General methods of data acquisition
localizer, axial and helical
Localizer scan Dose
similar to xray
Scouts take longer in _______ than _____
rad therapy; radiography
Localizer scans are ______ _______ images
low quality
T/F: All CT routines require at least two scouts
false
Purpose of Localizer Scan
Determine start and end of scan
Localizer Scan Mechanism
table moves through while tube is stationary
In a localizer scan, the __________ determines the orientation of the scan
position of the tube
Selecting the wrong patient orientation results in ___________-
incorrect landmarking on subsequent images
Consequences of incorrect localization
missed anatomy, decreased image quality, incorrect landmarking
There is a _____% dose variation with ____CM off isocentre
20; 4
Other names for Axial Scanning
conventional, serial, sequence, step and shoot
Axial Scanning Mechanism
table moves to a stationary position while the tube rotates
Step and shoot scanning results in images that are _________ to the z axis and ________ to every other slice
perpendicular; parallel
Modern use of axial scanning
dynamic and interventional CT
When would axial scanning be chosen over helical?
scan duration is not a major concern compared to resolution, gapped slices, exposure will be interupted
dynamic CT
re-scanning a specific area repeatedly often with contrast studies
advantages of axial scanning
highest image quality, contiguous or non contiguous, can repeat scans at same location
Dynamic CT is also called ________
cine
Gap image is taken when….
quick survey is needed
Disadvantages of axial scanning
interscan delay adds to total exam time, slice misregistration
slice misregistration
patient breathes differently with each scan
Interscan delay is particularly problematic for ________
CT angiography
SDCT slice thickness
limited by detector width
SDCT produces a _______ beam
fan
MDCT produces a ______ beam
cone
MDCT may contain _______ parallel rows
4-320
MDCT provides _____ and ______ z-axis coverage per gantry rotation
longer; faster
slice thickness in MDCT is determined by a combination of _______ and ______
beam width; detector configuration
How can slice thickness change in MDCT?
multiple rows of detector elements can be combined in various ways to yield different slice thicknesses
Uniform array
all detectors are the same dimension
non-uniform array
detectors of varying widths with thinner detectors in the middle
Non-uniform Array also known as ________
adaptive, hybrid, variable
Changing slice thickness changes ________ ________
voxel size
Slice thickness in SDCT is determined by…
collimation
What happens when collimation is opened beyond detector width in SDCT?
increased dose and scatter radiation
Effects of thicker slices
increased SNR and contrast resolution, decreased spatial resolution, less info for MPR
effects of thin slices
increased spatial resolution and decreased contrast resolution
Slice thickness
width of volume of tissue being examined
slice interval
distance between scan sections
Axial interval =
number of images/rotation x slice thickness
Helical scanning is also called…
spiral or volumetric
Characteristics of Helical Scanning
continually rotating x-ray tube, constant x-ray output, uninterrupted table movement
large volume of data can be scanned in a _________ in helical CT
single breath hold
Helical CT eliminates ________ _________
interscan delay
Advantages of helical scanning
optimization of contrast agent administration, reduces respiratory mis-registration and motion artifacts from organs
Major improvements leading to the development of helical scanning:
slip rings, more efficient tube cooling, higher x-ray output, smooth table movement, software adjusting for table motion, increased data storage, more efficient detectors
increased interpolation ______ image quality
decreases
Reconstruction of helical scan data depends on:
algorithm and pitch
Helical interpolation purpose
remove slant and blur from helical image
Interpolation is associated with ______
some loss of image resolution
pitch
describes CT table movement during a helical scan acquisition
When does pitch of 1 occur?
table feed and beam collimation are identical
Pitch less than 1 occurs when…
table feed is less than beam collimation
pitch greater than 1 occurs when…
scan overlap occurs
Pitch in SDCT
table speed varies according to slice thickness and gantry rotation speed
contiguous slices
slices follow one another
non-contiguous
overlap or gap in slices
Pitch above 1
increased slice angle and thus amount of interpolation required
Increasing pitch results in….
greater scan coverage lengthwise for given total acquisition time and decreased dose
decreasing pitch results in…
slower table movement, less anatomy covered and increased dose
Pitch 2 results in _____ with _____ tilt and ______ quality and dose
gapping, more, decreased
pitch 1 results in _____ slices, ______ tilt and ________ of quality and dose
contiguous, minimal, good balance
pitch 0.75 results in ______ of slices, _______tilt and quality _______ and _____ Dose
overlapping; very little; similar to axial, higher
beam width =
number of slices x slice thickness
Pitch in MDCT =
table movement per rotation/ beam width
T/F: In MDCT, collimation and slice thickness are synonymous
false
MDCT scan coverage =
pitch x total acquisition time x 1/rotation time x beam width
slip rings
electromechanical devices consisting of circular electrical conductive rings and brushes that transmit electrical energy across a rotating surface
slip ring designs
disk and cylindrical
Disk Slip Ring Design
conductive ring forms concentric circles in the plane of rotation
cylindrical slip ring resign
conductive rings positioned along axis of rotation to form a cylinder
Types of slip ring technologies
low and high voltage
low voltage slip ring
slip ring located before HV generator
high voltage slip ring
slip ring located after HV generator
With low voltage slip rings, the HVT….
rotates with the xray tube
Does the HVT rotate with the xray tube with high voltage slip rings?
no
Helical/spiral weighting algorithm
accounts for predictable table movement
general table incrementation speed in helical scanning
10mm/sec
Problems with helical scanning
no defined slice, streak artifacts if projection data used without algorithm, increased effective slice thickness
effective slice thickness is influenced by:
width of fan beam and speed of table incrementation
interpolation is required for _____
de-blurring
Common interpolation algorithms used
180 and 360 linear interpolation algorithms
360 linear interpolation algorithm
broadens slice sensitivity profile up to 20% decreasing image quality
180 degree linear interpolation algorithm
improves image quality by sampling using closer points to improve slice sensitivity profile
what is needed to reconstruct images?
full set of attenuation profiles from every angle for a specific location on the z axis
Axial scanning gives a _______ of attenuation profiles
complete set of 360
T/F: helical scanning doesn’t give a complete set of attenuation profiles thus interpolation is necessary
true
360 degree interpolation algorithm reconstructs image to represent….
a thicker region of the patient than the collimated thickness
180 degree interpolation algorithm reconstructs the image to represent….
a thinner region of patient more true to collimated thickness
pitch was initially referred to as ______
pitch ratio
pitch in SDCT =
table movement per rotation/width of collimation
lower pitch results in better _________ because slices are more true to the collimation
longitudinal resolution
image quality is best in….
helical mode with lower pitch
0.5 pitch has _____ dose
2x
pitch of one has a dose of ____
1
pitch of 1.5 has a dose of _____
x0.7
pitch of 2 has a dose of _____
x0.5
image quality
measure of accuracy of depiction of the original object
image quality parameters
spatial, temoral and contrast resolution