Objective measurements
-”objective” tests don’t require a behavioral repsonse
provide information about the function of various portions of the auditory system (but not perception!)
-Acoustic immittance test battery
tympanometry
acoustic reflex threshold
acoustic reflex decay
Acoustic impedance and admittance
-immitance = a hybrid term that refers to both acoustic impedance and acoustic admittance
acoustic impedance (Za)- the total opposition to the flow of sound energy (measured in ohms)
ratio of sound pressure (P) to sound flow (U) Za = P/U
acoustic admittance (Ya)- the total ease with which sound energy flows through a system (measured in mhos)
-These values have a reciprocal relationship
Ya = U/P or Ya = 1/Za
Acoustic immittance
-the ease (admittance) or opposition (impedance) of sound energy trasnmission through the external and middle ear can be affected by conductive pathologies-
middle ear effusion
eardrum perforation
ossicular disarticulation
otosclerosis/stapes fixation
-Sensory or neural pathologies have minimal effec
What factors affect impedance and admittance (immittance)?
-ease, or opposition, of energy transmission through a system (immittance) is affected by the:
system mass (its inertia)
system stiffness/elasticity/compliance (its restoring force)
system resistance (friction)
-The mass, stiffness and resistance of the system components modulate the total impedance/admittance of the system
Terms used to describe the effects of mass, stiffness, and friction for impedance
mass reactance, stiffness reactance, and resistance
Terms used to describe the effects of mass, stiffness, and friction for admittance
mass susceptance, stiffness susceptance (or compliance) and conductance
components of impedance
imagine how energy transfer through this system would be affected by the mass of the block, the stiffness of the spring, and the friction (resistance) of the surface the block moved on
components of admittance
these terms are the reciprocal of the components of impedance and dictate the ease of energy flow through the system.
frequency, impedance, and resonance
-opposition to energy flow due to mass and stiffness reactance varies with freq
mass reactance increases with freq
stiffness reactance decreases with freq
opposition due to resistance is constant with freq
-a system “resonates” at a certain freq based on its mass an stiffness
resonant freq occurs when “net reactance” is smallest
Immittance of our system
-immittance is due to
stiffness of
air in the EAC and ME
TM, ME ligaments and muscle contractions
mass of
ossicles, TM
system friction
perilymph, tendons, ligaments
immittance changes (primarily changes in stiffness) are associated with various pathologies
Clinical measurement of acoustic immittance
-clinical instruments measure the “admittance” of the ear in millimhos (mmhos) or milliliters (mLs)
most also provide separate measures of susceptance (B) and conductance (G)
components:
probe loud speaker
microphone
air pump
a second loudspeaker for reflex testing
-present pure tone at 85 dB SPL
generally LF 226
measuring middle ear admittance (vs total system admittance)
-diagnostically it is helpful to differentiate ME (YME) and total admittance (Ytotal)
-”stiffen” TM to measure admittance of EAC only (YEAC)
stiffen TM by increasing EAC pressure to +200 daPa (2000 Pa)
-we can infer ME admittance
YME = Ytotal - YEAC
Estimating ear canal volume via admittance measures
-In terms of admittance, a cavity (like the EAC) is essentially a tube of air- a stiffness susceptance
the stiffness/admittance of a tube of air varies systematically with its volume (cavity size)
For a 226 Hz tone, admittance (in mmhos) is equal to its volume (in mLs or cc)
this is one reason why we use a 226 Hz tone
-with a 226 Hz tone we can estimate ear canal volume while measuring middle ear admittance!
EAC volume can be diagnostically important
indicative of TM perforation
clogged or patent PE tube or
occluded EAC
What measures make up the immitance test battery?
-tympanometry: admittance as a function of EAC pressure
static acoustic admittance
compensated/uncompensated
tympanometric peak pressure
tympanometric width/gradient (we don’t use the gradient measure)
-Ear canal volume
-ipsilateral and contralateral acoustic reflex thresholds
-These are VERY informative measures that should be routinely obtained as part of the diagnostic test battery, but
-immitance measures should NOT be completed when contra-indications exist, e.g.
recent middle ear surgery (within 6 months)
patient reports extreme discomfort
if the referral source specifically requests no immittance
Tympanometry
the measurement of acoustic immittance in the external auditory meatus as a function of air pressure within the external auditory meatus