1/41
Looks like no tags are added yet.
Increasing MAs = _____ quantity of primary beam reaching the PT
Increase
Increase mas = _____ quantity and quality of exit beam reaching the IR
Increase
Decreasing MAs = _____ quantity in primary beam and exit beam
Decrease
Increasing kvp = _____ in quantity and quality of primary beam and exit beam
Increase
Decreasing kvp = _____ in quantity and quality of primary beam and exit beam
Decrease
Smaller and larger focal spot size has what effect on primary beam and exit beam
No effect
Increasing SID has what effect on quantity in primary beam and exit beam
Decreasing quantity
Decreasing Sid has what effect on quantity in primary beam and exit beam
Increase in quantity
Increasing or decreasing OID has what effect on primary beam
No effect
Increasing OID has what effect on exit beam reaching the IR
Decrease in quantity and scatter
Decreasing OID has what effect on exit beam reaching the IR
Increase in quantity and scatter
Increase CR angle =
Decrease in quantity of primary beam and exit beam
Increasing/decreasing grid ratio has what effect on primary beam
No effect
Increasing grid ratio has what effect on exit beam
Decrease in quantity and scatter
Decreasing grid ratio has what effect on exit beam
Increase in quantity and scatter
Increasing collimation has what effect in primary beam
Decreased quantity
Decreasing collimation has what effect in primary beam
Increased quantity
Increasing collimation has what affect on exit beam
Decrease quantity and scatter
Decreasing collimation has what effect on exit beam
Increase in quantity and scatter
Single phase generator has what effect on primary beam and exit beam
Decrease in quantity and quality - primary beam
Decrease in quantity and quality - exit beam
High frequency generator has what effect on primary beam and exit beam
Increased quantity and quality - primary beam
Increased quantity and quality - exit beam
Adding compensating filers has what effect on primary and exit beam
Decrease in quantity on both
Increase MAs =
Increase in IR exposure
No effect in brightness
No effect in contrast
No effect in spatial resolution
No effect in distortion
Decrease in MAs =
Decrease in IR exposure
No effect in brightness
No effect in contrast
No effect in spatial resolution
No effect in distortion
Increase in kvp=
Increase in IR exposure
No effect in brightness
Decrease in contrast
No effect on spatial resolution
No effect on distortion
Decrease in kvp=
Decrease in IR exposure
No effect in brightness
Increase in contrast
No effect spatial resolution
No effect distortion
Increase focal spot size =
No effect on IR exposure
No effect on brightness
No effect on contrast
Decrease in spatial resolution
No effect on distortion
Decrease focal spot size=
No effect on IR exposure
No effect on brightness
No effect on contrast
Increase in spatial resolution
No effect on distortion
Increased SID=
Decrease in IR exposure
No effect in brightness
No effect in contrast
Increase in spatial resolution
- magnification in distortion
Decreased SID =
Increase in IR exposure
No effect in brightness
No effect in contrast
Decrease in spatial resolution
+ magnification in distortion
Increased OID =
Decrease IR exposure
No effect in brightness
Increase in contrast
Decrease in spatial resolution
+ magnification in distortion
Increased CR angle =
Decreased IR exposure
No effect on brightness
No effect on contrast
Decrease in spatial resolution
+ shape distortion in distortion
Add grid =
Decrease in IR exposure
No effect in brightness
Increase in contrast
No effect in spatial resolution
No effect distortion
Remove grid=
Increase in IR exposure
No effect in brightness
Decrease in contrast
No effect in spatial resolution
No effect in distortion
Increase collimation =
Decrease IR exposure
No effect on brightness
Increase in contrast
No effect on spatial resolution
No effect in disortion
Decrease in Collimation =
Increase IR exposure
No effect on brightness
Decrease in contrast
No effect on spatial resolution
No effect in disortio
Excessive tube filtration =
Decrease IR exposure
No effect on brightness
Decrease in contrast
No effect on spatial resolution
No effect in disortio
Insufficient tube filtration =
Increase IR exposure
No effect on brightness
Increase in contrast
No effect on spatial resolution
No effect in disortio
Increase in PT thickness =
Decrease in IR exposure
No effect in brightness
Decrease in contrast
Decrease in spatial resolution
+ magnification in distortion
Decrease in PT thickness =
Increase in IR exposure
No effect in brightness
Increase in contrast
Increase in spatial resolution
- magnification
Patient motion has no effect on everything but what
Decrease in spatial resolution