What is the definition and primary objective of microtomy?
Preparing thin, transparent slices of biological tissue post-embedding for microscopic examination.
What are the three main structural components common to all microtomes?
The base (body), the knife plus knife attachment or holder, and the tissue or block holder.
How does a rotary microtome operate, and what is the typical thickness for paraffin sections?
The block moves up and down while advancing a preset number of μm per wheel revolution; typical paraffin section thickness is 3–6μm (often 4–5μm or one cell thick).
How does a sliding microtome function, and for what specimen types is it primarily used?
The tissue block remains stationary while the blade moves horizontally; it is used for cutting celloidin and large paraffin blocks.
What section thicknesses and knife materials are utilized in ultramicrotomy?
It cuts ultrathin sections (~90nm) for electron microscopy and ~0.5μm plastic sections for light microscopy using glass, diamond, or sapphire knives with a water bath adjacent to the blade.
Why has the benchtop freezing microtome largely been replaced by cryostats in modern laboratories?
It requires a bulky and hazardous CO2 cylinder, and CO2 freezing destroys antigens required for immunohistochemistry (IHC).
What are the shapes of microtome knife profiles A, B, C, and D?
Profile A is strongly planoconcave or biconcave; Profile B is planoconcave; Profile C is wedge-shaped; Profile D is plane-shaped.
What is the clearance angle on a microtome, and what are its standard ranges?
The angle between the block face and the lower cutting facet (the only angle adjusted on the microtome); it is set to ~3∘–8∘ for paraffin and 5∘–8∘ for resin or frozen sections.
What are the typical angle values for the bevel (cutting) angle and the wedge angle?
The bevel/cutting angle is ~27∘–32∘, and the wedge angle is ~15∘ (prior to sharpening).
Why must xylene never be used to clean a microtome knife?
Xylene leaves an oily residue on the blade edge that prevents section ribboning.
What are the standard operating temperatures of a cryostat for routine versus lipid-rich tissues?
Routine sectioning is maintained at ~−20∘C to −25∘C, whereas softer or lipid-rich tissues (such as liver, brain, spleen, and lymph nodes) are cut at ~−10∘C.
What are the primary clinical indications and diagnostic advantages of frozen sections?
Rapid STAT intraoperative diagnosis (~15min), staining of fat/lipids dissolved by routine solvents, demonstration of heat-labile enzymes, and preservation of undenatured proteins for immunofluorescence or IHC.
What is OCT compound, and how must tissues embedded in it be frozen and stored?
A glycol-based Optimum Cutting Temperature medium that supports tissue and prevents brittleness; it must be frozen rapidly (e.g., using a heat extractor or isopentane in liquid nitrogen) to prevent ice crystal holes and stored wrapped at −70∘C to prevent dehydration.
What sectioning defects occur if an anti-roll plate is too warm, too low, or too high?
If too warm, sections stick to the plate; if too low, sections ride over the top; if too high, the tissue block rubs against the plate.
What are the common causes and remedies for thick/thin or skipped sections during microtomy?
Causes: clearance angle too slight (block crushed/expanded) or too great, or loose screws/chuck. Remedies: adjust the clearance angle, tighten chuck and knife clamps, and service the unit if parts are worn.
What factors cause washboarding, chatter, or section curling, and how are they fixed?
Causes: clearance angle too steep, dull knife edge, or hard/over-dehydrated tissue. Remedies: reduce the clearance angle, change or shift the blade, and hydrate/cool tissue or cut at a slower speed.
How can a histotechnologist determine if vertical scratches in a ribbon are caused by knife damage versus tissue contents?
If a scratch stays in the exact same anatomical tissue structure across ribbons, the block contains hard material (e.g., calcifications or bone); if it occurs at a fixed blade point regardless of tissue position, the knife has a nick or debris.
How does the mechanical advancement system of a rotary microtome function?
A computerized motor or screw feed drives a ratchet wheel with 1,000 grooves (each notch representing 1μm or 0.001mm) engaged by a pawl (metal pin) to advance the carriage.
What distinguishes a semi-automated rotary microtome from a fully automated microtome?
A semi-automated microtome has one motor (coarse feed automatic and fine handwheel manual, or vice versa), while a fully automated microtome has two motors driving both coarse and fine handwheels (often operated by a foot pedal).
What are the specific definitions and values for knife angle of tilt, slant angle, and rake angle?
Angle of tilt is ~25∘ (~20∘ optimum in diagrams, ~30∘ in cryostats) between block face and the line bisecting the bevel angle; slant angle is 10∘–14∘ between lower knife side and block surface; rake angle is between the upper cutting facet and a line perpendicular to the block face.
What is the recommended protocol for facing (coarse-cutting) and cooling paraffin blocks?
Orient the block so the edge with least resistance hits first, coarse-cut at 10–15μm (block does not need to be cold before facing), then cool on a cold plate or ice water for ~1–2min up to 10–15min.
What causes the 'Venetian blind' effect, and what processing issue leads to mushy sections?
A 'Venetian blind' effect is caused by a dull knife, whereas mushy sections result from poor or insufficient tissue dehydration.
What is the distinction between cutting serial sections and cutting levels?
