Principles of Chromatography Separation
Separates mixtures between a mobile phase (liquid or gas) and a stationary phase (solid, or liquid on a solid support). Solid stationary phases separate by adsorption, whereas liquid stationary phases separate by relative solubility. Separation depends on the balance between solubility in the mobile phase and retention by the stationary phase (e.g., with a polar stationary phase and non-polar hexane mobile phase, non-polar compounds move faster due to greater solubility in hexane).
Thin-Layer Chromatography (TLC) Method & Precautions
Wear gloves (prevents skin contamination), draw a pencil baseline 1 cm up (ink would dissolve), spot tiny drops (prevents spots merging), place in a sealed chamber (prevents toxic solvent evaporation) with solvent < 1 cm deep (below pencil line so spots don't dissolve into solvent), remove 1 cm from top, mark solvent front, dry in fume cupboard, view colourless spots under UV lamp (or spray ninhydrin), and calculate Rf = distance moved by spot / distance moved by solvent.
Two-Directional TLC
A single mixture spot is separated with one solvent, then the TLC plate is rotated 90° and run in a second solvent. Used when components have identical Rf values in the first solvent or are insoluble in the first solvent.
Column Chromatography (CC) & HPLC
In CC, a glass column is packed with powdered solid stationary phase (silica or alumina, for high surface area) retained by a filter/plug, kept covered with solvent, and the mixture is run through to measure retention times. HPLC (High-Performance Liquid Chromatography) uses a solid silica stationary phase and a liquid mobile phase under high pressure.
Gas-Liquid Chromatography (GC) Principles
Separates volatile liquids using an inert carrier gas mobile phase (helium or nitrogen, so it does not react with sample components) and a high-boiling-point liquid absorbed onto a solid stationary phase. Retention time identifies substances; the number of peaks shows the number of components, and peak area is proportional to component abundance.
Factors Affecting GC Retention Time & GC-MS
Higher oven temperature, higher carrier gas flow rate, or a shorter column makes substances move faster, giving shorter retention times. Linking GC to a mass spectrometer (GC-MS) generates mass spectra for each separated component to match against a computer database for positive identification (used in forensics, airport security, environmental analysis, and space probes).