X-ray imaging is a _____ imaging of a 3D body where the image represents the total _____ of the x-ray beam as it passes through the patient
Transmission
Attenuation
The word tomography is derived from Ancient Greek tomos which means
Slice, section
CAT/CAT
Computed axial tomography
CT produces radiological images as ______ sections
Transaxial
Cross sectional images
Contrast in an x-ray is ___ %, while it is ____% to _____% for a typical CT image
2%
0.1%
0.3%
The mathematical principles of CT were first developed by Johann ____ in _____.
Radon
1917
The first commercially viable CT scanner was invented by Sir Godfrey _____ in ______.
Hounsfield
1967
A single transmission measurement through the patient made by a single detector at a given moment in time is called a
Ray
A series of rays that pass through the patient at the same orientation is called a _____ or a ______.
Projection
View
____ is the center of rotation of the CT gantry
Iso-center
The maximum ____ is defined by the physical extent of the curved detector arrays (the fan angle)
Field of view
____at the iso-center is defined as the ratio of the source to detector distance to that of source to iso-center
Magnification
The ____ is related to the detector width
Slice thickness
____and _____ make the use of tube output much more effective
Multi-detector arrays and opening up the collimator
The x-ray ___ or must not vary over the image acquisition cycle, as any variation would be treated as absorption differences
Energy must also be ___, because absorption coefficients are dependant on the x-ray energy
Beam stability should be better than ____%
Intensity
Stable
1%
The sum of the individual voxel data is called the ______.
ray-sum
The image is formed by calculating the individual voxel values within the matrix using ______ and ______ techniques
Image processing and reconstruction techniques
What is a pixel?
The 2D element of a digital image
Each pixel corresponds to a voxel in the patient
The voxel has two dimensions equal to the pixel and one extra dimension which is the slice thickness
Image reconstruction is used via
Iterative techniques (guess-compare cycle)
Back projection (reverse of projection)
Derivatives of back projection
High resolution CT requires ____ number of projections
Insufficient number of projections causes ____ in the reconstructed image
High
Streaks
The detector collects information about the total ____between the source and the detector
Attenuation
Spacial resolution includes?
In plane resolution
Slice thickness
____ are a type of chemical that produce visible light when interacting with high energy photons
Scintillators
_____ detectors are made of inert gas chambers that are sensitive to high energy photons and produce an electric signal upon interaction with the photons
Gas detectors
What are the 3 types of detector systems that are suitable for CT machines?
Which one is favourable in spite of the lower efficiency in detecting photons?
1. Multiple scintillation detectors with photomultiplier tubes
2. Multiple scintillation detectors with photodiodes
3. A single multi-chamber inert gas detector
Gas detectors
Scintillation phosphors produce a brief pulse of light for each photon detected
The light pulse is converted into an _____ by a photomultiplier tube
This can multiply the signal by 10^6 or more
Electric signal
Only a small amount of energy is required to ionize _____.
Ion production is linearly proportional to _____.
Gas detectors rely on ________ and the collection of the induced charge by electrodes. A high voltage is applied to collect the electrons.
Why is amplification necessary?
Gas
Energy absorbed
ionization of gas atoms
Because the current is small
What is a Hounsfield Unit?
A unit of measurement for CT images
Allows comparison of images from different machines
Only tissues with ______ or _____ CT values can be seen on the image
0 or positive
What determines slice thickness?
What is the sensitivity profile?
Collimation of the beam
Represents one measure of "real" slice thickness
The perfect sensitivity profile is ______.
Due to divergence, the finite size of the focal spot on anode, the actual SP is not.
The ____ determines the degree of interferences from the adjacent slices in the image
The ____ the SSP slope, the smaller the overlap, or smaller partial volume effects
Rectangular
Steepness
Steeper
____ is the first x-ray image taken from a patient before the actual CT
Scout view
(Scanned projection radiograph)
A ____ degree image is taken, the table is moved further into the gantry, and another image is taken, and the process is repeted slice by slice
Newer machines have a fast data acqusition (1-2s)
What controls the slice thickness?
Basic sequential acquisition
360
The primary and secondary diaphragms (collimators)
A larger slice dose profile contributes to increased
In sequential scanning, the dose profiles ___ and increase the dose to the patient
Surface dose
Overlap
How is image data obtained?
