General methods of data acquisition

localizer, axial and helical

2
New cards

Localizer scan Dose

similar to xray

3
New cards

Scouts take longer in _______ than _____

rad therapy; radiography

4
New cards

Localizer scans are ______ _______ images

low quality

5
New cards

T/F: All CT routines require at least two scouts

false

6
New cards

Purpose of Localizer Scan

Determine start and end of scan

7
New cards

Localizer Scan Mechanism

table moves through while tube is stationary

8
New cards

In a localizer scan, the __________ determines the orientation of the scan

position of the tube

9
New cards

Selecting the wrong patient orientation results in ___________-

incorrect landmarking on subsequent images

10
New cards

Consequences of incorrect localization

missed anatomy, decreased image quality, incorrect landmarking

11
New cards

There is a _____% dose variation with ____CM off isocentre

20; 4

12
New cards

Other names for Axial Scanning

conventional, serial, sequence, step and shoot

13
New cards

Axial Scanning Mechanism

table moves to a stationary position while the tube rotates

14
New cards

Step and shoot scanning results in images that are _________ to the z axis and ________ to every other slice

perpendicular; parallel

15
New cards

Modern use of axial scanning

dynamic and interventional CT

16
New cards

When would axial scanning be chosen over helical?

scan duration is not a major concern compared to resolution, gapped slices, exposure will be interupted

17
New cards

dynamic CT

re-scanning a specific area repeatedly often with contrast studies

18
New cards

advantages of axial scanning

highest image quality, contiguous or non contiguous, can repeat scans at same location

19
New cards

Dynamic CT is also called ________

cine

20
New cards

Gap image is taken when….

quick survey is needed

21
New cards

Disadvantages of axial scanning

interscan delay adds to total exam time, slice misregistration

22
New cards

slice misregistration

patient breathes differently with each scan

23
New cards

Interscan delay is particularly problematic for ________

CT angiography

24
New cards

SDCT slice thickness

limited by detector width

25
New cards

SDCT produces a _______ beam

fan

26
New cards

MDCT produces a ______ beam

cone

27
New cards

MDCT may contain _______ parallel rows

4-320

28
New cards

MDCT provides _____ and ______ z-axis coverage per gantry rotation

longer; faster

29
New cards

slice thickness in MDCT is determined by a combination of _______ and ______

beam width; detector configuration

30
New cards

How can slice thickness change in MDCT?

multiple rows of detector elements can be combined in various ways to yield different slice thicknesses

31
New cards

Uniform array

all detectors are the same dimension

32
New cards

non-uniform array

detectors of varying widths with thinner detectors in the middle

33
New cards

Non-uniform Array also known as ________

adaptive, hybrid, variable

34
New cards

Changing slice thickness changes ________ ________

voxel size

35
New cards

Slice thickness in SDCT is determined by…

collimation

36
New cards

What happens when collimation is opened beyond detector width in SDCT?

increased dose and scatter radiation

37
New cards

Effects of thicker slices

increased SNR and contrast resolution, decreased spatial resolution, less info for MPR

38
New cards

effects of thin slices

increased spatial resolution and decreased contrast resolution

39
New cards

Slice thickness

width of volume of tissue being examined

40
New cards

slice interval

distance between scan sections

41
New cards

Axial interval =

number of images/rotation x slice thickness

42
New cards

Helical scanning is also called…

spiral or volumetric

43
New cards

Characteristics of Helical Scanning

continually rotating x-ray tube, constant x-ray output, uninterrupted table movement

44
New cards

large volume of data can be scanned in a _________ in helical CT

single breath hold

45
New cards

Helical CT eliminates ________ _________

interscan delay

46
New cards

Advantages of helical scanning

optimization of contrast agent administration, reduces respiratory mis-registration and motion artifacts from organs

47
New cards

Major improvements leading to the development of helical scanning:

slip rings, more efficient tube cooling, higher x-ray output, smooth table movement, software adjusting for table motion, increased data storage, more efficient detectors

48
New cards

increased interpolation ______ image quality

decreases

49
New cards

Reconstruction of helical scan data depends on:

algorithm and pitch

50
New cards

Helical interpolation purpose

remove slant and blur from helical image

51
New cards

Interpolation is associated with ______

some loss of image resolution

52
New cards

pitch

describes CT table movement during a helical scan acquisition

53
New cards

When does pitch of 1 occur?

