Introduction to Medical Devices and Hospital Environment

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/31

flashcard set

Earn XP

Description and Tags

Flashcards reviewing medical devices, hospital environments, diagnostic tools, therapeutic equipment, laboratory instruments, radiology systems, health IT, and SMPS power electronics.

32 Terms

1
New cards

Identify two medical devices commonly found in each of the following hospital areas: Intensive Care Unit (ICU), Obstetrics and Gynaecology (O&G) Ward, and Imaging Department.

  1. Intensive Care Unit (ICU): Mechanical ventilator, multi-parameter patient monitor (or infusion pump, defibrillator). 2. Obstetrics and Gynaecology (O&G) Ward: Cardiotocograph (CTG), fetal Doppler (or infant incubator). 3. Imaging Department: X-ray machine, ultrasound machine (or CT scanner, MRI scanner).


2
New cards

Briefly describe the following three categories of medical devices: Radiology & Imaging, Laboratory, and Diagnostics.

  1. Radiology & Imaging: Devices that use radiation, magnetic fields, or sound waves to create visual representations of internal body structures for clinical assessment (e.g., X-ray, CT, Ultrasound). 2. Laboratory: Equipment utilized to analyze biological specimens such as blood, urine, or tissue to aid diagnosis and research (e.g., Centrifuge, Spectrophotometer, Blood Gas Analyzer). 3. Diagnostics: Equipment used to detect, measure, and evaluate physiological signals to identify medical conditions (e.g., ECG, Spirometer, Patient Monitor).
3
New cards

A 3232-year-old patient named Daniel arrives at the emergency department with severe chest pain and shortness of breath. Which five medical devices could be used during his emergency assessment and treatment?

  1. Electrocardiograph (ECG) machine. 2. Multi-parameter patient monitor. 3. Defibrillator. 4. Volumetric infusion pump or syringe pump. 5. Medical suction unit (or emergency mechanical ventilator).
4
New cards

Classify the three risk levels of medical devices and give one example for each class.

  1. Class I (Low Risk): Minimal potential for harm (e.g., thermometer, manual stethoscope, tongue depressor). 2. Class II (Moderate Risk): Moderate potential for harm (e.g., ECG machine, syringe pump, volumetric infusion pump). 3. Class III (High Risk): Life-supporting, life-sustaining, or high potential for serious harm (e.g., defibrillator, mechanical ventilator, cardiac pacemaker).
5
New cards

Assign each of the following three patients to their appropriate specialized hospital clinic: Patient A has persistent chest pain and irregular heartbeat; Patient B is a 44-year-old child with persistent fever and loss of appetite; Patient C has knee pain and difficulty walking.

  1. Patient A: Cardiology Clinic. 2. Patient B: Paediatric Clinic. 3. Patient C: Orthopaedic Clinic.
6
New cards

During emergency ambulance transit of an injured mother with breathing difficulties and an infant needing temperature control, identify two appropriate devices used and their roles.

  1. Multi-parameter Transport Monitor (or Pulse Oximeter): Continuously tracks the mother's unstable vital signs, such as heart rate, oxygen saturation (SpO2SpO_2), and blood pressure. 2. Transport Incubator: Provides a temperature-controlled, stable microenvironment to protect the infant from hypothermia while monitoring vital functions.
7
New cards
<p>Identify the hospital environment depicted in the image and state two essential medical devices located in this setting along with their functions.</p>

Identify the hospital environment depicted in the image and state two essential medical devices located in this setting along with their functions.

The image shows an Operating Theatre (OT) or surgical suite. Two essential devices: 1. Operating Theatre (OT) Surgical Light: Provides high-intensity, shadowless illumination over the surgical field. 2. Operating Table: Safely supports and adjusts patient positioning during surgical interventions (other valid devices include the anaesthesia machine, patient monitor, and electro-surgical unit).

8
New cards

What is the definition and primary clinical function of a cardiotocograph (CTG)?

A Cardiotocograph (CTG) is an electronic diagnostic monitoring device used during pregnancy and labor to simultaneously record and display the Fetal Heart Rate (FHR) and uterine contractions.

