Radiographic Quality Control: Equipment Tests, Tolerances, and Frequencies

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/30

encourage image

There's no tags or description

Looks like no tags are added yet.

31 Terms

1
New cards

What is Quality Assurance (QA)?

Focuses on PEOPLE and the overall quality of health care.

2
New cards

What is another name for QA?

Continuous Quality Improvement (CQI).

3
New cards

What is the primary goal of QA?

Provide an accurate and timely diagnosis.

4
New cards

What is the secondary goal of QA?

Minimize radiation exposure to the patient.

5
New cards

What is Quality Control (QC)?

Focuses on EQUIPMENT and the actual testing of imaging equipment.

6
New cards

What is another name for QC?

Total Quality Management (TQM).

7
New cards

What is the purpose of QC?

Evaluate imaging equipment to ensure safe and reliable operation.

8
New cards

What are the types of QC testing listed?

Acceptance testing, preventive maintenance (PM) testing, and error testing.

9
New cards

What is the total filtration formula?

Total filtration = inherent filtration + added filtration.

10
New cards

What is the minimum total filtration above 70 kVp?

At least 2.5 mm aluminum equivalent.

11
New cards

What is the frequency of filtration or beam quality tests?

Annually.

12
New cards

What is the half-value layer (HVL)?

Amount of filtration that reduces x-ray beam intensity or air kerma to 50% of its original value.

13
New cards

How does HVL relate to beam hardness?

Higher HVL = harder beam with higher average energy. Lower HVL = softer beam.

14
New cards

What is the collimation light field to radiation field alignment tolerance?

Within 2% of the source-to-image receptor distance (SID).

15
New cards

What is the frequency of collimation alignment tests?

Semi-annually (every 6 months).

16
New cards

What is the Positive Beam Limitation (PBL) tolerance?

Automatic collimators must be within 2% of SID.

17
New cards

What is the SID accuracy tolerance?

Actual SID must be within ±2% of the indicated SID.

18
New cards

What is the frequency of SID accuracy tests?

Annually.

19
New cards

What is the centering tolerance?

The x-ray beam center must align with the centering indicator within 1% of SID.

20
New cards

What is the frequency of centering tests?

Annually.

21
New cards

What is the effective focal spot size tolerance?

Within 50% of the size stated in the equipment specifications.

22
New cards

What is the frequency of effective focal spot tests?

Annually.

23
New cards

What is the kVp calibration tolerance?

Actual kVp must be within 5% of the indicated kVp.

24
New cards

What is the frequency of kVp calibration tests?

Annually.

25
New cards

What is the timer accuracy tolerance?

Exposure time must be within 5% of the indicated exposure time.

26
New cards

What is the timer test method and frequency?

Spinning top test; performed annually.

27
New cards

What is exposure linearity?

Exposure per mAs should not change more than 10% between different mA stations.

28
New cards

What is the frequency of exposure linearity tests?

Annually.

29
New cards

What is exposure reproducibility?

Exposure values produced using identical technical parameters should be within 5%.

30
New cards

What is the frequency of exposure reproducibility tests?

Annually.

31
New cards

What is the line focus principle?

Explains the relationship between the actual focal spot on the anode and the effective focal spot size