Required equipment for speech audiometry

  1. Speech audiometer: audiometer with a circuit that allows use of a speech stimulus

  2. Sound booth: threshold testing MUST be done in a sound booth

    • Microphone in booth to detect response from patient


2
New cards

Ways to present stimuli

  1. Recorded (CD or wave files)Pros

  2. Monitored live voice (MLV): Tester presents words into a microphone AND monitors the level on the VU meter


3
New cards

Pros and cons of recorded stimuli

  • PROS: control over level, no inter-tester variance or intra-tester variance

  • CONS: harder to control presentation rate

    • we usually prefer this presentation


4
New cards

Pros and cons of MLV stimuli

  • PROS: easier to control presentation rate

  • CONS: harder to control the level, inter-tester variance, intra-tester variance


5
New cards

Patient response options

  1. Raise hand, press button

  2. Repeat word heard

  3. Point to a picture of the word heard

  4. Write down or type the word heard (not often used clinically, often in research)


6
New cards

Threshold procedures

  • speech reception threshold (SRT)

  • speech detection threshold (SDT)


7
New cards

Supra-threshold tests

  • Speech recognition tests


8
New cards

SRT

the lowest level in DB HL which speech can be barely understood (~50% correct)

  • Stimulus: spondee words: bi-syllabic words with equal stress on each syllable


9
New cards

SRT procedure

  1. One ear at a time

  2. 1 spondee at 30dB HL

  3. If no response, present at 50dB, keep going by 10dB HL till response

  4. When response occurs, drop in 10dB steps

  5. Follow “down 10 up 5” procedure used in pure tone testing until 3 correct responses at one level

  6. Threshold = lowest level at which at least 50% of the responses are correct, minimum of at least 3 correct responses at that intensity


10
New cards

Pure tone average

threshold at 500, 1000 and 2000 Hz averaged

  • SRT should correspond to the PTA within 10 dB 


11
New cards

Cross-check principle

comparing PTA with SRT

  • way of confirming your audiogram is correct

  • can be used in counseling regarding degree of hearing loss


12
New cards

SRT Familiarization

Presenting the test stimuli at suprathreshold level or in written form

  • Take out any words the listener does not know

    • Often not done by many audiologists


13
New cards

SDT

lowest level in dB HL that a listener can just detect the presence of speech

  • Does NOT imply understanding!

  • Used for listeners who can’t repeat words back


14
New cards

Stimuli for SDT

sentences presented rapidly and monotonously

15
New cards

Comparing SRT and SDT

SRT is ALWAYS higher than SDT and SAT (a little worse)

  • It’s harder to tell what word is repeated back than just knowing a word is there


16
New cards

Comparing SRT and PTA

SRT is within 10dB to PTA

  • SRT can be much better than PTA, especially when sloping HL is occurs