Quality assurance:
the routine, periodic evaluation of an ultrasound system to guarantee optimal image quality
How often should QA be performed?
periodically and routinely according to your specific lab
Who is responsible for implementing a quality assurance action plan?
the sonographer
What are the 4 requirements of QA?
Assessment of system components
Repairs
Preventative maintenance
Record keeping
Are technique and protocols a part of QA?
No
What are the 4 QA devices?
AIUM 100 mm test object (first ever device; no longer used)
Tissue equivalent phantom
Doppler phantom
Beam profile/slice thickness phantom
Methods for QA:
Test under known conditions
Use constant instrument settings
Use a phantom with measurable characteristics
Image in an identical environment
An objective statement is…
completely unbiased
A subjective statement is…
one that is influenced by an individual’s experience, beliefs, or assumption
QA programs should be based on _____ standards
objective
Tissue equivalent phantom is used to…
evaluate characteristics such as gray scale, tissue texture, multi focus & adjustable focus phased array transducers
Tissue equivalent phantoms are similar to soft tissue in the 4 following ways:
Speed of sound
Attenuation
Scattering characteristics
Echogenecity
Gray scale is evaluated because it contains _____ that act like soft tissue
small scatterers
Structures embedded in the phantom mimic what?
hollow cysts and solid masses
What may be embedded in the phantom to produce reflections at strategic locations
nylon strings
Evaluated when the pins in the test object are parallel to the sound beam’s main axis
axial resolution
Evaluated when the pins in the test object are perpendicular to the sound beam
lateral resolution
Evaluated by comparing the distances between reflections on the display with the actual distances in the test object
electronic caliper accuracy
Evaluated by scanning the pins located at the top of the test object very close to the transducer
dead zone
Doppler/flow phantom is used for…
evaluating Doppler systems
Includes a circulation pump to propel fluid through vessels embedded in a tissue equivalent phantom
Doppler phantom
Other Doppler phantoms contain…
vibrating strings or moving belts
Doppler modalities:
Pulsed
Continuous wave
Color
Power mode
Slice thickness phantom evaluates…
elevational resolution
Slice thickness phantom:
assesses the slice thickness and its effect upon image accuracy
Slice thickness is most likely to…
degrade image quality because the imaging plane is thicker than either the beam width or the pulse length
When the ultrasound beam is overly thick…
cystic structures may appear filled in
Thicker slices diminish…
spatial resolution and reduce the ability to visualize small, low contrast reflectors
Contains a diffuse scattering plane that is at an angle to the incident beam & measures the beam geometry perpendicular to the imaging plance
slice thickness phantom
Sensitivity:
refers to the ability of a system to display low level echoes (weak echoes)
What are the 2 forms of sensitivity?
normal and maximum
Normal sensitivity:
those at which all the pins, solid masses, and cystic structures in the test phantom are accurately displayed
Maximum sensitivity:
evaluated with the output power and amplification of the system set to the maximum practical levels
Maximum sensitivity is used with…
a tissue equivalent phantom
Dead zone:
is the region closest to the transducer to the shallowest depth and results from transducer ringing and the time it takes the system to switch from the transmit to the receive mode
Information within the dead zone is…
unreliable and may not be used for diagnostic setting
Higher frequency transducers have a ____ dead zone than lower frequency transducers
thinner
Increasing dead zone may indicate…
a cracked crystal, detached backing material, or a longer pulse duration
What allows for accurate imaging of important superficial structures?
a stand off pad
Registration accuracy:
the ability of the system to place reflections in proper positions while imaging from different orientations
Range accuracy or Vertical depth calibration:
describes the system’s accuracy in placing reflectors at correct depths located parallel to the sound beam
The accuracy of reflector depth positioning
depth calibration
Horizontal calibration:
the system’s ability to place echoes in their correct position when the reflectors are perpendicular to the sound beam
Focal zone:
surrounds the focus (the depth at which intensity is the highest and the beam is narrowest)
______ resolution is excellent in the focal zone because the beam is narrow
Lateral
The focus of phased array transducers is…
sonographer controlled
Systems with dynamic receive focusing should produce…
narrow reflections over a wide range of depths
Axial resolution:
the smallest distance at which two pins parallel to the sound beam are displayed as two distinct echoes
Axial resolution is evaluated by…
scanning a set of closely spaced pins within the phantom
Lateral resolution:
the minimum distance that two side by side rods are displayed as two distinct images
Compensation operation or Uniformity
describes the system’s ability to display similar reflectors in the phantom with echoes of equal brightness
With proper TGC, identical reflectors should have the same appearance on the monitor regardless of their ____
depth
What phantom is used to evaluate the dimension, texture, and fill in of cysts
tissue equivalent
A tissue equivalent phantom system should accurately display hollow structures as ____ and solid structures as ____
anechoic, hyperechoic
Adjusting the systems output power and amplification should produce changes in the…
grayscale display
The strength of the reflection is related to the characteristics of…
the tissue creating the reflection
Direction of the flow =
color map
Sample volume =
accuracy of sample volume placement
Velocity =
accuracy of velocity estimation
Vertical length =
axial
Width =
lateral
What results in better elevational resolution?
a shorter, thinner beam
What is elevational resolution determined by
the height of the ultrasound beam