cassette-based systems
the cassette is simply a light-proof container that protects an imaging plate from light and handling (damage)
computed radiography (CR)
a type of indirect capture digital radiographic system
detector + reader
cassette
durable, lightweight plastic
backed by aluminum or lead to absorb backscatter x-ray photons
antistatic material (usually felt) protects against static electricity build-up, dust collection, and mechanical damage
imaging plate
records the image on a thin sheet of plastic known as the imaging plate
photostimuable imaging plates (PSP)
constructed of several layers: protective layer, phosphor layer, light reflective layer, conductive layer, support layer, light shielding layer, backing layer, and barcode label
imaging plate construction
protective layer:
thin, tough, clear plastic protects the phosphor layer
phosphor layer or active layer (most important layer)
contains photostimuable phosphor that “traps” electrons during exposure
most common — barium fluorohalide family
may also contain a dye
barcode label or sticker
for each new examination, the patient-identifying barcode and the barcode label on the cassette must be scanned and connected to the patient position of examination menu
colored mark or sticker — indicates the appropriate orientation of the cassette in relation to the patient
orientation affects image display
acquiring and forming the image
the patient is radiographed exactly the same way as in conventional radiography
differences lie in how the exposure is recorded
In PSP, the remnant beam interacts with electrons in the barium fluorohalide crystals contained within the imaging plate
interactions gives energy to electrons in the crystals, trapping them in an area of the crystal known as the color of phosphor center
this trapped signal remains for hours or even days and is never completely lost
latent image
deterioration begins almost immediately
small amount of exposure remains trapped; imaging plates are never completely erased
residual trapped electrons are so few that they do not interfere with subsequent exposure
build-up is problematic
remnant beam
the remaining x-ray beam that completely passes through the patient’s body and hits the image receptor
the reader
two types of PSP readers:
point scan
line scan
after exposure, the cassette is fed into a reader that removes the imaging plate and scans it with a laser to release the stores electrons
there are two scan directions within:
fast scan direction
slow scan direction
point scan readers
have an optical stage, a scanning laser beam, translation mechanics, light pick-up guides, a photomultiplier, a signal transformer/amplifier, and an analog-to-digital converter (ADC)
at any point in time, only a single laser point radiates the image plate
line scan readers
based on simultaneous stimulation of the imaging plate one line at a time
acquisition of the photostimulated luminescence (PSL) occurs with a charge-coupled device (CCD) linear array photodetector
the scanning module contains several linear laser units and optical light collection lenses
the system requires a lens array to focus each laser beam to a corresponding point on the CCD array
photostimulated luminescence (PSL)
refers to the emission of light from the phosphor layer after stimulation by the relevant light source
fast scan direction
the movement of the laser across the imaging plate
also known as “scan”
slow scan direction
the movement of the imaging plate through the reader
also known as “translation or subscan direction”
the laser
a laser creates and amplifies a narrow, intense beam of coherent light
light amplification by stimulated emission of radiation
requires a constant power source to prevent output fluctuations
passes through beam-shaping optics to an optical mirror, which directs the laser beam to the surface of the imaging plate
using the laser to read the imaging plate
the cassette is put into the reader, and the plate is extracted
the plate is scanned with a helium laser beam or solid-state laser diodes
the reader scans the plate with red light in zigzag or raster, pattern
the laser gives energy to the trapped electrons
the red laser light is emitted at ~2 eV, which is necessary to energize the trapped electrons
the extra energy allows the trapped electrons to escape the active layer, where they emit visible blue light at an energy of 3 eV as they relax into lower energy levels
the imaging plate moves through the reader or a scanning mechanism moves across the stationary imaging plate
the laser scans across the imaging plate multiple times
the scanning process produces lines of light intensity information detected by a photodetector
the photodetector amplifies the light and sends it to an ADC
translation speed of the plate must be coordinated with the scan direction of the laser or the spacing of the scan lines will be affected
a beam deflector moves the laser beam rapidly back and forth across the imaging plate to stimulate the phosphors
mirrors are used to ensure that the beam is positioned consistently
because the type of phosphor material in the imaging plate has an effect on the amount of energy required, the laser and the imaging plate should be designed to work together
the typical throughput is 50 cassettes per hour
some manufacturers claim up to 150 cassettes per hour, but based on average hospital department workflow, 50 cassettes per hour is much more realistic
the light collection optics direct the released phosphor energy to an optical filter and then to the photodetector
analog
represents changing values as continuously variable physical quantities
digitalizing the signal
the assignment of a numeric value to each light photon - discrete values
the scanning process results in the conversions of the light emitted from the storage phosphor into an electrical signal
the electrical signal is sampled and digitalized to represent a specific location within the image matrix an displays as a specific brightness
the typical number of pixels in a matrix range from about 512 Ă— 512 to 1024 Ă— 1024 and can be as large as 2500 Ă— 2500
the image is digitalized by positioning (spatial location) and by intensity (gray level)
