Doppler
used to detect motion, primarily of blood flow
The Doppler Effect
• The Doppler shift or Doppler Frequency that describes the apparent change in pitch due to a change in position (movement) between the sound source and receiver
• The frequency of a sound changes when the sound source and receive move closer together or farther apart
If a wave reflects from a moving object, the reflected wave will …
have a different wavelength than the original wave
The faster the moving object …
the higher the frequency
Who first described the doppler effect ?
Christian Johann Doppler
Wavelength and frequency are
inversely related:
long wavelength = low frequency
small wavelength = high frequency
source of soundwave =
transducer
reflector of sound wave =
red blood cells
Positive doppler shift
Reflectors moving TOWARD the transducer will create a shorter wavelength than the original. This makes a higher frequency
Negative Doppler shift
Reflectors moving AWAY from the transducer will create a longer wavelength than the original. This makes a lower frequency
Motion that is exactly perpendicular (90 degrees)
will not show any shift in frequency
As RBCs move faster…
the reflected frequency is higher
Demodulation
• Signal processing that happens within the ultrasound system
• Looks at Doppler only, gets rid of gray scale
Doppler Shift =
received frequency – transmitted frequency
What units are used for transmitted and reflected frequencies ?
• Transmitted and reflected frequencies are in MHz
• This shift falls into an audible range (2-20kHz or 20-20,000Hz)
Positive Doppler Shifts
• When red blood cells move toward the transducer
• This is often assigned RED on color Doppler and appears ABOVE the baseline in spectral Doppler
Negative Doppler Shifts
• When red blood cells move away from the transducer
• This is often assigned BLUE on color Doppler and appear BELOW the baseline in spectral Doppler

spectral doppler
Speed gives information about ..
magnitude only. How fast is something traveling?
Velocity gives information about …
magnitude AND direction. How fast, and in what direction, is something traveling
Doppler equation
doppler shift = 2 x velocity of blood x transducer frequency x cosine theta / propagation speed (1.54)

Desired angle
60 degrees (or less) is the angle to achieve
• angle correct and make the cursor parallel to the vessel wall to allow the machine to make accurate velocity calculations.
Nondirectional
like a fetal doptone, measures the PRESENCE of red blood cells and makes audio with single speaker to hear it. Cannot determine direction
Bidirectional
Uses Doppler shift to determine direction of flow (positive or negative shift determines toward or away from the transducer)
• 2 speakers- one toward and one away. Think of headphones- one ear piece hears towards, the other hears away.
• Allows visual/graphic display of information- forward flow is above the baseline, away flow is below the baseline
Phase Quadrature
• Aka Quadrature detection
• Signal processing that separates the echoes into negative and positive echoes (allows directional information)
• Obtains Directional Information
Demodulation vs Quadrature detection
• Both are signal processing techniques
• Demodulation separates signals to extract Doppler frequency from transmitted frequency but DOES NOT give info about direction
• Quadrature detection separates echoes into negative and positive
Continuous wave Doppler crystal
• Two crystals required in the transducer
• One constantly transmits, One constantly receives
• Very accurately measures very high velocities

continuous wave doppler
Continuous wave Doppler
• CANNOT determine the exact location of the signal
• Listens along the ENTIRE LENGTH OF THE BEAM PATH
• Does NOT create an image
Continuous wave doppler advantages
Accurately measures very high velocity blood flow
Continuous wave doppler disadvanted
• Exact location of red blood cells cannot be determined.
• RANGE AMBIGUITY- Signals arise from all blood cells in the region of overlap between the transmit and receive beams
Pulsed Wave Doppler crystal
• Crystal alternates between transmitting and receiving
• 5-30 cycles per pulses with pulsed wave (2D is more like 2-3 cycles)
Pulsed wave doppler location
can be determined by the sonographer!
• huge advantage over continuous wave
• called range resolution: Range specificity, Freedom from range ambiguity artifact
Sample volume or gate
the small marker that can be adjusted by the sonographer to determine the location to sample to Doppler signal

sample volume
Pulsed wave doppler advantages
• Ability to select the exact location where velocity is measured.
• This is called Range resolution, range specificity, or freedom from Prange ambiguity artifact
Pulsed wave doppler disadvantages
• Inaccurate measurement of high velocity signals.
• High velocity flow in one direction is incorrectly displayed as traveling in the opposite direction (Aliasing)
Aliasing
• most common error associated with Doppler ultrasound.
• a false depiction of very high velocities incorrectly displayed as going the opposite direction
• When flow is very fast and goes all the way to the top of the spectral image (forward flow), it then wraps around incorrectly and reappears at the very bottom of the display

aliasing
Nyquist limit
the very top of the display with the maximum velocity that can be displayed before aliasing occurs
• Or the opposite, the very bottom of the display is the maximum away velocity that can be displayed before aliasing wraps around to the top of the screen

