AOI
is a specific part of the body that is being focused on, studied, or imaged during a medical procedure.
Collimator
limits the x-ray beam to the AOI using adjustable shutters.
Shutters
metal plate that controls the size and shape of the x-ray beam by limiting the field of radiation
every 6 months
Light field alignment and scale accuracy test:
annually
visual inspection:
100 cm
Procedure: set the FFD to (blank) cm
Scale settings
Procedure: Compare actual light size to the (blank) on the collimator
Shutter Chain Drive
used to control the movement of the shutter
x-ray engineer
if misalignment or scale inaccuracy is found and cannot be corrected easily, call a
engineer/bio-med technicians
the ones who change the collimator light bulb
quartz bulb
high temperature resistant glass bulb made from fused quartz or silica instead of regular glass.
Halogen bulb
a type of incandescent lamp that contains a small amount of halogen glass (iodin/bromine) along with tungsten filament.
tungsten filament
a material that extends the light of the bulb
2%
the difference between the light field and the x-ray field should not be more than (blank) of the SID in any direction.
once a year or everytime the bulb or mirror is changed/ater service changed/after
Frequency for x-ray beam alignment test: