory motion management

47
New cards

How do you measure coincidence of PET and CT components

phantom that has wires with radioactive solution

Image using both

Tolerance is 1 PET voxel

48
New cards

What are the HU and ED values for lung, soft tissue, bone, and titanium?

Lung: 300 HU and 0.3 g/cc

Soft tissue: 100 HU and 1.05 g/cc

Bone: 500 HU and 1.3 g/cc

Titanium: HU saturates and 4.3 g/cc

49
New cards

How do we test for B0 homogeneity

two images with different TE and post process them

50
New cards

What artifacts would a hip implant cause

photon starvation (not enough signal reaches the detector)

Beam hardening - streaking

51
New cards

How do you get rid of artifacts from a hip implant

take an MV image because more photons would get through

IR DECT can get rid of beam hardening but not starvation

52
New cards

What are the components of a CT

X-ray tube – produces the x-ray beam

Filtration and collimation – shape and filter the beam

Detector array – measures

Gantry – the home for the x-ray tube and detectors

Bow Tie filter – attenuates the lateral portion of the beam

53
New cards

CT elevator pitch

The x-ray tube rotates around the patient to acquire projections from different angles

As the x-ray passes through the patient it will be attenuated by different tissue types

After the x-ray has passed through the patient the detector element measures the remaining intensity

The projection data is collected and processed to reconstruct cross sectional images

54
New cards

What is filtered backprojection?

Takes the projections from the different gantry angles and evenly projects them backward across the slice

With many different angles used an image will start to form

Back projection alone will cause there to be 1/r blurring

A ramp filter can be used to eliminate the blurring by suppressing low spatial frequencies, but increases noise from the unsuppressed high frequency portion of k-space

Roll off is used to reduce the high frequency noise amplification

Convolution with a kernel helps with sharpness and smoothing

55
New cards

What is iterative reconstruction?

starts with an estimated image and repeatedly compares the forward-projected data from that image with the actual measured projection data. The algorithm calculates the difference (error) and updates the image until the difference is minimized.

56
New cards

What is the impact of thinner slices?

Thinner slices will improve longitudinal spatial resolution but cause an increase in noise

57
New cards

What is pitch?

Pitch is the ratio of the distance the CT table travels during one 360° gantry rotation to the total beam collimation width.

  • Pitch = 1: Table advances one beam width per rotation, with no overlap or gap in nominal coverage.

  • Pitch greater than 1: Table moves farther per rotation, resulting in less overlap, faster scanning, and generally lower radiation dose. However, image noise and artifacts may increase depending on scanner settings and reconstruction.

  • Pitch less than 1: Table moves less per rotation, resulting in overlapping data, increased radiation dose, and potentially improved image quality.


58
New cards
<p></p>


Ring artifact

detector element malfunction and calls for recalibrating the pixel map

59
New cards
term image

Beam hardening

Lower-energy photons are preferentially attenuated by dense, higher-Z structures, so the average energy shifts upward with increasing path length.

60
New cards
term image

Streak artifact

attenuation exceeds the dynamic or effective linear range of the detectors

61
New cards
term image

Photon starvation

Too few photons surviving the longest densest path, projections are dominated by noise

62
New cards
term image

Partial volume

occur when the reconstructed voxel dimension is large relative to structures in the patient, so a voxel encompasses several tissue types and their attenuation is averaged, diluting the contrast of the small structure

63
New cards
term image

Aliasing

too few projection views to reconstruct high-frequency objects

64
New cards
term image

Cupping

mostly caused by scattered radiation, with beam hardening contributing and also causing streaking

Reduce by anti scatter grid between the object and detector

65
New cards

Why are PET/CT scanners combined?

Images are easy to register since the patient is in the same position

CT gives attenuation correction factors to get a more accurate PET image

66
New cards

MRI elevator pitch

Strong magnetic field is applied to a patient

Protons align with the magnetic field and precess

RF energy is applied to convert longitudinal magnetization into the transverse plane

After the pulse, lateral magnetization decays and protons dephase back to longitudinal plane

Receiving coils detect emitted RF waves and record signal

Signal comes from the changes in bulk magnetization

67
New cards

How are MRI images reconstructed?

All voxels in a row are read out simultaneously

Data has to be deconvolved by an inverse Fourier transform in k-space

K-space is a frequency map

68
New cards
term image

Susceptibility

Usually appears as darkening

Comes from changes in the magnetic field from ferromagnetic, paramagnetic, or diamagnetic materials

69
New cards
term image

Wrap around

Most common in the phase encode direction

Caused by sampling below the Nyquist frequency (lowers apparent frequency)

Increase field of view, change phase encode direction, or oversampling

70
New cards
term image

Distortion

Non-uniformity of the magnetic field that leads to inaccurate encoding of the spatial position

Check for proper shimming, use lower field strength scanner, place ROIs in the field center (area of lowest distortion)

71
New cards
term image

Chemical shift

Appears as dark or light edges on borders with different chemical compositions

Differences in intrinsic shielding of structures\

Use a higher bandwidth or use a lower field strength scanner

Fat and water signals

72
New cards

What are the differences between types 1 and 2 chemical shifts?

Type 1 causes spatial displacement of fat relative to water, while Type 2 causes signal cancellation at fat–water boundaries.