matrix
a group of squares that make up the image formation
pixel
also called picture element; each square of the matrix
the more __ there are, the greater the image resolution for a fixed field of view
erasing the image
the image reading process returns most of the electrons to a lower energy state, effectively removing the image from the plate
imaging plates are extremely sensitive to scatter radiation
plates should be run at least once a week under an erase cycle to remove built-up background radiation and scatter
erasure mode
systems erase the plate by flooding it with light to remove any trapped electrons left after the initial plate reading
cassettes should be erased before use if the last erasure time is unknown
erasure mode
allows the plate surface to be scanned without recoding the generated signal
preprocessing, processing, and forwarding the image
after the imaging plate has been read, the signal is sent to the computer, where it is preprocessed
the data then goes to a monitor that displays the image
the technologist can do the following:
review the image
manipulate it if necessary (postprocessing)
send it to the quality control station and ultimately to the picture archiving and communications systems (PACS)
all systems use a region of interest (ROI) to define the area where the part to be examined is recognized and the exposure outside the region of interest is subtracted
each vendor has a specific tool for different situations, such as neck, breasts, hips, or pediatrics, in which the anatomy requires some special recognition
part selection
selecting the proper body part and position is important for the proper conversion to take place
if the improper part or position is selected, the image may be processed incorrectly and fail to display properly
it is not acceptable to select a body part of position different from that actually being performed to improve the quality of the displayed image
if the proper exam/part selection results in a suboptimal image, service personnel should be notified to correct this issue
improper menu selections may lead to overexposure of the patient and to repeated exams
technical factors
kVp, mAs, and distance are chosen in exactly the same manner as for F/S radiography
kVp was selected for penetration and tissue type
mAs was selected according to the number of photons requires for the body part to be imaged
attenuation of the x-ray beam is exactly the same as in F/S radiography
the proper balance between patient dose and part penetration is vital
kilovoltage peak selection
digital image contrast is no longer dependent on kVp
higher kVp values can be used, allowing for lower mAs values
the dynamic range of digital receptors and exposure latitude are much higher/wider/larger than F/S and inherently produces a wide variety of gray values
contrast is determined by computer processing
equipment selection
imaging plate selection
two important factors should be considered when selecting the PSP imaging cassette:
type
size
grid selection
because PSP systems are more sensitive to low levels of radiation, the use of grids is much more critical than in F/S radiography
grid selection factors include frequency, ratio, and focus
more lead, higher patient dose
imaging plate selection
typically, two types of imaging plates:
standard
high resolution
cassettes should be marked on the outside to indicate high-resolution imaging plates
contain a thinner phosphor layer
the thinner layer reduces light spread yielding greater image sharpness
are limited to smaller cassette sizes and are most often used for extremities and mammography
PSP digital images are displayed in a matrix of pixels, and the pixel size is an important factor in determining the resolution of the displayed image
appropriate image plate selection eliminates scatter outside the initial collimation by reducing the amount of imaging plate not being exposed
this increases contrast resolution
grid selection
grid lines that are projected onto the imaging plate when using a stationary grid can interfere with the image
this results in a Moire pattern:
a wavy artifact caused by grid lines running parallel to the laser scanning motion
grid feeding direction is important
missing data from digital system is another reason to cause this artifact
side/position markers
conventional lead markers should be used the same way they were used in F/S systems
electronic markers can be easily used to mark images with left and right side markers or other position or text markers after the exposure has been made
marking the patient exam at the time of exposure identifies the patient’s side and the technologist performing the exam
legal issue
artifacts
defined as any undesirable objects on the processed image other than those caused by scatter and fog
examples:
adhesive tape used to secure lead markers to the cassette can leave residue on the imaging plate
if static exists because of low humidity, hair can cling to the imaging plate
backscatter — dark lines
worm, cracked areas of the cassette lead coating allow scatter to image these weak areas
horizontal white lines — caused by direct on the light guide in the reader
service needs to clean the light guide
loading multiple imaging plates in a single cassette — only 1 plate usually extracted, leaving the other to be exposed multiple times
insufficient erasure settings — residual image left in PSP plate before the next exposure
also occurs if the erasure lamp needs repair
results vary depending on how much residual image is left and where it is located
orientation of a stationary grid relative to the direction of the laser scanner is critical to reduce the likelihood of the Moire pattern error
the grid lines must be perpendicular to the laser scan direction
moving grids generally do not demonstrate the moire effect because the grid lines are blurred out
vendors have identified specific grid frequencies for their equipment to prevent the artifact
proper collimation and regular cassette inspection help to eliminate these problems
operator errors
underexposure procedures quantum mottle
overexposure reduces subject contrast
proper technical factors are critical for:
patient dose
image quality