Machine precision in rotation of the fan beam influences?
A fan beam is rotated around the patient
The resolution of the data
How is the resolution of the image influenced?
If the patient moves
Pauses between slice acquisition (can miss anatomical details)
Drawbacks of sequential acquisition
Increased dose at the skin surface due to slice dose profile overlap
Loss of anatomical information
Inter-scan delay times
What was created to overcome the disadvantages of sequential acquisition?
What differs?
What are fixed cables replaced by?
Allows complete anatomical volumes to be scanned in a continuous fashion in 20-60s
Spiral (helical) aquisition
The data is collection with the patient moving through the gantry while the fan beam is continuously rotating
Slip rings (transmission of power and electrical signals from a stationary to a rotating structure)
What allows for fast rotation times? 0.8-3s
Slip rings and brushes
Advantages of continuous data acquisition via spiral
No interscan delay
Allows for a single breath hold
Thinner slices
Pitch of 1 vs pitch of 2 vs pitch less than 1
1: No slice gap
2: gap equal to one x-ray beam width, increases width of slice profile
<1: overlapping radiation beams, increases dose in patient
______is defined as the full-width at half maximum of the slice sensitivity profile
Slice thickness
The higher the pitch, the ____ dose to the patient, ____ table increment, ___ scan times, _____ images
Decreased, increased, decreased, lower resolution
Multislice CT imaging requires a _____ beam instead of a fan beam
Comes in 4, 8, 16, or 64 slice geometries
cone-shape
Patient dose is related to _____.
The higher the pitch, the ____ the imaging time, the ____ the patient's dose, ____ quality images
The lower the pitch, the ____ the imaging time, the _____ the patient's dose, ____ quality images
In multislice CT, higher pitch, more slices are acquired, so images with better quality can be acquired while keeping the patient's dose low
Pitch
Shorter, lower, lower
Longer, higher, higher
What is cardiac imaging used for?
Measurements of coronary artery calcification
____ finds the exact mathematical solution for reconstructing the image slice from the image date
Slow
Rounding errors
All data must be collected before reconstruction can begin
Iterative Technique
Back projection
CT collects X-ray measurements from many angles.
The computer projects the measurements backward along their original paths.
The projections overlap to create an image.
Problem: The image is blurry and can have a starburst effect.
Filtered back projection
CT collects X-ray measurements.
The computer applies a mathematical filter to the measurements.
The filtered measurements are projected backward.
Result: A sharper image with reduced blurring and starburst artifacts.
What does convolution do?
It makes the filtered back projection sharper
What determines/describes the image quality?
Uniformity
Spacial resolution
Contrast
Noise
Image artifacts
What is uniformity?
Determines the accuracy of CT numbers or measurements for the same slice
A measure of the homogeneity over the cross section of a homogenous phantom
____ is carried out by performing a single scan of a water phantom
An uncalibrated system will have influence over image quality including resolution, contrast, and noise
Calibration
____ is the ability to resolve small details
Spatial resolution
____ is the ability to discriminate object from background
Measured as differences between the object density in CT numbers and the background (water)
Influenced by image reconstruction process
Involves image noise and spacial info
Contrast
____ is the standard deviation of CT numbers in a uniform image
Indicates how many photons have reached the detector to form the image
Noise gives the grainy appearance to the image
_____ slices are noisier
If slices are too thin _____
Noise
Thin
Soft tissue contrast is lost
Beam hardening
Influenced by patient thickness and tissue material (worse for bone)
As lower energy photons are removed when the beam enters the patient, the energy of the x-ray increases, the values of the linear attenuation coefficients decrease
Non-uniform response in detecting x-rays results in ____ in the image
Artifacts (visual flaws in the image)
__D volumes can be acquired via spiral CT
3
Fluoroscopy
Visualization of body parts in motion by projecting x-ray images on a luminous fluorescent screen
Multi slice advantages
These factors improve the spatial, temporal, and contrast resolution of the images, significantly increasing the diagnostic accuracy
Longer anatomical coverage
Longer z-axis coverage
Shorter acquisition time with fewer motion artifacts
Creation of thinner or thicker slices from the same data
Improved 3D rendering with less spiral artifacts
Increased volume coverage
Contrast can be delivered faster
The conventional single slice CT scanner has one x-ray tube and ____ row of detectors
-500-900 detector elements
-Describe an arc in the transverse (XY) plane
-One channel of spacial data
The multislice helical CT scanner has one x-ray tube directed at ____ rows of detectors
-35000 detector elements
-2D curved array in the transverse plane
-Multiple channels (4, 8, 16, 64)
A single
Multiple
Factors to consider for optimized image quality
Collimation (narrow)
Rotation time
Pitch value
Isotropic resolution
Achieved using small ___, and thin ____.
