1/36
Name three types of contrast in MRI
T1, T2, PD (Proton Density)
To get a T2W image, how do we set our parameters on the MR scanner?
Long TE, Long TR
To get a T1W image, how do we set our parameters on the MR scanner?
Short TE, Short TR
To get a PD weighted image, how do we set our parameters on the MR scanner?
Short TE, Long TR
If you have Tissue 1 with T2 = 100 ms, T1 = 500 ms, and Tissue 2 with T2 = 300 ms, T1 = 2000 ms, what is the most correct set of parameters to highlight Tissue 2?
TE = 250, TR = 1500 ms
Which material is used as exogenous contrast for MRI?
Gadolinium
Gadolinium can cross the Blood Brain Barrier under which conditions?
When the BBB is compromised (i.e., in a diseased state)
What property of tissues does Gadolinium affect for contrast?
T1
What is the unit for k-space?
Spatial frequency (cm^-1)
Higher frequencies are located where in k-space?
On the edges
Lower frequencies are located where in k-space?
In the middle
A low-pass filter in k-space is when:
high spatial frequencies are lost
A high-pass filter in k-space is when:
low spatial frequencies are lost
The object in image space (our imaging subject) is reconstructed by performing which operation equivalent to the Fourier transform:
Summing all sinusoidal functions containing the spatial frequencies
The time between samples of your FID echo is called
Dwell Time
Movement in k-space is calculated via the:
Area of your imaging gradient
The receiver bandwidth is related to the dwell time via which equation?
BW(Rx) = 1/DW
Movement in k-space is calculated via the __________ of our applied gradients
Vector sum
In a spin echo, the 180 pulse will
invert the polarity of gradient effect in k-space
Your image resolution in any dimension is defined by what in k-space?
k_max in that dimension
Your image FOV in any dimension is defined by what in k-space?
dk in that dimension
Zero-padding k-space means to
Artificially extend k-space by appending zeros to the matrix
A 2D FFT can only be done on ________ k-space points
uniformly spaced
Applying a gradient during echo acquisition will allow us to
pick up position-dependent frequencies
Ghosting artifact is due to
phase mismapping
Which direction is more prone to ghosting artifacts and why?
Phase encoding because the time between steps are longer
What is a valid strategy to reduce motion artifacts?
Patient holds their breath
What is Respiratory Ordered Phase Encoding?
The act of acquiring k-space points in an ordered sequence based on the respiratory cycle
What is the main drawback of swapping frequency and phase encoding in order to reduce motion artifact?
May increase scan time
Aliasing occurs when which of the following conditions is not met?
When your sampling frequency is less than twice your signal frequency
What happens in the image space when your signal is aliased?
Your FOV is too small, causing a “wraparound effect”
The chemical shift effect is a result of
Frequency mismapping, therefore a shift in the frequency encode direction
What’s the best strategy for mitigating chemical shift artifacts without using an additional pulse?
Increase receiver bandwidth
Why does a high susceptibility object cause an artifact in an MR image?
High inhomogeneity causes rapid dephasing, leading to a short T2* and signal loss
What are valid techniques to reduce susceptibility artifacts?
Better shimming, using spin echo, decrease TE
What is the source for the slice cross talk artifact?
An imperfect rectangular slice profile
How does the truncation artifact appear in image space?
As a series of ripples along sharp edges