Quality Assurance (QA)
The complete radiology quality program. It addresses overall quality, including customer service, image interpretation, diagnostic accuracy, and distribution of radiologists’ reports.
2
New cards
Quality Control (QC)
The part of the quality program that specifically addresses the safe and reliable operation of imaging equipment.
3
New cards

The Joint Commission.

Quality Control Program Requirement Mosby states that a quality control program is required by

4
New cards
Quality Control Program Responsibility
QC may involve radiologists, radiology managers, radiation physicists, QC technologists, and staff radiographers depending on department size.
5
New cards
Filtration / Beam Quality Test
Tests beam quality with a digital dosimeter; a half-value layer measurement is required.
6
New cards
Half-Value Layer (HVL)
The amount of material needed to reduce x-ray beam intensity to one-half; used to evaluate beam quality and filtration.
7
New cards
Light Field–Radiation Field Alignment Test
Checks whether the visible collimator field matches the actual radiation field. Mosby: within ±2% of SID.
8
New cards
SID Indicator Accuracy Test
Checks whether the indicated SID agrees with the actual source-to-image receptor distance. Think: displayed distance versus actual distance.
9
New cards
Central Ray Alignment / Congruency Test
Checks whether the central ray and center of the Bucky/image receptor are properly aligned. Mosby digital QC: within ±1% of SID.
10
New cards
Effective Focal-Spot Size Test
Checks effective focal-spot size against the equipment specification. Mosby: should be within 50% of the stated size.
11
New cards
kVp Accuracy Test
Checks whether the actual kVp matches the selected kVp. Mosby: accurate within ±5% of indicated kVp.
12
New cards
Timer Accuracy Test
Checks whether actual exposure time matches selected time. Mosby: within 5% for exposures longer than 10 ms.
13
New cards
Exposure Linearity Test
Compares adjacent mA stations to ensure radiation output changes appropriately. Mosby: adjacent mA stations within 10% of one another.
14
New cards
Exposure Reproducibility Test
Repeats the same exposure to determine whether the machine produces consistent radiation output. Mosby: sequential exposures should not vary by more than 5%.
15
New cards
AEC Reproducibility / Phantom Test
Uses phantoms of different thicknesses to verify that AEC produces consistent receptor exposure despite changes in patient thickness.
16
New cards
AEC Detector Calibration
Tests AEC accuracy using a variety of exposures to ensure consistent performance. Mosby digital QC: within ±5%.
17
New cards
AEC Backup Timer
A safety device that stops the exposure if the AEC fails to terminate it, preventing an excessively long exposure.
18
New cards
Fluoroscopy Exposure-Rate Test
Uses a digital dosimeter to measure entrance skin exposure rate. Mosby page reviewed: no more than 100 mGy/min.
19
New cards
Automatic Brightness Control (ABC) QC
Check performed on an image intensifier to ensure the radiation dose reaching the input phosphor remains consistent.
20
New cards
CR Plate Reader Calibration
CR plates/readers are tested regularly to ensure accurate representation of images.
21
New cards
CR Erasure Thoroughness Test
Verifies that the CR system completely erases residual energy so ghosting artifacts do not appear on later images.
22
New cards
CR Plate Uniformity Test
A uniformly exposed CR plate should produce uniform image brightness. Mosby: areas should be within ±0.2 of each other.
23
New cards
Spatial Resolution QC Test
A wire-mesh pattern is exposed and examined for geometric distortion and image blur. Mosby: no more than 2% image blur.
24
New cards
Grayscale Standard Display Function
Ensures displayed contrast and gray shades remain consistent throughout the pixel-value range of the image.
25
New cards
Luminance Uniformity Test
Checks that monitor brightness is uniform across the entire screen. Mosby measures several screen areas; readings should be within 20% of each other.
26
New cards
Luminance
The amount of light/brightness emitted by a display monitor.
27
New cards
CR (Computed Radiography)
Uses an imaging plate containing PSP that stores x-ray energy and must be scanned by a laser in a separate reader.
28
New cards
PSP (Photostimulable Phosphor)
The material in a CR imaging plate that absorbs and stores energy from the remnant x-ray beam until laser stimulation.
29
New cards
CR Reader
The device that removes/scans the CR plate with a laser so stored energy is released as visible light and processed into a digital image.
30
New cards
ADC (Analog-to-Digital Converter)
Converts an analog electrical signal into numerical digital data for computer processing.
31
New cards
DR (Digital Radiography)
Uses a flat-panel detector connected to the computer workstation and produces an image more quickly than CR because no separate plate reader is required.
32
New cards
Technologist Workstation
Used primarily to review acquired images for positioning and exposure before they are sent for interpretation.
33
New cards
Radiologist Workstation
Uses a high-resolution diagnostic monitor so the radiologist can evaluate maximum detail and contrast resolution.
34
New cards
PACS
Picture Archiving and Communication System; used for image acquisition, storage, retrieval, management, and distribution. Think: PACS = pictures/archive.
35
New cards

DICOM

the standard protocol that allows imaging systems and modalities to communicate and exchange image information.

36
New cards

RIS

manages information and workflow specific to the radiology department, such as procedures and reporting.

37
New cards

HIS

manages information across the hospital rather than only the radiology department.

38
New cards

EHR

the patient's electronic health information/medical record used across healthcare.