Film Reject/Repeat Analysis

is an important aspect of a quality

management program in an imaging department.

2
New cards

To avoid future repeats

Film Reject/Repeat Analysis Goal:

3
New cards

- Radiographic Equipment Performance

- Radiologic Technologist Competence

- Efficiency of Departmental Protocols

Film Reject/Repeat Analysis is a statistical representation of the following key performance indicators:

4
New cards

Rejects

Refers to all rejected images, including repeats.

5
New cards

Repeats

Refers to patient images repeated, resulting in additional exposure

6
New cards

1. Improved department efficiency

2. Lower department costs

3. Lower patient doses

Film Reject/Repeat Analysis, Advantages:

7
New cards

Additional workload, in terms of administrative tasks

Film Reject/Repeat Analysis, Disadvantages:

8
New cards

Repeat Analysis Program

is a systematic process of cataloguing rejected

images and determining the nature of the repeat, so that repeat images

can be minimized or eliminated in the future

9
New cards

- Equipment and accessory performance

- Departmental Procedures

- Skill level of the Technical Staff

Repeat Analysis Program provides the following important data:

10
New cards

American Association of Physicists in Medicine, Task Group 151

suggested that reasons for rejection should be standardized and included in all RAP programs. An option to add additional

user-specified reasons should also be available.

11
New cards

1. Rotation

2. Anatomy cut-off

3. Incorrect projection

4. Incorrect marker

Positioning error:

12
New cards

1. Overexposure

2. Underexposure

Exposure error:

13
New cards

1. Cut-off

2. Decentering

3. No grid

4. Grid lines

Grid error:

14
New cards

1. Detector

2. Foreign body

3. Table/Support/X-ray Tube

Artifact:

15
New cards

Causal repeat rate

is the percentage of repeats from a specific cause, such as positioning error or technique error

16
New cards

Causal repeat rate

... concept can also be utilized to do an in-depth quality and performance of imaging department staff

17
New cards

Determine the percentage of repeats for a specific cause

Main function of Causal Repeat Rate:

18
New cards

- Radiologic Technologist

- RT Intern

- X-ray Room

- Shift (Team)

Causal repeat rate,

Secondary function:

Determine the percentage of repeats for a specific cause:

19
New cards

total repeat rate

The ... is the percentage of repeat films in a departmental aspect.

20
New cards

Total repeat rate

This presents the overall repeat rate of an imaging department

21
New cards

- Equipment Quality

- Technical Staff (RT) Competence

- Patient Population

- Data Collection Method

- Shift

- Number of images accepted by the radiologist to diagnose

Total repeat rate presents the overall repeat rate of an imaging department and is

influenced by the following:

22
New cards

8%

Repeat rates for radiographic procedures should not exceed ... for general

radiography

23
New cards

5%

Repeat rates for radiographic procedures should not exceed ... for pediatric radiography

24
New cards

±2%

Repeat rates for radiographic procedures should not exceed 8% for general

radiography and 5% for pediatric radiography, with corrective action limits

of ... (AAPM TG 151).

25
New cards

±2%

Mammographic repeat rates cannot change by more than ... each quarter (AAPM TG 151)

26
New cards

10% to 12%

Any departments with repeat rates exceeding ... should be examined seriously.

27
New cards

inefficient and high patient dose

Any departments with repeat rate exceeding 10% to 12% should be examined seriously.

Any department exceeding these values are considered ... and contributes to ...

28
New cards

compliance and acceptance of poor-quality images

A department repeat rate that is too low may signal problems with ... and ...

29
New cards

film analysis programs

After establishment of the QAP, ... programs should be created to minimize repeats.

30
New cards

Monthly, QA/QC Committee

... reports of film analysis should be kept on file and reviewed by the designated ...

31
New cards

- Tabulated Data

- Graphs

- Charts

- Trend

Methods to present a Film-Reject Analysis:

32
New cards

- Assess the efficiency of the whole department

- Assess the efficiency of equipment

- Assess the efficiency of technical staff

A Film-Reject Analysis can be used to:

33
New cards

- HR Decisions

- Staff Training

- Equipment replacement/upgrade

A Film-Reject Analysis can also be used to justify the following: