1
New cards
Average photon energy of a bremsstrahlung beam
Approximately one-third of the maximum/nominal energy.
2
New cards
Co-60 photon energies
1.17 and 1.33 MeV.
3
New cards
Average Co-60 photon energy
~1.25 MeV.
4
New cards
Co-60 dmax
~0.5 cm.
5
New cards
6 MV dmax for 10×10 cm²
~1.5 cm.
6
New cards
10 MV dmax for 10×10 cm²
~2.4 cm.
7
New cards
18 MV dmax for 10×10 cm²
~3.3 cm.
8
New cards
PDD at 10 cm for Co-60, 10×10 cm²
~55%.
9
New cards
PDD at 10 cm for 6 MV, 10×10 cm²
~65%.
10
New cards
PDD at 10 cm for 10 MV, 10×10 cm²
~75%.
11
New cards
PDD at 10 cm for 18 MV, 10×10 cm²
~80%.
12
New cards
TMR at 10 cm for 6 MV
~0.78.
13
New cards
TMR at 10 cm for 10 MV
~0.85.
14
New cards
TMR at 10 cm for 18 MV
~0.90.
15
New cards
Approximate attenuation rate for 6 MV
~3% per cm.
16
New cards
Approximate attenuation rate for 10 MV
~2.5% per cm.
17
New cards
Approximate attenuation rate for 18 MV
~2% per cm.
18
New cards
Surface dose for Co-60
~50%.
19
New cards
Surface dose for 6 MV
~25%.
20
New cards
Surface dose for 10 MV
~23%.
21
New cards
Surface dose for 18 MV
~20%.
22
New cards
Scatter at 1 m from phantom
~1/1000 of primary beam.
23
New cards
MV f-factor in air, conversion for Roentgent to cGy
0.876 cGy/R.
24
New cards
MV f-factor in tissue, conversion for Roentgen to cGy
~0.97 cGy/R.
25
New cards
I-125 exposure rate constant
1.45 R cm2/mCi-hr
26
New cards
Ir-192 exposure rate constant
4.69 R cm2/ mCi-hr
27
New cards
Physical wedge out-of-field scatter
~2.5% of central axis dose.
28
New cards
Dynamic wedge out-of-field scatter
~1.0% of central axis dose.
29
New cards
Dose 10 cm outside field at dmax
~1% of central axis dose.
30
New cards
Dose 30 cm outside field at dmax
~0.2% of central axis dose.
31
New cards
Intraleaf MLC leakage
~2%.
32
New cards
Interleaf MLC leakage
~3%.
33
New cards
MLC end leakage
~15%.
34
New cards
LINAC head leakage limit
0.1% of central axis dose at 1 m.
35
New cards
Neutron head leakage 10 MV
~0.01%.
36
New cards
Neutron head leakage 18 MV
~0.15%.
37
New cards
6 MeV electron dmax
~1.2 cm.
38
New cards
9 MeV electron dmax
~1.9 cm.
39
New cards
12 MeV electron dmax
~2.1 cm.
40
New cards
20 MeV electron dmax
~2.5 cm.
41
New cards
6 MeV electron surface dose
~80%.
42
New cards
12 MeV electron surface dose
~90%.
43
New cards
20 MeV electron surface dose
~95%.
44
New cards
6 MeV electron x-ray contamination
~1%.
45
New cards
12 MeV electron x-ray contamination
~2%.
46
New cards
20 MeV electron x-ray contamination
~5%.
47
New cards
Lead thickness to stop electrons
Approximately E0(MeV)/2 in mm of lead.
48
New cards
Electron practical range Rp
Approximately E/2 in cm of water.
49
New cards
Electron therapeutic range R80/R90
Approximately E/3 in cm of water.
50
New cards
Electron R50 relation to Energy
R50 ≈ E/2.33 in cm.
51
New cards
Electron penumbra trend with energy
Penumbra generally increases with energy.
52
New cards
Proton range uncertainty common estimate
~3.5% of range plus 1–3 mm.
53
New cards
70 MeV proton range in water
~4 cm.
54
New cards
100 MeV proton range in water
~7.7 cm.
55
New cards
130 MeV proton range in water
12.3 cm
56
New cards
150 MeV proton range in water
~17.7 cm.
57
New cards
190 MeV proton range in water
~23.8cm
58
New cards
200 MeV proton range in water
~26 cm.
59
New cards
230 MeV proton range in water
~33 cm.
60
New cards
250 MeV proton range in water
~38 cm.
61
New cards
Proton calibration protocol often referenced
IAEA TRS-398; AAPM TG-185 for proton reference dosimetry.
62
New cards
LDR dose rate at prescription point
0.5–2 Gy/hr.
63
New cards
HDR dose rate at prescription point
12Gy/hr
64
New cards
Ir-192 half-life
73.8 days.
65
New cards
Ir-192 daily decay rule
~1% per day (~0.94% exact).
66
New cards
HDR source replacement interval
About every 3 months.
67
New cards
approximate dose from Cs-137 to Ir-192
Ir-192 = D(Cs-137) * 0.65
68
New cards
Time 1 m from exposed HDR source to occupational limit
Order of ~1 hour, depending on source strength.
69
New cards
Definition of 1 U, equivalent units
1 cGy*cm2/hr
70
New cards
TG-43 geometry function purpose
Accounts for inverse-square and source geometry effects.
71
New cards
TG-43 radial dose function purpose
Accounts for attenuation and scatter along transverse plane.
72
New cards
TG-43 anisotropy function purpose
Accounts for angular dose variation around source.
73
New cards
I-125 half-life
~59.4 days.
74
New cards
Pd-103 half-life
~17 days.
75
New cards
Cs-131 half-life
~9.7 days.
76
New cards
Cs-137 half-life
~30.2 years.
77
New cards
Co-60 half-life
~5.27 years.
78
New cards
Ra-226 half-life
1600 yrs
79
New cards
Au-198 half-life
2.7 days
80
New cards
I-125 average photon energy
~28 keV.
81
New cards
Pd-103 average photon energy
~21 keV.
82
New cards
Cs-131 average photon energy
~30 keV.
83
New cards
Ir-192 average photon energy
~380 keV.
84
New cards
Au-198 energy
420 keV
85
New cards
Cs-137 energy
660 keV
86
New cards
Ra-226 energy
830 keV
87
New cards
Co-60 photon energies
1250 keV
88
New cards
I-125 patient release limit
less than 1 mR/hr at 1 m.
89
New cards
Pd-103 patient release limit
less than 3 mR/hr at 1 m.
90
New cards
I-131 patient release limit
less than 7 mR/hr at 1 m.
91
New cards
Sealed source leak test removal threshold
More than 185 Bq (0.005 μCi) leakage.
92
New cards
Brachy medical event
Wrong patient, site, or radionuclide; total dose 20%, single fraction 50%
93
New cards
Teletherapy medical event
wrong patient, site; total dose differs by 20% (3+fx); total dose differs by 10% (-3fx)
94
New cards
Point A vs Point B
Point A from tandem, point b from midline
95
New cards
Occupational dose limit
50 mSv/year.
96
New cards
Pregnancy occupational dose limit
5 mSv for entire pregnancy.
97
New cards
Public dose limit
1 mSv/year.
98
New cards
Average occupational exposure in U.S. radiation worker
~2 mSv/year.
99
New cards
Average annual background + medical exposure
~4–5 mSv/year.
100
New cards
Radon annual contribution
~1 mSv/year.