73
New cards

Ultrasound Elevator pitch

Uses high frequency sound waves to produce images

We are measuring the waveform (need time and amplitude)

Transducer emits an ultrasound beam that travels through the patient’s body

Different tissue have different acoustic impedances which impact how much of the ultrasound power is reflected to the transducer to be detected

The computer uses the time between emission and reflection to determine the object distance

Produces an image based on the distance to reflection and the intensity of the reflection

74
New cards

Describe an ultrasound transducer

Used to produce and detect the ultrasound waves

Takes electrical signals and converts them into mechanical signals

Use piezoelectric crystals to create sound pressure waves

Crystal mechanically deforms when electrical signal is applied creating the ultrasound wave

75
New cards

What are the QA tests for ultrasound

knowt flashcard image
76
New cards
term image

Shadowing

Region of hypo intense signal often distal to high attenuation objects

77
New cards
term image

Enhancement

Hyper intense signal that is distal to low attenuation objects

78
New cards
term image

Reverberation

Hyper intense repeating signal

Repeated reflections between two closely spaced objects

79
New cards
term image

Speed Artifact

Abrupt mis-mapping of an object along the beam axis

Caused by variations of the speed of sound between beam projections

80
New cards
term image

Mirroring

Second inverted object appears beyond a highly reflective surface

Caused by multiple beam reflections between the object and a highly reflective surface

81
New cards

Describe MV imaging

Uses the linac’s photon beam and imaging panel to get portal images

Primarily used for patient positioning and alignment using bony anatomy

Doesn’t have good soft tissue contrast because there is greater Compton scatter and less photoelectric effect

82
New cards

Describe kV imaging

kV imaging

Uses an x-ray source and detector panel that are mounted onto the linac

Gives better image contrast and spatial resolution than MV imaging

Generally, delivers less imaging dose than MV

83
New cards

What corrections are needed for EPID images

  1. Dark-field correction: Removes background signal and electronic noise present when no radiation is delivered.

  2. Flood-field (gain) correction: Corrects for variations in pixel sensitivity and detector response to produce a uniform image.

  3. Bad-pixel correction: Identifies and interpolates defective or nonresponsive detector pixels.

  4. Geometric correction: Corrects image distortion, scaling, and alignment errors.

  5. Backscatter correction: Accounts for radiation scattered from the EPID support arm and detector housing.

  6. Lag/ghosting correction: Accounts for residual signal from previous irradiations that can affect subsequent images.



84
New cards

What are the MRI safety zones

Zone 1 uncontrolled

Zone 2 uncontrolled, where screening occurs

Zone 3 controlled, control room, level 1 escort

Zone 4 MRI, level 2 escort.

85
New cards

MRI QA Tolerances

Percent integral Uniformity >82%f or 3T or 87.5% for 1.5T and below)

Percent signal ghosting < 2.5%

LCV: >38 of 40 spokes for 3T. System dependent

Distortion < 1 mmSRS, 2 mm other

Geometric accuracy 190 ± 2mm

Resolution 1 mmshould be visible (10 lp/cm)

Slice thickness 5 ± 0.7 mm (this is testing RF BW)

Slice position: this is testing slice selection gradient

B0 homogeneity RMS < 0.5 ppm. (1.5 μT variation for 3T*0.5/1E6)

B0 field drift <0.25 ppm/day(after commissioning)

Transmitter gain <5%(analogous to 5 HU tolerance for water on CT)

Transmitter gain = power required (on 𝐵HRF pulse) to flip by desired flip angle

86
New cards

What is the attenuation for ultrasound

0.5 dB/cm/MHz

87
New cards

What are the impacts of increasing kV

raises the max energy of the bremsstrahlung spectrum

shifts the spectrum to higher energies

exposure is proportional to kV²

88
New cards

How is CTDI measured

100 mm pencil ionization chamber in a 16 or 32 cm PMMA phantom

normalized by nominal beam width

volume correction 1/3 center 2/3 peripheral

89
New cards

Amplitude sorting

bins by geometric displacement level of respiratory trace

Can miss data if amplitudes vary

90
New cards

Phase sorting

divides each cycle into equal time fractions

assumes each respiratory cycle has the same duration and shape

91
New cards

What are the EPID requirements for dosimetry

positioning: 2mm

Linearity: 2%

Uniformity: 2%

Reproducibility: 2%

92
New cards

What impacts MRI slice thickness

z = BW/(gyromagnetic ratio * slice select gradient strength)

93
New cards

Typical imaging doses

CT: 10-30 mGy

4DCT: 20-60 mGy

MV: 2 cGy

kV; 1 mGy

kV CBCT: 10s mGy

MV CBCT: 10s cGy

94
New cards

Explain the different gradients used in MRI

MRI uses three magnetic field gradients to localize the signal. First, a slice-selection gradient is applied during RF excitation, causing protons at different positions to have different Larmor frequencies. An RF pulse excites only protons within the desired frequency range, selecting a slice. Next, a phase-encoding gradient is briefly applied, producing position-dependent phase shifts along one direction. Finally, a frequency-encoding gradient is applied during signal acquisition, producing position-dependent frequencies along the other direction. The acquired signals are stored in k-space and reconstructed into an image using a Fourier transform.

95
New cards

How is k-space filled?

K-space is filled line by line using phase and frequency encoding. After slice selection, a phase-encoding gradient is applied to establish a particular phase shift. A frequency-encoding gradient is then applied during signal acquisition, collecting data across one line of k-space. The sequence repeats with different phase-encoding gradient strengths until all required lines are collected. The center of k-space contains low spatial frequencies that primarily determine image contrast, while the periphery contains high spatial frequencies that determine fine detail. Once filled, a 2D Fourier transform converts k-space data into the final image.

96
New cards

What contrast agents are used for MRI

Gadolinium

Iron oxide

97
New cards

Why is the CT to ED curve different for different energies

TC-to-electron density curves differ at 80 and 120 kVp because the X-ray energy spectrum changes with tube voltage, and different tissues have energy-dependent attenuation properties.

Lower kVp produces lower-energy photons, increasing the relative contribution of photoelectric absorption, particularly in high-atomic-number materials such as bone.