9
New cards

If a cardiotocography (CTG) machine displays the Fetal Heart Rate (FHR) on screen but fails to print the FHR chart on paper, what three immediate steps should technical personnel take?

  1. Check the paper tray to ensure thermal recording paper is loaded, oriented correctly, and not jammed. 2. Verify that the recorder/print function is activated and that the paper door/latch is securely closed. 3. Inspect the thermal printhead and printer cable connections for dust, damage, or dislodgement.
10
New cards
<p>Identify the biomedical diagnostic device shown in the figure and state two critical operational safety procedures to observe during its use.</p>

Identify the biomedical diagnostic device shown in the figure and state two critical operational safety procedures to observe during its use.

The device is an Electrocardiograph (ECG) machine. Two critical safety procedures: 1. Ensure the device and patient cables are properly grounded and electrically isolated to avoid microshock hazard. 2. Verify skin contact sites are clean and dry, use fresh conductive gel or proper electrode placements, and avoid placing electrodes over broken skin or uninsulated conductive surfaces.

11
New cards

Name the two transducers utilized in a Cardiotocograph (CTG) and specify the physiological parameter monitored by each.

  1. Ultrasound Transducer: Measures Fetal Heart Rate (FHR) via pulsed Doppler ultrasound. 2. Tocodynamometer (TOCO transducer): Measures the frequency and relative strength of maternal uterine contractions using a pressure-sensitive gauge.
12
New cards

List four standard physiological parameters monitored by a multi-parameter patient monitoring system.

  1. Electrocardiogram (ECG) / Heart Rate. 2. Pulse Oxygen Saturation (SpO2SpO_2). 3. Non-Invasive Blood Pressure (NIBP). 4. Body Temperature (or Respiration Rate).
13
New cards

Contrast the transmission mode and reflection mode of pulse oximeter (SpO2SpO_2) sensors.

  1. Transmission Mode: The light emitters (red and infrared LEDs) and the photodetector are placed opposite each other; light travels completely through the tissue bed (e.g., finger clip sensor, earlobe). 2. Reflection Mode: The light emitters and the photodetector are placed side by side on the same tissue surface; the detector measures backscattered light reflected off underlying tissue and bone (e.g., forehead sensor).
14
New cards

Describe Einthoven's Triangle, including the polarity and electrode attachments for Leads I, II, and III.

Einthoven's Triangle is an imaginary equilateral triangle centered on the heart formed by three standard bipolar limb leads: 1. Lead I: Negative electrode on the Right Arm (RA), positive electrode on the Left Arm (LA). 2. Lead II: Negative electrode on the Right Arm (RA), positive electrode on the Left Leg (LL). 3. Lead III: Negative electrode on the Left Arm (LA), positive electrode on the Left Leg (LL). (The Right Leg [RL] serves as reference ground).

15
New cards
<p>Identify the diagnostic device shown in the figure and list five safety and operational procedures required before operating it.</p>

Identify the diagnostic device shown in the figure and list five safety and operational procedures required before operating it.

The device is a Spirometer (Schiller Spirovit SP-1). Five procedures: 1. Fit a clean, single-use disposable mouthpiece or viral/bacterial filter to prevent cross-contamination. 2. Perform volume calibration using a certified calibration syringe (e.g., 3 dm33\,dm^3 syringe). 3. Verify electrical cord integrity and battery charging status. 4. Ensure ambient conditions (temperature, barometric pressure, humidity) are properly zeroed/calibrated in the system. 5. Instruct and seat the patient comfortably to prevent falls due to hyperventilation or dizziness.

16
New cards

List four common clinical types of infusion devices.

  1. Volumetric Infusion Pump. 2. Syringe Pump. 3. Patient-Controlled Analgesia (PCA) Pump. 4. Ambulatory / Elastomeric Infusion Pump.
17
New cards
<p>Identify the two infusion devices shown in the image and describe their primary functional distinction.</p>

Identify the two infusion devices shown in the image and describe their primary functional distinction.