Identical X, Y, Z voxel dimensions
Focal spot
Slice collimation
How does a bone beam differ from a fan beam?
The relation between the source and individual detectors is variable
Detector elements in the periphery are exposed more obliquely to those of the center
Multi slice CT
Thicker slices are useful for reducing? Decreasing?
Number of images to be viewed
Image noise
Clinical advantages of a multi slice CT
Increased speed in scan
Increased volume coverage
Isotropic voxels
-Accurate 3D reconstruction
-Suitable for angiography and visual endoscopy
In Winnipeg we receive annually ___mSv from natural sources (food, radon gas, uranium, flying)
Public limit of artificial radiation dose is ___mSv per year
4.1
1
The increase in dose to head and body in CT imaging is due to?
Advanced technology
Increase in heat capacity of the anode in the x-ray tube
Factors that influence patient dose
Beam energy
Filtration
Collimation (slice thickness)
Slice number and spaces
-overlapping slices = more dose
Scan pitch factor
-increasing pitch decreases dose
Image quality and noise
The two most significant factors affecting image quality are
Detector collimators help increase contrast while reducing dose to the patient
Noise
Scatter
When dealing with radiation to produce an image, these two things should ALWAYS be considered
1. Image quality
2. Reduce dose to the patient (ALARA)
Roles and responsibilities of Imaging Medical Physicists
"Pharmacist of radiation oncology"
Commissions all new technology and techniques in RT and is in charge of all quality assurance
What is 3D CRT?
Planning based on 3D CT imaging
Allowed for full visualization of treatment region, better plans
What is IGRT/VMAT?
Dose conforms around the tumour
What is the role of CT in treatment planning?
Shows soft tissue
Highly accurate
Cheap
____ can be related to tissue density
Voxel values
Describe the RT process
Immobilization
Image acquisition
Planning
Patient positioning
Treatment
What does 4DCT allow?
___ 3D datasets are reconstructed to represent the entire breathing period
Can be used to visualize the ___ of the tumour
Takes movement into account
Respiratory correlated CT datasets (motion during breathing)
Trajectory
Metal artifacts affect _____ and ____.
Contouring and dose calculation
Dose calculation
___ can be related to electron density
Can also carry out tissue mapping
CT number
What does tomography mean?
Slice imaging
Provides ____ imaging instead of projection
Volumetric
What do radiotherapy physicists do?
Ensure that equipment is operating safely
What do imaging physicists do?
Ensure that the medical imaging obtained are the best that can be achieved
Technical advice
Quality assurance
Research
Teaching
An expert in the physics of how images are formed?
Imaging physicist
Cons of each imaging technique
Does the # of cancers detected justify the potential harm/risk incurred due to radiation dose from screening?
Mammography
Cons of each imaging technique
Image-guided procedures can be complex and long
Dose to skin can be high
Potential for radiation burns, hair loss, eye damage
Fluoroscopy
Cons of each imaging technique
High radiation dose relative to other x-ray procedures
CT
Decreasing image noise requires ___ dose
Increasing
___ means sufficient to make an accurate diagnosis while minimizing patient dose
"Good enough"
Goal of an x-ray imaging physicist
Maintain or improve image quality while reducing image dose
Emission images vs transmission images
E: A radioactive substance is injected into the patient and the energetic photons are generated within the body and detected outside the body
T: Patient is irradiated with x-rays from outside and transmitted x-rays are detected outside the body
Isotope
ex: Carbon 12
11C, 13C, 14C
Same atomic number
Different # of neutrons
Isotones
Ex: 12C6
11B, 13N
Same # of neutrons
Different atomic number
Isobars
ex: 12C6
12B, 12Be, 12N, 12O
Same atomic mass
Different atomic number
Different # of neutrons