Primary types of electron microscopes
Scanning Electron Microscope (SEM) and Transmission Electron Microscope (TEM)
Operating principle of Scanning Electron Microscopy (SEM)
Fires electrons onto the sample surface and records electrons that bounce back from the surface
Operating principle of Transmission Electron Microscopy (TEM)
Transmits electrons through a thin specimen and records elastic and inelastic scattering onto a detector on the opposite side
Key physical requirement for TEM specimens
Specimens must be ultra-thin so electrons can transmit through without creating a completely dark image
Dual-column microscope setup used in sample preparation
A FIB-SEM system containing an SEM column on top for imaging and a Focused Ion Beam (FIB) column at an angle for milling
Function of Focused Ion Beam (FIB) milling
Blasts high-current ion beams at frozen cells to ablate unwanted cellular material and generate a thin specimen section
Ideal thickness sweet spot for a cryo-FIB lamella
Approximately 200 nanometers thick
Strategy used during FIB milling to protect biological structures
Starting with harsh high-current ion beams far from the target, then switching to delicate lower currents closer to the biological feature
Titan Krios microscope function
A state-of-the-art 300 kV transmission electron microscope used for high-end cryo-EM and cryo-ET data acquisition
Purpose of acquiring a tilt series in cryo-electron tomography
Collecting a sequence of 2D projection images while incrementally tilting the sample stage to enable 3D computational volume reconstruction
Standard angular tilt range for sample stages in cryo-ET
Typically from minus 60 degrees to plus 60 degrees
Primary physical factor limiting electron dose and total tilt images
Radiation damage, as excessive electron exposure destroys fragile biological structures
Central Slice Theorem definition
The 2D Fourier transform of a projection image corresponds to a central slice through the 3D Fourier transform of the 3D object
Cause of the missing wedge problem in tomography
Mechanical stage rotation limits capped at around plus or minus 60 degrees prevent collecting high-tilt projection data
Consequence of the missing wedge on 3D tomograms
Severe anisotropic resolution, missing structural information, and elongation artifacts along the beam direction
Nature of a cryo-EM micrograph capture
Recording a rapid movie sequence of multiple fast frames rather than a single static exposure
Primary purpose of recording fast movie frames in cryo-EM
Enables tracking and correcting sample movement and stage drift before averaging frames together
EER file format advantage in direct electron detection
Electron Event Representation records exact XY coordinates and arrival timing of individual electron hits for maximum spatial and temporal resolution
Standard file formats used in cryo-electron microscopy
MRC, TIF, and EER files
Main sample-related cause of motion during imaging
Beam-induced motion caused by electron beam charging of the ice, leading to subtle particle displacement and micro-lensing
Main mechanical cause of image blur during tilt acquisition
Mechanical instability and subtle drift of the microscope stage arm holding the sample grid
Why individual tilt series images have very low signal-to-noise ratio
The total allowed electron dose must be split across dozens of tilt images to avoid destroying the sampl