Serial sections retain every consecutive section cut from the block, whereas levels sample different tissue depths by collecting sections from ribbons every 8 to 10 sections.
What temperature should a water flotation bath be set at relative to paraffin wax?
Maintained at 5–10∘C below the paraffin melting point; because paraffin typically melts at 55–58∘C, the bath is usually ~45–46∘C.
What floating technique should be used, and what artifact results from over-soaking or excessive water heat?
Float sections shiny side down, touching the trailing end first to flatten within ~30s; excessive heat, soaking, or stretching creates tissue separation or a 'parched earth' artifact in delicate structures.
What three bonding mechanisms facilitate tissue adhesion to glass slides, and which is strongest?
Van der Waals forces, hydrogen bonding, and ionic interactions (which are the strongest).
What slide type represents the current routine standard for histology and IHC?
Positively charged slides, due to their strong section retention and low background staining.
What risks or drawbacks are associated with using gelatin and Mayer's egg albumin as section adhesives?
Gelatin supports microbial growth and can cause acid-fast artifact contamination in the bath; Mayer's egg albumin requires heat-fixing (~70∘C for 30min) and can cause heavy background staining if applied too thickly.
What are the temperature parameters for slide drying, and what artifact results from water entering xylene?
Drying is carried out at ~60∘C (~7–15min forced air or ~1hr oven), or <60∘C (40–50∘C or room temperature) for IHC to avoid nuclear bubbling; water carried into xylene blocks deparaffinization and causes intensely staining 'white spots.'
What is the sequence of chemical solutions used in deparaffinization?
Xylene (or clearing agent) removes paraffin, graded alcohols remove xylene, and water hydrates the tissue section for aqueous staining.
What is the ideal refractive index of mounting media, and when must aqueous media be selected over resinous media?
The ideal refractive index is ~1.54; aqueous mounting media are required when dehydration and clearing would dissolve or ruin the stain (e.g., fat/lipid stains, immunofluorescence, metachromatic stains, and enzyme histochemistry).
What causes a cloudy appearance under a resinous mountant, and what morphological artifact arises if a slide air-dries before coverslipping?
Cloudiness is caused by residual water remaining on the slide; air-drying creates brownish staining and glossy nuclei that mimic pyknotic cells seen in malignancies.
How do heterochromatin and euchromatin differ in structure and staining appearance?
Heterochromatin is condensed and intensely basophilic (staining darkly), whereas euchromatin is genetically active/extended and stains lightly or appears pale and clear.
What are the four physical principles of staining, and which one governs lipid staining?
Capillarity, adsorption, selective solubility, and permeability/porosity; selective solubility governs lipid staining (the dye is more soluble in tissue lipids than in its solvent).
How are chromophores, chromogens, and auxochromes defined in dye chemistry?
A chromophore is a color-bearing chemical group; a chromogen is an aromatic compound containing a chromophore (colored but unable to bind tissue); an auxochrome is an ionizing group (−NH2, −COOH, −SO3H, −OH) that enables salt formation and tissue binding.
How are basic and acid dyes classified based on ionic charge and target tissue binding?
They are named for the charge of their colored ion (not solution pH); basic (cationic) dyes bind anionic/acidic targets such as nucleic acids in nuclei, while acid (anionic) dyes bind cationic/basic targets such as proteins in cytoplasm, collagen, and red blood cells.
What are the distinct definitions of a mordant, an accentuator, and dye ripening?
A mordant is a metal or compound that links dye to tissue to form an insoluble dye lake; an accentuator increases staining intensity or speed without forming a lake; ripening is the chemical or atmospheric oxidation of hematoxylin into hematein.
What are two examples of chemical accentuators in histotechnology?
Potassium hydroxide (KOH) in methylene blue, and phenol in carbol fuchsin.
What is the procedural difference between progressive and regressive staining?
In progressive staining, tissue is stained until the desired endpoint intensity is reached and stopped; in regressive staining, tissue is intentionally over-stained and then differentiated to selectively remove excess dye from specific structures.
What are the four mechanisms used to differentiate regressive stains, with histological examples?
Acid differentiation (dilute HCl, H2SO4, or CH3COOH breaking metal-dye bonds in hematoxylin), mordant differentiation (weak FeCl2 in Verhoeff's hematoxylin), solvent differentiation (dilute alcohols after eosin), and oxidizing differentiation (bleaching with potassium ferricyanide or permanganate).
How do argyrophilic and argentaffin silver staining mechanisms differ?
Argyrophilic tissues bind silver ions but require an external reducing agent (e.g., hydroquinone or formalin) to produce black metallic silver; argentaffin tissues possess endogenous chemical groups that directly reduce silver ions without an external reducer.
What are four natural dyes used in histology, along with their biological sources?
Carmine (female cochineal bug), hematoxylin (heartwood of the logwood tree Hematoxylum campechianum), orcein (lichens), and saffron (dried Crocus stigmata).
How do formalin, dichromate, and picric acid fixatives alter subsequent staining affinity?
Formalin can reduce eosin uptake; dichromate decreases hematoxylin uptake and increases eosin uptake; picric acid enhances the binding of acidic (anionic) dyes; and over-fixation generally reduces nuclear basophilia.