NCRP 102 of aprons MUST be vs SHOULD be
must: .25 mm Pb
should: .5 mmPb
NCRP 102 of gloves MUST be
.25 mm Pb
primary barrier (for primary beam) lead equivalent number
1.6 mm OR 1/16 inch
secondary barrier (for scatter) lead equivalent number
.8 mm OR 1/32 inch
thyroid shielding lead equivalent
.5 mm Pb
glasses shielding lead equivalent
.35 mm Pb
gloves/drapes/bucky slot shielding lead equivalent
.25 mm Pb
control booth shielding lead equivalent
.8 mm Pb
control booth WINDOW shielding lead equivalent
1.6 mm Pb
inverse square law formula
new distance = (old intensity)(old distance/new distance)²
NCRP annual occupational exposure
50 mSv (.05 sieverts)
NCRP 116 occupational exposure to lens, skin, hands, feet
lens: 150 mSv
skin/hands/feet: 500 mSv
public exposure & students effective dose: continuous, infrequent, lens, skin/hands
continuous: 1 mSv
infrequent: 5 mSv
lens: 15 mSv
skin/hands: 50 mSv
NCRP 116 fetus: monthly, entire gestation
monthly: 0.5 mSv
entire gestation: 5 mSv