most simple way to measure growth
turbidity
useful countable plate has how many colonies
30-300
in turbidity, light scatter is proportional to
mass
what does a spectrophotometer measure
fraction of incident light measured through culture
OD =
Absorbance
Absorbance =
-logT (transmittance)
Transmittance (T) =
I/I o incident light/intensity after passing through material
transmittance and absorbance are ____. if all light passes through, it’s ____ A and ___ T. If all light is absorbed, it’s ___ A and ___ T
inverse. 0, 100%. infinate, 0.
OD is typically the y axis compared to ___ on the x axis
cells/ml
absorbance in terms of a b c
A = absorbance (OD) x b path length x concentration
beer lambert law is used to determine
concentration
beer lambert law shows a _____ relationship between absorbance and concentration. the ______ line can give concentration of unknown culture
linear. regression
units of a b c for beer lambert
coefficient ml/CFU*cm, cm, CFU/ml
CFU/mL equation
number of colonies/ (dilution (10^x)) x (volume plated mL)
A (absorbance) can also mean
OD600
in cell counting, tryphan blue dyes what cells
dead
for electronic cell counting, cells are put through a ____ and electrical conductivity ______. this causes a ____ ___
pore. decrease. voltage pulse
most direct way to measure growth
dry weight and protein
electronic counting finds
cell number and size distribution
flow cytometry finds
cell-associated flourescence
limitation of viable cell counting
there can be clumps and poor plating can lead to underestimation of viability
dry weight/protein finds ___. limitation is
biomass. slower and destructive
CFU/mL on a plate estimates
cells’ ability to form colonies
limitation of CFU/mL counting colonies
some can be clumped together, so it could be an underestimate
what should you blank the spectrometer with
uninoculated medium
when interpreting OD and CFU/mL together, a steady OD does not mean
viability
two A equations
-log10T and abc