analog image
continuous X,Y locations
image produced directly on film
cant be manipulated after aquisition in a completely analog system
no quantitative analysis
poor images must be reacquired
digitla image
discrete X, Y locations
images stored in computer memroy
can be processed and manipulated
quantiative analysis possilbe
display settings can be adjusted
ADC = analog to digital converter
converts analog signalrs into digital values that the computer can store
converts continuous positional info into discrete pizel locations
Field of view (FOV)
the area of he patient or object that the camera images
matrix
the grid arrangement in which image events are stored
pixel
individual picture element wihtin the matrix
pixel depth
maximum number of counts that can be recorded in an individual pixel
pixel size equation
fov / matrix dimension
larger matrix = __ pixels
smaller pixels
64×64 matrix size has
large pixel size
higher counts per pixel
lower perceived noise
128×128 matrix has
intermediate pixes
intermediate counts per pixel
intermediate noise
256×256 matrix has
small pizels
lower counts per pixel
higher noise
when total counts remain constant increasing matrix size divides the counts among more pizels so…
smalle rpixels
fewer counts per pixel
greater statistical noise per pixel
64×64 when
counts are low an dexcellent resolution is not essential
128× 128 when
good resolution is needed but counts are not high
256×256 when
good resolution is needed and plenty of counts are available
byte mode
8 bits
1byte memory
255 max counts
possilbe pixel saturation
lower memory requirement
word mode
16 bits
2 bytes memory
65.535 max counts
pixel saturation much less likely
twice as much memor requirement
pixel saturation
occurs when individual pixel receives mroe counts than its storage capacity
computer may stop acquiring in that pixel
stop the entire image acquisition
reset that pixel to zero and being again
static image
single image
shows radiactive distribution at a particular time
dynamic
sequential images
shows physiologic changes over time
gated
images organixed by ECG timing
shows cardiac function
list mode
stores all individual events and associated timing information
allows flexible reconstruction and ananysis
static/frame more requires _ image buffer
one
dynamic mode requried __ buffers
2
one stores the completed fram while the other acquires the next
gates mode required __ buffers
16 image buffers
beats too long or too short may be rejected
list mode requires
much more memory than frame mode
smoothing
reduces statistical noise
makes images appear less grainy
uses mathematical filters to adjust pixel values based on neighboring pizels
can remove real details if applied excessively
repeated smoothing can eventually produce a flat image
digital zoom
magnifies an image
multiplies x and y coordinates beofore assigning events to the matrix
improves displayed spatial resolution up to the cameras limits
decreases field of view
gray scale
displays vount density using different shades of gray
darker shades generally represent higher counts
helps visulize differences in radiactive distribution
color scale
uses colors ot represent diff count densities
often uses combos of red, green and blue
color bar helps idenfify relative count density
Look up table (LUT)
assigns a particular shade of color to each pixel count value
determines how counts are displayed
can be adjusted to change contrast
does not change the underlying number of counts
nonlinear scales
logarithmic scale: increases contrast in low count region
exponential scare: increases contrast in high count region