22 Terms

1
New cards

Primary types of electron microscopes

Scanning Electron Microscope (SEM) and Transmission Electron Microscope (TEM)

2
New cards

Operating principle of Scanning Electron Microscopy (SEM)

Fires electrons onto the sample surface and records electrons that bounce back from the surface

3
New cards

Operating principle of Transmission Electron Microscopy (TEM)

Transmits electrons through a thin specimen and records elastic and inelastic scattering onto a detector on the opposite side

4
New cards

Key physical requirement for TEM specimens

Specimens must be ultra-thin so electrons can transmit through without creating a completely dark image

5
New cards

Dual-column microscope setup used in sample preparation

A FIB-SEM system containing an SEM column on top for imaging and a Focused Ion Beam (FIB) column at an angle for milling

6
New cards

Function of Focused Ion Beam (FIB) milling

Blasts high-current ion beams at frozen cells to ablate unwanted cellular material and generate a thin specimen section

7
New cards

Ideal thickness sweet spot for a cryo-FIB lamella

Approximately 200 nanometers thick

8
New cards

Strategy used during FIB milling to protect biological structures

Starting with harsh high-current ion beams far from the target, then switching to delicate lower currents closer to the biological feature

9
New cards

Titan Krios microscope function

A state-of-the-art 300 kV transmission electron microscope used for high-end cryo-EM and cryo-ET data acquisition

10
New cards

Purpose of acquiring a tilt series in cryo-electron tomography

Collecting a sequence of 2D projection images while incrementally tilting the sample stage to enable 3D computational volume reconstruction

11
New cards

Standard angular tilt range for sample stages in cryo-ET

Typically from minus 60 degrees to plus 60 degrees

12
New cards

Primary physical factor limiting electron dose and total tilt images

Radiation damage, as excessive electron exposure destroys fragile biological structures

13
New cards

Central Slice Theorem definition

The 2D Fourier transform of a projection image corresponds to a central slice through the 3D Fourier transform of the 3D object

14
New cards

Cause of the missing wedge problem in tomography

Mechanical stage rotation limits capped at around plus or minus 60 degrees prevent collecting high-tilt projection data

15
New cards

Consequence of the missing wedge on 3D tomograms

Severe anisotropic resolution, missing structural information, and elongation artifacts along the beam direction

16
New cards

Nature of a cryo-EM micrograph capture

Recording a rapid movie sequence of multiple fast frames rather than a single static exposure

17
New cards

Primary purpose of recording fast movie frames in cryo-EM

Enables tracking and correcting sample movement and stage drift before averaging frames together

18
New cards

EER file format advantage in direct electron detection

Electron Event Representation records exact XY coordinates and arrival timing of individual electron hits for maximum spatial and temporal resolution

19
New cards

Standard file formats used in cryo-electron microscopy

MRC, TIF, and EER files

20
New cards

Main sample-related cause of motion during imaging

Beam-induced motion caused by electron beam charging of the ice, leading to subtle particle displacement and micro-lensing

21
New cards

Main mechanical cause of image blur during tilt acquisition

Mechanical instability and subtle drift of the microscope stage arm holding the sample grid

22
New cards

Why individual tilt series images have very low signal-to-noise ratio

The total allowed electron dose must be split across dozens of tilt images to avoid destroying the sampl