what are stages of digital processing
pre-processing
post-processing
what happens during pre-processing
data prepared for processing and display
what happens during post-processing
editing of the image through computer operation available before and after image is displayed
what are processes of digital data
histogram creation and analysis
automatic rescaling
look up table (LUT) application
what happens to raw image data before it can be displayed
several operations occur to prepare it so a quality image can be displayed
what type of imperfections is software correction applied to
bad pixels
dead pixels
what type of cassettes does software correction occur with
DR type IRs
what is flat fielding
process of correcting nonuniformity of pixel values throughout image
what is considered the first step in digital imaging processing
histogram creation
what is a histogram
graphic representation of a data set
what does a histogram include
all pixel values that represent the invisible image after software corrections applied
what is another term for a histogram
exposure histogram
what does a x-axis in histogram represent
range of pixel values assigned
what does y-axis in histogram represent
frequency of pixels for each value
what does location of graph along x-axis in histogram represent
image brightness levels
what does shape of histogram graph represent
image contrast (grayscale)
histogram analysis is an important processing technique for
identifying edges of an image
assessing exposure data/pixel value before image processing and display
what factors impact location and shape of histogram
anatomy imaged
exposure technique used
positioning used
collimation
what is histogram data compared to by a computer
stored preestablished histogram specific to the anatomic part imaged
what do stored histogram models have
distinct and characteristic shape based on anatomic region and projection
what is value of interest (VOI)
range of histogram data set that should be included in displayed imaged
in a raw exposure data histogram, what needs to be excluded and why
area outside of collimation to prevent processing error
what is automatic rescaling
process of mapping the gray scale to VOI to present a specific display of brightness
what does automatic rescaling maintain
consistent image brightness despite overexposure or underexposure
what does automatic rescaling not compensate for
underexposure
image noise
what can cause automatic rescaling errors
poor collimation
positioning
image artifacts
what is a LUT
method of altering the image based on reference histogram for image display
what kind of exposure response and dynamic range do digital detectors have
linear exposure response
wide dynamic range
what must be altered in raw images for improve visibility
image contrast
what does LUT alter and how
brightness and grayscale using computer algorithms
what happens to a LUT graph if original image is altered and why
no longer a straight light (S-shaped) due to different original pixel values
why is it important to choose correct LUT
anatomy appearance can be altered
what does calculation of pixel intensities in LUT display
differing contrast levels in an image
LUT creates a final look in terms of
brightness and contrast
what do exposure indicators provide in terms of exposure to IR
numeric value indication level of exposure to digital IR
what is a feature of digital imagining process
creating image with appropriate amount of brightness regardless of exposure to IR
how is numerical expression of exposure to receptor calculated
pixel VOI in histogram
what does exposure indicator assess
image quality in terms of noise
exposure indicator is NOT a measure of
patient exposure
what is symbol for exposure indicator
EI #
S #
varies between vendors
exposure indicator can be impacted by
collimation
positioning
kVp
how are target EI values determined
by department officials based on acceptable image quality standards
when was deviation index established
2008
what is deviation index
value that reflects the difference between the desired or target exposure to the image receptor and actual exposure of IR
DI value and actual exposure to IR
>+3.0 — Excessive exposure by a factor of 2 or 100% more than the EIt.
>+1.0 — Overexposed
<+0.5 — Exposure within target range
0 — No difference between target (desired) and actual exposure to IR
>−0.5 — Exposure within target range
<−1.0 — Underexposed
<−3.0 — Insufficient exposure by ½ or 50% of EIt.
