The Image Intensifier has significantly ___ patient does compared to conventional fluoroscopy
Lower
What is the Input Phosphor?
Converts incident x-rays into light

What is the Photocathode?
Turns the light into electrons

What is the Output Phosphor?
Turns electrons back into light

What is the Brightness Gain formula?
Minification gain x flux gain
What is Minification Gain?
Focuses electrons to a smaller area in order to increase brightness (signal)
What is Flux Gain?
Ratio of more light photons going out than x-ray photons going in
What is Magnification Mode?
1. Less area of the input phosphor is used for the signal
2. Creates a larger image
3. Decreased FOV
4. Better visualization of structures
5. Increased patient dose
How does Magnification Mode increase patient dose?
The signal must be increased to maintain brightness
What is Pulsed-Progressive Fluoroscopy?
1. Beam pulsed instead of continuous
2. Controls image acquisition frame rate
3. Higher frame rate = increased dose

What is the biggest benefit of pulsed-progressive fluroscopy?
Decreases patient dose
What is an example of Continuous Fluoroscopy?
30 pulses per second, 33 msec pulse width
What is an example of Grid-Controlled Fluoroscopy?
15 pulses/sec, 10 msec pulse width
Wider pulse width (more msec) = ___ patient dose
Increased
Temporal Resolution = ___
Time
What is last image hold?
Allows review of most recent image before giving patient another pulse of radiation

What is the main benefit of last image hold?
Reduces patient dose

Limiting fluoroscopic field size ___ patient dose
Reduces
Collimation = decreased ___ (intregral) dose
Volumetric
Collimation means there is ___ tissue irradiated and ___ scatter produced
Less; Less
What type of shutters are used for fluoroscopic collimation?
1. Iris
2. Rectangular

How much of a collimated border should be visualized in fluoroscopic collimation?
1"
What does Dose Area Product (DAP) take into account?
Collimation

DAP is a better indicator of ___ risk
Stochastic

What is DAP measured in?
cGy-cm^2

What is the formula for DAP?
Dose x field size (collimation)

DAP ___ with increasing field size
Increases

What is SSD?
Source-skin distance
Increased ___ reduces ESE
Distance
What is the SSD for fluoroscopes?
15"
What is the SSD for mobile?
12"
Filtration ___ patient ESE
Lowers
What is the HVL for fluoroscopy?
3-4mm Al
What is the total filtration above 70 kVp?
2.5 mm Al
What is the Cumulative Timer for fluoro?
Audible alarm reset after 5 minutes
True or False: Beam-on time must be documented
True
Higher occupational dose on the ___ side
Tube
True or False: Never stand in the primary beam
True
Which side should you stand on?
Image-Intensifier side
How far away should you stand from the patient (source)?
6" at 90 degrees
What is the bucky slot shield lead equivalent?
0.25 mm lead equivalent
What is the benefit of the bucky slot shield?
Protects occupational workers at gonadal level
What is the lead equivalent for spot film device protective curtains to protect from scatter?
0.25 mm lead equivalent
Patient position plays a part in ___ rate
Exposure
The image intensifier is the ___ ___ & is considered a ___ barrier
Image receptor; Primary
Bringing the II closer to the patient & raising the table will ___ dose
Reduce
The II is a ___ barrier & must be ___mm lead equivalent
Primary; 2
How do you create decreased SSD?
Place the tube under the table closer to the patient
Decreased SSD = ___ patient ESE
Increased
How do you create increased OID?
The patient table is not raised closer to the II
Increased OID = ___ magnification & patient ESE
Increased
What are the negatives of having the patient table lower closer to the tube?
1. High exposure (increased patient dose)
2. Blurred image (beam divergence reduces spatial resolution)
3. More scatter reaching the IR (reduces contrast resolution)
What is a dead man control switch?
Only continuous pressure applied will allow radiation emission
What does the Eye pedal indicate?
Regular fluoroscopy
What does the Camera pedal indicate?
Cine or High-Level Control (HLC)
What are the regular fluoroscopy ESE guidelines?
1. 10 R/min (88 mGya/min) with 12" SSD
2. If HLC is available, the regular rate is 5 R/min & unlimited HLC
What are the HLC rate (cine) guidelines?
1. 20 R/min (176 mGya/min)
2. Must have continuous audible signal for HLC
True or False: Chart in EMR is ESE reaches 1-2Gy to track reactions
True
True or False: The FDA is trying to limit serious cutaneous skin reactions with long fluoroscopy times and use of HLC
True
In digital, there is constant image contrast ___ of exposure
Regardless

Why is technique much more flexible with digital imaging?
Contrast resolution is CONTRAST

What controls the contrast resolution in digital imaging?
Computer linear response

What manipulates the gray scale in digital imaging?
Window width
In digital, kVp still controls the ___ of the part & ___ production
Penetration; Scatter
Why is kVp less important now in digital?
Digital computer windowing
What is digital very sensitive to?
Scatter
Collimation is ___ in digital
Imperative
What does collimation control in digital?
Controls the production of scatter as well as patient dose
Grids are ___ in digital
Imperative
What do grids control in digital?
Control scatter reaching the IR, but increase patient dose
DR receptors are more sensitive than CR PSP plates so less ___ can be used
Dose
True or False: In general, there are fewer repeat images in digital due to the technique factors
True
What does Linear Response ensure?
Constant image contrast
Wide exposure latitude = more ___ technique
Flexible
What is Windowing?
Computer manipulation of grayscale
True or False: In digital, you can overexpose or underexpose and still get a good image
True
What is Exposure/Dose Creep?
In digital, it is easy to increase patient radiation dose bu using more technique than necessary
What is the Exposure Indicator?
1. Midpoint in histogram VOI (volume of interest)
2. Different ranges for different manufacturers

What is the tube housing made of?
Lead

What is the function of the tube housing?
Protects from off-focus & leakage radiation

What is off-focus and leakage radiation considered?
Secondary, nondiagnostic

How are x-rays emitted?
Isotropically

The tube housing ___ the primary beam
Restricts

What is the restriction rate for the tube housing?
1 mGya (100 mR) per hour @ 1 meter from the source

True or False: Even at the high voltage & current with continuous operation, 1 mGya (100 mR) per hour @ 1 meter from the source should not be exceeded
True

Beam on time must always be ___
Documented
What are 2 notable characteristics of a control panel?
1. Audible tone when the exposure ends
2. Positive indication when the tube is energixed (x-ray on)
What is the protective barrier for the control panel?
1. Lead impregnated glass
2. Still allows for patient visualization
The table has ___ thickness
Uniform
What is the table made of?
1. Carbon fiber (strong)
2. Radiolucent (minimal absorption to reduce patient dose)
What is APR on the control panel?
Anatomically programmed radiography

What is Manual mAs Mode on the control panel?
1. Setting mAs manually
2. No AEC

When needs to be chosen when selecting technique on the control panel?
1. Table bucky
2. Wall bucky
3. Tomography
4. Fluoroscopy

The control panel indicates when the grid is ___
Activated

What is HS on the control panel?
1. High speed (more sensitivity)
2. Used for motion control

What is the HU% on a control panel?
Heat units

What is Load 100% on a control panel?
Heat load to target

What can be changed when setting Manual mAs Mode?
1. kVp
2. mA
3. mAs
4. sec

The control panel has ___ configuration
AEC
What can be selected for the screen on the control panel?
1. High speed
2. Medium
3. Detail