Left: Syringe Pump; Right: Volumetric Infusion Pump. Functional distinction: The syringe pump is designed for precise, low-volume, and low-flow-rate delivery of concentrated, potent drugs (e.g., inotropes, anaesthetics). The volumetric infusion pump delivers larger volumes of intravenous fluids, medications, or total parenteral nutrition (TPN) over sustained periods at higher flow rates.

18
New cards

State two cardiac arrhythmias treatable with a manual defibrillator, and name the clinical condition associated with an Automated External Defibrillator (AED).

Manual Defibrillator: 1. Ventricular Fibrillation (VF). 2. Pulseless Ventricular Tachycardia (pVT). Automated External Defibrillator (AED): Sudden Cardiac Arrest (SCA) exhibiting shockable ventricular fibrillation or pulseless ventricular tachycardia.

19
New cards

Distinguish between the two operational modes of a defibrillator: Asynchronous mode and Synchronized mode.

  1. Asynchronous Mode (Defibrillation): Delivers electrical discharge immediately when triggered, without regard to the ECG rhythm; indicated for uncoordinated rhythms like Ventricular Fibrillation. 2. Synchronized Mode (Cardioversion): Synchronizes shock delivery with the peak of the ECG's R-wave to prevent discharging during the vulnerable T-wave repolarization phase (preventing VF induction); indicated for unstable ventricular tachycardia with pulse or atrial fibrillation.
20
New cards

State three primary functions of a hospital anaesthesia workstation.

  1. Deliver precisely regulated concentrations of medical gases (oxygen, nitrous oxide, medical air) and volatile anaesthetic vapors. 2. Provide continuous artificial ventilation and oxygenation for patients under general anaesthesia. 3. Remove exhaled carbon dioxide (CO2CO_2) and scavenge waste anaesthetic gases from the operating environment.
21
New cards

Define the roles of the scavenging system, breathing circuit, ventilator, and vaporiser within an anaesthesia workstation.

  1. Scavenging System: Captures and safely evacuates waste anaesthetic gases to prevent operating theatre pollution. 2. Breathing Circuit: Provides the conduit delivering fresh gas and anaesthetic mixtures to the patient's airway and carrying expired gas away. 3. Ventilator: Mechanically inflates and deflates the patient's lungs when spontaneous respiration is suppressed. 4. Vaporiser: Accurately converts liquid volatile anaesthetic agents into vapor and blends them into the fresh gas flow.
22
New cards

Explain the four physical mechanisms of heat loss that must be managed to maintain neonatal body temperature.

  1. Conduction: Direct heat transfer to a colder solid surface in contact with the infant (e.g., cold mattress or scale). 2. Convection: Heat loss from the infant's skin to cooler surrounding air currents. 3. Radiation: Transfer of radiant heat from the infant's body to cooler surrounding objects without direct contact (e.g., incubator walls, cold windows). 4. Evaporation: Heat lost as moisture or fluid evaporates from wet skin or respiratory mucosa.
23
New cards

Describe three good engineering handling practices for maintaining an optical spectrophotometer.

  1. Keep optical paths and sample cuvettes clean; handle cuvettes solely by frosted edges and wipe optical faces with lint-free wipes. 2. Allow adequate warm-up time for the light source lamp to achieve thermal and luminous stability before measurement. 3. Perform routine zero/blank baseline calibration with pure solvent and conduct periodic wavelength accuracy checks.
24
New cards

Describe the function and operating principle of the pCO2pCO_2 and pO2pO_2 electrode systems used in an Arterial Blood Gas (ABG) analyzer.

  1. pCO2pCO_2 Electrode (Severinghaus Electrode): A modified pH glass electrode covered by a gas-permeable silicone membrane and a bicarbonate electrolyte buffer. Dissolved CO2CO_2 diffuses across the membrane, generating carbonic acid which alters buffer pH proportionally to the pCO2pCO_2. 2. pO2pO_2 Electrode (Clark Electrode): A polarographic sensor featuring a platinum cathode and silver/silver chloride anode behind an oxygen-permeable membrane. Oxygen diffuses across the membrane and undergoes electrochemical reduction at the polarized cathode, generating a measurable electric current directly proportional to pO2pO_2.
25
New cards
<p>Based on the optical microscope diagram, identify the components designated by letters A through F.</p>

Based on the optical microscope diagram, identify the components designated by letters A through F.

