Implantable Devices (Pacemaker, ICD, CRT, Loop Recorder)

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/16

encourage image

There's no tags or description

Looks like no tags are added yet.

17 Terms

1
New cards

What is the advantage of using biphasic over monophasic shock waveforms for ICDs and external defibrillators?

The reversing polarity of the devices makes biphasic waveforms more effective such that VF is terminated at lower energies with less injury to the myocardium.

2
New cards

The principle of “(R-wave) synchronization” is needed for what shocking modality?

Cardioversion

3
New cards

During a pacemaker gen change, the leads are disconnected and reconnected. What might have gone wrong during the procedure if, post-procedure, the ventricular impedance reading suddenly skyrockets and we have loss of capture?

The lead pin was not fully (re)-inserted

4
New cards

What is pace-sense conductor fracture?

An episode on an ICD where the near-field channel shows non-physiologic spikes, and the far-field channel shows normal rhythm. There is also a spike in impedance intermittently. This may be provoked by reaching one’s arms upward.

5
New cards

What is SVT discrimination on an ICD?

SVT discriminators work for presenting rhythms that fall within the programmed zone (i.e. think about delivering a shock for VT at or above 170 bpm). Algorithms can reference many things, including: rhythm onset, R-R (ir)regularity, QRS morphology, PR interval, AV association, A and V rates, rhythm onset and in what chamber, etc.

6
New cards

Which ICD ventricular rate sensing configuration is more prone to oversensing diaphragmatic myopotentials?

Integrated bipolar sensing uses the tip as one electrode and the RV shock coil as the second electrode, which widens the sensing field. Thus, diaphragmatic and far-field signals are picked up more easily.

7
New cards

What is chronotropic incompetence, and how can it be managed with a pacemaker?

Chronotropic incompetence means that the heart rate does not increase adequately enough to support a given level of activity or metabolic demand. Usually 80%+ of the age-predicted max is an acceptable heart rate; otherwise, rate-responsive pacing will raise the paced rate with activity.

8
New cards

A stimulator delivers a 5 mA pulse lasting 0.5 ms. How much charge is delivered per pulse?

Q = Current x Time. So Q = (5E-3 A) x (0.5 E-3 s) = 2.5E-6 C = 2.5 microcoulombs

9
New cards

For an ICD, what is non-committed shock reconfirmation?

An ICD will reconfirm that the patient is still in VT/VF before delivering a shock after having charged. If sinus rhythm has returned, the shock is aborted and the charge is dumped internally.

10
New cards

For an ICD: If a high-voltage shock impedance reading is significantly lower than the one before it, what might be wrong?

There is an insulation breach in the shock circuit and some of the current is taking a path of least resistance through a short, so the shock impedance is lowered

11
New cards

For a pacemaker check: If the RV lead impedance has significantly dropped (below 200 -250 ohms) and there is intermittent loss of capture from the lead, what is most likely wrong with the lead?

Insulation breach - current is leaking through the damaged lead insulation, lowering the output to the myocardium and potentially not capturing

12
New cards

What might happen to impedance in the event of:

  1. Pacemaker conductor fracture

  2. Loose set screw


  1. Raise impedance. Conductor fracture may cause inappropriate shocks or failed therapy.

  2. Raise impedance


13
New cards

During a CRT implant, if a contrast injection through the CS sheath leaves a persistent linear stain along the vessel wall with slow clearance, and the patient remains hemodynamically stable, what is the most likely complication?

Coronary sinus dissection, a complication of sheath manipulation or forceful injection. May need to stop injections and/or resume procedure later

14
New cards

Formulas to calculate:

  1. Energy per pacing pulse delivered

  2. Charge


  1. E = (V² x R x t) or Energy = Voltage squared *lead impedance*time duration

  2. C = I x t or Charge = Current x Time


15
New cards

Why is a lower shock impedance associated with a dual-coil ICD system (RV coil plus SVC coil active with the can) compared to a single-coil system?

The extra coil adds a parallel path and lowers total resistance, so shock impedance falls.

16
New cards

Why might an LV lead for a CRT-D patient cause hiccup-like twitching for every paced beat? What is the best fix?

Capturing phrenic nerve along LV lateral wall and stimulate the diaphragm. Reprogram the LV vector or output to avoid phrenic capture.

17
New cards

Say we have a patient with a 9-year-old dual-chamber pacemaker who is being upgraded to a CRT. Contrast venography through the left arm vein shows abrupt cut-off of the left subclavian vein with multiple collateral channels draining toward the neck. What is the correct interpretation, and what should be done?

There is chronic occlusion of the subclavian vein, and it’s been occluded long enough that collateral veins have formed. The upgrade to a CRT requires venoplasty (vein widening), a different access, or lead extraction to regain the vein.