what is dual-energy subtraction
technique that removes superimposed structures to make anatomic area of interest more visible
how is dual-energy subtraction possible digitally
computer subtracts selected brightness values
what is contrast enhancement
technique that alters pixel values to increase image contrast
what is edge enhancement (aka high-pass filtering)
postprocessing technique that improves visibility of small, high-contrast structures
what is a downside of edge ehancement
increased image noise
what is inversion
postprocessing technique that reverses grayscale of original image
what is soothing (aka low-pass filtering)
postprocessing technique that supresses image noise
what is downside of smoothing
decreased spatial resolution
what is equalization
postprocessing function where underexposed areas are made darker and overexposed areas are made lighter
what is benefit of equalization
image with lower contrast for better visibility of dense and lucent structures
what is region of interest in postprocessing
provides calculations of selected pixels values within area of interest to provide quantitative information about tissue
what is stitching in post processing
viewing multiple images as one
what is new technology for postprocessing
grid subtraction
scatter correction
what does new technology for postprocessing improve
image quality
reduced patient exposure
how are digital artifacts classified
according to origin
equipment
procedural
what can digital artifacts be related to
calibration of individual DEL, row, or column of detector matrix
artifacts during data extraction related to electronic readout mechanisms associated with rows and columns of DEL comprising detector matrix
erasure error
improper centering of CR to table grid
quality of image is affected by what features of a monitor
luminance
resolution
ambient viewing light and monitor
what is electronic masking
altering regions viewed on displayed image
what does window level do
set midpoint of range of brightness visible in image
high level brightens image
low level darkens image
what does window width do
varies range of shades of gray visible in image
narrow increases contrast
wide decreases contrast
what is the primary monitor for diagnostic interpretation
high-res 5-megapixel (2049 × 2560 pixels)
bit depth of 8 - 256 gray shades
what is display monitor diagonal measurement requirement
must be 80% of viewing distance
53 cm (21 in) for images sized 35 × 43 cm (14 × 17 in)
what is pixel pitch for standard image viewing
.200 mm at viewing distance of 60 cm (24 in)
what is aspect ratio
relationship between monitor width and height
3:4
4:5
16:9
16:10
what is luminance
measurement of light intensity emitted from surface of monitor
what is unit luminance is expressed in
candela per square meter (cd/m2)
what is luminance range for primary display monitors
1.0 cd/m2 to 350 cd/m2
what is luminance ratio (LR)
ratio of maximum to minimum luminance
what is recommended LR
greater than 250
what does a monitor with a LR of 350 have
improved contrast
what does pixel bit-depth determine
contrast resolution
14 bit-depth yield 16384 shades of gray
how does LR affect contrast
higher ratio displays greater shades of gray
what are additional monitor concerns
geometric distortion
veil glare
display noise
what is AI applied to
digital data sets of medical images
what systems deal with managing patients health information
electronic health record (EHR)
health information systems (HIS)
medical image management and processing system (MIMPS)
what does health information management consist of
health informatics and data management
what does radiology specific health informatics deal with
collection
documentation
maintenance
protection and privacy
what is health information management
a complex system-wide effort to ensure that people, equipment, and processes are in place to expand healthcare communications and improve patient outcomes
modern radiology practice requires efficient and effective methods for
acquisition, storage, retrieval, transmission, and display of digital image from multiple modalities throughout healthcare systems
what do digital communication needs in radiology also include
processing and delivery of patient data
subsequent interpretation of radiological procedures
protected health information (PHI)
patients health related information that’s collected, stored, and shared
hospital information system (HIS)
stores PHI
electronic health record (EHR)
consists of all PHI on patient from all healthcare encounters throughout lifetime
includes electronic medical record (EMR)
radiology information system (RIS)
system specific to radiology
what does clinical informatics involve
data management for radiologic services both within and outside the healthcare system
what is accurate PHI critical to
healthcare revenue streams and financial stability
what guidelines does PHI require proficiency with
ICD-10-CM
current procedural terminology, 4th edition (CPT-4)
what is medical image management and processing system (MIMPS)
sophisticated computer network designed for digital imaging departments and enterprises
formerly PACS
what does MIMPS deal with
acquisition
distribution
storage
display
what kind of information is MIMPS permitted to share
healthcare specific not vendor specific
integrating the healthcare enterprises (IHE)
what are platforms that chare medical image and information across modalities and platforms
health level seven standard (HL7)
DICOM
what will growth of VNA’s create
health information exchanges (HIEs)
what does DICOM header on digital images include
accession number
patient name and identification number
date and time of examination
radiographic procedure
total number of images in study
exposure related information: kVp, mAs, exposure indicator, pre or post processing
what does DICOM worklist do
assist radiographer in maintaining accuracy of information
what is grayscale standard display function (GSDF) for
consistently displaying characteristics such as grayscale appearance and image quality