RTR.6 Analyzing IQ and Responding

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/32

encourage image

There's no tags or description

Looks like no tags are added yet.

33 Terms

1
New cards

Spatial Resolution

  • ability to image small objects that have high subject contrast

  • affected by focal spot size and detector material, smaller the better


2
New cards

Contrast Resolution

  • ability to image small objects that have high subject contrast

  • improves with low kVp


3
New cards

Subject Factors that Affect IQ

  • subject contrast

  • patient thickness → thicker = more absorption

  • tissue mass density → denser = more absorption

  • atomic number → higher = more scatter

  • object shape


4
New cards

Window Width

the number of Hounsfield units represented on a specific image

5
New cards

Wide WW

  • Best for imaging tissues with high contrast  

  • Reduces image contrast, but also suppresses noise, thus often used for obese pts or for metallic objects 


6
New cards

Narrow WW

best for low contrast tissues eg. brain

7
New cards

Window Level

  • The centre value of the WW thus selects which Hounsfield values are displayed as shades of gray

  • Should be set at a point that is roughly the same value as the average attenuation number of the tissue of interest


8
New cards

Factors that Increase Scatter Radiation

  • high kVp and also high mA

  • large field size

  • increased patient thickness


9
New cards

Lossless Image Compression

can be reconstructed to be the same as the original image, 10:1 or 2:1

10
New cards

Lossy Image Compression

can provide compression factors of up to 100:1 or greater; not acceptable for archiving any images as it represents a lack of data

11
New cards

XR Preprocessing Components

  • interpolation = for dead pixels

  • flatfielding = equalizes the response of each pixel to a uniform X-ray beam, eg., anode heel effect


12
New cards

CT Artifact - Aliasing

  • Insufficient projection data (aka undersampling), due to extended helical pitch

  • Consist of evenly spaced lines, easy to distinguish from anatomic structures, thus rarely render an image undiagnostic


13
New cards

CT Artifact - Beam Hardening

  • Can produce either cupping artifacts (the periphery of the image is lighter) or streak artifacts (of dark bands or streaks between dense objects)

  • Filtration, bow tie filter, increase kVp, select appropriate SFOV


<ul><li><p><span style="background-color: transparent;">Can produce either cupping artifacts (the periphery of the image is lighter) or streak artifacts (of dark bands or streaks between dense objects)</span></p></li><li><p><span style="background-color: transparent;">Filtration, bow tie filter, increase kVp, select appropriate SFOV</span></p></li></ul><p></p>
14
New cards

CT Artifact - Partia Volume

  • When a dense object is lying at the edge of FOV, it may only show up in a small number of views

  • Use thinner slices


<ul><li><p><span style="background-color: transparent;">When a dense object is lying at the edge of FOV, it may only show up in a small number of views</span></p></li><li><p><span style="background-color: transparent;">Use thinner slices</span></p></li></ul><p></p>
15
New cards

CT Artifact - Edge Gradient

  • Due to irregularly shaped objects that have a pronounced difference in density from surrounding structures

  • Unavoidable, but can be reduced by thinner slices


<ul><li><p><span style="background-color: transparent;">Due to irregularly shaped objects that have a pronounced difference in density from surrounding structures</span></p></li><li><p><span style="background-color: transparent;">Unavoidable, but can be reduced by thinner slices</span></p></li></ul><p></p>
16
New cards

CT Artifact - Ring

  • Caused by imperfect detector elements—either faulty or simply out of calibration

  • Appear as a ring or concentric rings centred on the rotational axis


<ul><li><p><span style="background-color: transparent;">Caused by imperfect detector elements—either faulty or simply out of calibration</span></p></li><li><p><span style="background-color: transparent;">Appear as a ring or concentric rings centred on the rotational axis</span></p></li></ul><p></p>
17
New cards
<p>What type of Artifact is this?</p>

What type of Artifact is this?

Motion Artifact

18
New cards
<p>What type of Artifact is this?</p>

What type of Artifact is this?

Metallic Artifact

19
New cards
<p>What type of Artifact is this?</p>

What type of Artifact is this?

Out of Field Artifact

  • Caused by anatomy beyond the SFOV hardening the beam, but ignored in the reconstruction process eg. arms being by the side


20
New cards

QC - Exposure Reproducibility

ensuring that an X-ray generator always produces the same intensity each time with the same set of technical factors; maximum variability is +-5%

21
New cards

QC - Beam Quantity

X-ray generators should emit a specific amount of radiation per unit of X-ray tube current and time; maximum variability is +-10%

22
New cards

QC - Kilovolt Accuracy

the kV stated on the control panel should produce an X-ray beam with a comparable and consistent amount of energy; maximum variability is +-5%

23
New cards

QC - Timer Accuracy

to ensure accurate exposure timer; +-20% for exposure times less than 10 ms

24
New cards

QC - mA and Exposure time Linearity and Reciprocity

  • to ensure constant radiation output with various mA and exposure time combinations

  • eg. g. 50 mA at 1sec and 100 mA at 0.5sec, both with the same kVp, should yield the same amount of radiation output

  • +-10%


25
New cards

QC - Linearity

sequential increase in mAs should produce the same sequential increase in the exposure measured, +-10%

26
New cards

QC for DSA

needs a weekly QC to evaluate DSA function with a DSA phantom

27
New cards

QC for Fluoro

Quarterly QC to evaluate table angluation accuracy and smooth motion

28
New cards

QC - Electronic Display Device Performance

Yearly, SMPTE and/or TG-18 to assess for any

  • Distortion

  • Reflection

  • Luminance

  • Resolution

  • Noise

  • Glare

  • Chromacity (quality of colour)


<p>Yearly, <span>SMPTE and/or TG-18 to assess for any </span></p><ul><li><p>Distortion</p></li><li><p>Reflection</p></li><li><p>Luminance</p></li><li><p>Resolution</p></li><li><p>Noise</p></li><li><p>Glare</p></li><li><p>Chromacity (quality of colour)</p></li></ul><p></p>
29
New cards

Bolus (arterial) Phase

  • Immediately follows injection

  • HU difference of 30 or more between the aorta and IVC


<ul><li><p><span style="background-color: transparent;">Immediately follows injection</span></p></li><li><p><span style="background-color: transparent;">HU difference of 30 or more between the aorta and IVC</span></p></li></ul><p></p>
30
New cards

Non-equilibrium (venous) Phase

  • 1min post injection

  • HU difference between 10-30


<ul><li><p><span style="background-color: transparent;">1min post injection</span></p></li><li><p><span style="background-color: transparent;">HU difference between 10-30</span></p></li></ul><p></p>
31
New cards

Equilibrium (delayed) Phase

  • 2min post injection

  • HU difference less than 10


32
New cards

Wide Histogram =

  • peaks on the far ends of the graph = lots of white and black shades

  • high contrast image


<ul><li><p>peaks on the far ends of the graph = lots of white and black shades</p></li><li><p>high contrast image</p></li></ul><p></p>
33
New cards

Narrow Historgram

peak is at the center of the graph = similar shades of gray, low contrast