red: aliasing
yellow: nyquisit limit
When does aliasing occur ?
• Occurs with high velocities (because the velocity needs to be measured at such a rapid rate)
• Occurs with PW doppler due to pulses
• Never happens with CW doppler: able to measure high velocities, continuous sampling
PRF
Number of pulses per second
• As depth decreases, PRF increases: the system only waits a short time for the return and can create new pulses more rapidly.
• As depth increases, PRF decreases: the system waits for a longer time for the return which affects the frequency of creating new pulses.
Nyquist limit=
½ the PRF OR PRF/2
• Nyquist limit and PRF are DIRECTLY related
If PRF is too low
the blood flow cannot be sampled fast enough.
Shallow Sample
• High PRF: rapidly moving bloodis measured frequently
• Nyquist limit (speed limit) is high
• High velocities measured accurately
Deep Sample
• Low PRF: rapidly moving blood is measured infrequently
• Nyquist limit (speed limit) is low
• System is unable to measure high velocity flow
• Prone to aliasing.
Aliasing summary
• Occurs when then Doppler shift (speed) exceeds the Nyquist limit (speed limit)
• Avoid this error by raising the Nyquist limit (speed limit)
• Reduce the Doppler shift

A: Aliasing appears (black arrow)
B: aliasing is eliminated by reducing the doppler shift
C: Aliasing is eliminated by increasing the nyquist limit
Ways to fix/avoid aliasing
Adjust the scale to it’s maximum
Find a different view with a shallower sample volume
Use a lower frequency transducer
4. Shift the baseline
5. Use continuous wave Doppler
Gray Shades of a Spectrum
• Amplitude of the reflected signal or Number of blood cells creating the reflection
• Brighter areas on the spectrum (waveform) mean that there are MORE red blood cells at that particular point, creating the Doppler shift.
X-axis in grey shades
time
y-axis in grey shades
velocity
• Above the baseline is a positive shift (towards the transducer)
• Below the baseline is a negative shift (away from the transducer)
Color Flow Doppler display
• Two-dimensional Doppler: Not in a single location (PW) or along a single transmit line (CW)
• Black and white= anatomy
• Color= blood flow
Color flow doppler
• Gives information on location so it HAS to be a pulsed ultrasound
technique
• Has range resolution (or range specificity)
• IS subject to aliasing
• Gives information about direction (BASICALLY IS A COLOR DEPICTION OF THE DOPPLER SHIFT)
• Is semiquantitative
Semiquantitative
• Angles are less important than with PW
• Provides mean(average) velocities
• CW and PW give peak velocities
Color Maps
• Convert measured velocity information into color that appears on gray scale image
• The dictionary or look up table that is used for this conversion is displayed is a vertical bar of various colors with black at it’s center
Two common dictionaries of color maps
1. Velocity mode
2. Variance mode

two different velocity mode maps
1: displays flow direction info
2: displays changes in velocity
Color with Rectangular Images: home
• To figure out if flow is moving toward or away
• To find “home” find the corner of the color box where you can draw a straight line down and not hit the colored part of the vessel.
• If the vessel is blue/below baseline on the color map, the blood is moving AWAY from the transducer (imaginary “home” line created)
• Vice versa for red

Power Doppler
• AKA energy mode or color angio
• Identifies PRESENCE of a Doppler Shift
• DOES NOT evaluate speed or direction
• Non directional color Doppler
• All vessels have the same color in Power Doppler
• Strength (amplitude) is processed without deciphering direction or speed
Amplitude in power doppler.
directly related to the number of moving blood cells
• Larger amplitude signals are brighter
• Weaker amplitude signals are darker.
Advantages of Power Doppler
• Increased sensitivity to low flow/velocity: Venous flow , Small vessels
• Unaffected by Doppler angles: EXCEPT exactly 90 degrees
• No aliasing : Velocity information is ignore
Disadvantages of Power Doppler
• No measurement of velocity
• Lower frame rates
• Flash artifact or bursts of color can occur from motion of the transducer, patient, or soft tissue.
Ghosting
low frequencies that appear on color Doppler
Clutter
low frequencies that appear on Spectral Doppler
• Wall filter (aka high pass filter) eliminates low frequency Doppler shifts, essentially rejects tiny Doppler shifts.
Crosstalk
mirror image of spectral display with baseline being the mirror
• Fix this with decreasing wave gain or changing angle
Spectral Analysis
• Not all blood cells travel at the same speed or direction
• breaks the signals into individual signals that make up the entire Doppler signal.
Two methods for spectral analysis
1. Fast Fourier Transformation (FFT)
• PW and CW
2. Autocorrelation
• Color