A: Eyepiece (Ocular Lens); B: Coarse and Fine Adjustment Knobs; C: Revolving Nosepiece / Objective Lens (40×40\times shown); D: Specimen Stage (with Slide Clips); E: Condenser and Iris Diaphragm; F: Light Source (Illuminator).

26
New cards
<p>Identify the laboratory equipment shown in the figure and name the two primary rotor types used in this device.</p>

Identify the laboratory equipment shown in the figure and name the two primary rotor types used in this device.

The device is a Laboratory Centrifuge (Hettich MIKRO 220). The two primary rotor types are: 1. Fixed-Angle Rotor (holds sample tubes at a rigid, fixed angle during rotation). 2. Swing-Out / Swinging-Bucket Rotor (allows buckets to swing horizontally perpendicular to the rotational axis during spinning).

27
New cards
<p>Identify the medical imaging device shown and name three of its key structural or operational sub-assemblies.</p>

Identify the medical imaging device shown and name three of its key structural or operational sub-assemblies.

The device is an Ultrasound Imaging System. Key components include: 1. Ultrasound Transducer / Probe (houses piezoelectric crystals). 2. Central Processing Unit (CPU) / Beamformer Console (processes returning acoustic echoes). 3. Video Display Monitor (visualizes real-time B-mode, M-mode, or Doppler images).

28
New cards

During Planned Preventive Maintenance (PPM) of a diagnostic X-ray machine, name three key parameters evaluated and the test analyzer used.

Key parameters: 1. Tube potential accuracy (kVpkVp). 2. Exposure duration / timer accuracy (msms). 3. Radiation output linearity and repeatability (mAsmAs or mGymGy). Test analyzer: Non-invasive X-ray Multimeter / Beam Quality Analyzer (with solid-state detectors or ionization chamber).

29
New cards

Azhar suffered a broken leg, chest swelling, and suspected internal brain bleeding following a road accident. Which radiology/biomedical diagnostic devices are indicated for each injury?

  1. Broken Leg: X-ray Radiography system (to visualize bone fractures and bone alignment). 2. Chest Swelling: ECG machine (to assess cardiac rhythm and detect cardiac trauma) and Chest X-ray / Ultrasound. 3. Internal Brain Bleeding: Computed Tomography (CT) Scanner (or MRI) of the head to detect intracranial hemorrhage and tissue injury.
30
New cards

List five key components comprising a hospital Picture Archiving and Communication System (PACS).

  1. Medical imaging modalities (e.g., CT, MRI, X-ray, Ultrasound). 2. Secure high-speed PACS local area network (LAN/WAN). 3. Central PACS archive/server storage system. 4. Diagnostic display workstations. 5. DICOM interface and Hospital/Radiology Information System (HIS/RIS) gateway software.
31
New cards

Explain two primary distinctions between an Electronic Medical Record (EMR) and an Electronic Health Record (EHR) regarding data sharing and accessibility.

  1. Data Sharing Scope: An EMR is an internal digital patient chart restricted to a single clinical facility or practice and is not designed to easily travel outside it. An EHR contains a comprehensive health summary designed to be shared across multiple healthcare organizations, specialists, and laboratories. 2. Accessibility & Interoperability: EMR access is limited to internal clinic staff; EHRs are built on standardized interoperability protocols allowing authorized clinicians across different healthcare facilities and regional health networks to collaborate and view patient history.
32
New cards

What are four core safety precautions a biomedical technician must execute prior to opening and troubleshooting a Switch Mode Power Supply (SMPS)?

  1. Completely disconnect the unit from the AC mains power outlet. 2. Discharge all high-voltage primary bulk filter capacitors safely using a power resistor before touching internal components. 3. Use an AC isolation transformer when connecting diagnostic test equipment (such as an oscilloscope) to primary floating circuits. 4. Wear safety eyewear, remove conductive personal jewellery, and observe electrostatic discharge (ESD) and single-hand testing protocols on energized points.