table feed and beam collimation are identical

54
New cards

Pitch less than 1 occurs when…

table feed is less than beam collimation

55
New cards

pitch greater than 1 occurs when…

scan overlap occurs

56
New cards

Pitch in SDCT

table speed varies according to slice thickness and gantry rotation speed

57
New cards

contiguous slices

slices follow one another

58
New cards

non-contiguous

overlap or gap in slices

59
New cards

Pitch above 1

increased slice angle and thus amount of interpolation required

60
New cards

Increasing pitch results in….

greater scan coverage lengthwise for given total acquisition time and decreased dose

61
New cards

decreasing pitch results in…

slower table movement, less anatomy covered and increased dose

62
New cards

Pitch 2 results in _____ with _____ tilt and ______ quality and dose

gapping, more, decreased

63
New cards

pitch 1 results in _____ slices, ______ tilt and ________ of quality and dose

contiguous, minimal, good balance

64
New cards

pitch 0.75 results in ______ of slices, _______tilt and quality _______ and _____ Dose

overlapping; very little; similar to axial, higher

65
New cards

beam width =

number of slices x slice thickness

66
New cards

Pitch in MDCT =

table movement per rotation/ beam width

67
New cards

T/F: In MDCT, collimation and slice thickness are synonymous

false

68
New cards

MDCT scan coverage =

pitch x total acquisition time x 1/rotation time x beam width

69
New cards

slip rings

electromechanical devices consisting of circular electrical conductive rings and brushes that transmit electrical energy across a rotating surface

70
New cards

slip ring designs

disk and cylindrical

71
New cards

Disk Slip Ring Design

conductive ring forms concentric circles in the plane of rotation

72
New cards

cylindrical slip ring resign

conductive rings positioned along axis of rotation to form a cylinder

73
New cards

Types of slip ring technologies

low and high voltage

74
New cards

low voltage slip ring

slip ring located before HV generator

75
New cards

high voltage slip ring

slip ring located after HV generator

76
New cards

With low voltage slip rings, the HVT….

rotates with the xray tube

77
New cards

Does the HVT rotate with the xray tube with high voltage slip rings?

no

78
New cards

Helical/spiral weighting algorithm

accounts for predictable table movement

79
New cards

general table incrementation speed in helical scanning

10mm/sec

80
New cards

Problems with helical scanning

no defined slice, streak artifacts if projection data used without algorithm, increased effective slice thickness

81
New cards

effective slice thickness is influenced by:

width of fan beam and speed of table incrementation

82
New cards

interpolation is required for _____

de-blurring

83
New cards

Common interpolation algorithms used

180 and 360 linear interpolation algorithms

84
New cards

360 linear interpolation algorithm

broadens slice sensitivity profile up to 20% decreasing image quality

85
New cards

180 degree linear interpolation algorithm

improves image quality by sampling using closer points to improve slice sensitivity profile

86
New cards

what is needed to reconstruct images?

full set of attenuation profiles from every angle for a specific location on the z axis

87
New cards

Axial scanning gives a _______ of attenuation profiles

complete set of 360

88
New cards

T/F: helical scanning doesn’t give a complete set of attenuation profiles thus interpolation is necessary

true

89
New cards

360 degree interpolation algorithm reconstructs image to represent….

a thicker region of the patient than the collimated thickness

90
New cards

180 degree interpolation algorithm reconstructs the image to represent….

a thinner region of patient more true to collimated thickness

91
New cards

pitch was initially referred to as ______

pitch ratio

92
New cards

pitch in SDCT =

table movement per rotation/width of collimation

93
New cards

lower pitch results in better _________ because slices are more true to the collimation

longitudinal resolution

94
New cards

image quality is best in….

helical mode with lower pitch

95
New cards

0.5 pitch has _____ dose

2x

96
New cards

pitch of one has a dose of ____

1

97
New cards

pitch of 1.5 has a dose of _____

x0.7

98
New cards

pitch of 2 has a dose of _____

x0.5

99
New cards

image quality

measure of accuracy of depiction of the original object

100
New cards

image quality parameters

spatial, temoral and contrast resolution