The ultrasound PRF is equal to the ______ pulse repetition frequency of the pulser
voltage
Increased voltage amplitude produced by the pulser increases the ______ and _____ of ultrasound pulses produced by the transducer
amplitude, frequency
If a 6 MHz transducer images to a depth of 10cm, to avoid range ambiguity, the maximum PRF permitted is ______
7.7 kHz

Functions performed by the signal processor include:
filtering, detection, compression
What does detection do?
Converts pulses from radio frequency to amplitude form

What does amplification do?
Increases all amplitudes

What does compression do?
Decreases dynamic range
What does compensation do?
Corrects for tissue attenuation
What does filtering do?
Reduces noise
If the input voltage to an amplifier is 1mV and the output voltage is 10mV, the voltage amplification ratio is ____, the power ratio is ___, the gain is ____ dB
10,100,20

An amplifier with a gain of 60 dB has 1 mW of power applied to the input. The output power is:
1 W

An amplifier with a gain of 60 dB has 10 mV of voltage applied to the input. The output voltage is what?
10 mV
Time gain compensation is accomplished in the ____
beam former
Compensation takes into account reflector ____
depth
Compensation decreases the dynamic range to a range that the _______ and ______ can handle
display, human vision
If a display has a dynamic range of 20 dB and the smallest voltage it can handle is 200 mV, then the largest voltage it can handle is ___
2 V
Detection converts voltage pulses from _____ form to _____ form
radio frequency, amplitude
True or false: filtering widens bandwidth
False
Filtering is accomplished in the ______
signal processor
The compression or dynamic range control reduces the range of echo amplitudes displayed by reducing the weakest to _____ and/or the strongest ones to the ____ value and assigning the others to increasing values. This produces a ____ contrast images with elimination of weakest and maximizing of strongest echoes.
zero, maximum, higher
An amplifier has a power output of 100mW when the input power is 0.1mW. The amplifier gain is:
30dB
If the beam-former output to the transducer is reduced by 3, 6, or 9 dB, the ultrasound pulse output intensity is reduced by:
50%, 75%, 87.5%
One watt is ___ dB below 100W
20
One watt is ___ dB above 100 mW
10
Id the input power is 1mW and the output is 10,000 mW, the gain is:
40 dB
If an amplifier has a gain of 15 dB, the ratio of output power to input power is ____
32
If the output of a 22dB gain amplifier is connected to the input of a 23 dB gain amplifier, the total gain is _____ dB. The overall power ratio is _____
45, 32000
If a 17 dB electric attenuator is connected to a 15 dB amplifier, the net gain is _____ dB. The net attentuation is ______ dB. For a 1W input, the output is ____W.
-2, 2, .63

For the digital memory shown, enter the number store in each pixel location
Lower right:
Middle right:
Upper right:
Upper middle:
Upper left:
6, 9, 10, 13, 14
The contrast resolution for an instrument that has an exact dynamic range of 43 dB and 32 shades is ___ per shade
1.3
The contrast resolution for an instrument that has a memory and a 45 dB echo dynamic range is ____ dB per shade
0.7
Pixel:
picture element
Preprocessing:
assignment of stored numbers
Digital:
discrete
Bit:
binary digit
Analog:
proportional
Postprocessing:
assignment of displayed brightness
Typical image pixel dimensions are at least _____
1024 × 768
How many bits per pixel are required for each number of shades:
16
32
64
128
256
4
5
6
7
8
______ total memory elements are required for a 100 × 100 pixel, 5 bit digital memory
50,000
Memories of _____ bits per pixel are common in ultrasound today
8-16
Digital memories store what?
Numbers
_____ is commonly controllable by the operator
Postprocessing
In binary numbers, how many symbols are used?
2
The term binary digit is commonly shortened into the single word ____
bit
Each binary digit in a binary number is represented in the memory by a memory element, which, at any time, is in one of _____ states that are ______ or _______
2, on, off
Match the 8-bit binary number hgfedcba to the decimal number represented by a 1 in the column
A 1
B 2
C 4
D 8
E 16
F 32
G 64
H 128
The binary number 10110 represents zero 1, one 2, one 4, and one 16. So the decimal number is 0 + 2 + 4 + 0 + 16 = 22. What decimal number does the binary number 11001 represent?
25
The decimal number 13 is made up of one 1, zero 2, one 4, and one 8 (8+4+0+1 = 13). The number therefore is represented by the binary number ____
1101
What is the binary number for the following:
1
5
10
15
20
25
30
40
50
100
0000001
0000101
0001010
0001111
0010100
0011001
0011110
0101000
0110010
1100100
How many binary digits are required in the binary numbers representing the following decimal numbers?
0
1
5
10
25
30
63
64
75
100
1
1
3
4
5
5
6
7
7
7
How many bits are required to represent each decimal number in binary form?
7
15
3
511
1023
63
255
1
127
31
3
4
2
9
10
6
8
1
7
5
How many bits are required to store numbers representing each number of different gray shades?
2
4
8
15
16
25
32
64
65
128
1
2
3
4
4
5
5
6
7
7
The primary formats of image presentations are:
They are used in what types of clinical imaging respectivelly
B mode, M mode, A mode
all, cardiac, opthalmic
____ mode is used for studying the motion of a structure such as a heart valve
M
B scan presents a cross-section through the _____
scan plane
A display that shows various echo strengths as different brightnesses is called a _____ or _____ display
grayscale, B mode
The _______ stores the gray-scale image and allows it to be displayed on a computer monitor
image memory
Which arrays have rectangular image and sector image
Rectangular: linear array
Sector: convex array, phased array, vector array
If the PRF of an instrument is 1kHz and it displays (single focus) 25 fps, there are ____ lines per frame
40
The PRF is ______ is Hz if 30 frames (40 lines each) are displayed per second (single focus)
1200
True or false imaging involving 10cm depth penetration, a single focus, 100 scan lines per frame, and 30 frames per second can be accomplished without range ambiguity.
True
The maximum frame rate permitted for 15cm penetration, 3 foci, and 200 scan lines per frame is _____
8.5
The primary components of a diagnostic ultrasound imaging system are what?
Transducer, beam former, signal processor, image processor, and display
The information that can be obtained from an M mode display includes _____
distance and motion pattern
The TGC control compensates for ____
attenuation
A grayscale display shows _____
a range of echo amplitudes
The dynamic range of an ultrasound system is ______
the ratio of the maximum amplitude to the minimum echo strength that can be displayed
A _______ formats scan line data to image form
scan convertor
Which of the following is not performed by a signal processor?
digital-to-analog conversion
In a digital memory, echo intensity is represented by what?
A number stored in the memory
If there were no attenuation in tissue, ____ would not be needed
TGC
Echo imaging includes ultrasound generation, propagation and reflection in tissues, and reception of returning _____
echoes
The diagnostic ultrasound systems in common clinical use today are the _____ type
pulse-echo
Grayscale instruments show echo amplitude as ____ on the display
brightness
Pulse-echo instruments look for three things:
strength, direction, and arrival time of echoes returning from tissues
An image memory stores image information in the form of ____ numbers
binary
Imaging systems produce a visual ____ from the electric ____ received from the transducer
display, voltages
The transducer is connected to the signal processor through the ______
beam former
The transducer receives voltages from the ____ in pulse-echo systems
beam former
The _______ receives digitized echo voltages from the beam former
signal processor
Increasing gain generally produces the same effect as what?
Decreasing attenuation
The ADC is part of the _______
beam former
Voltage pulses from the pulser are applied through delays to the _____
transducer
Detection is a function of the:
signal processor
If gain is reduced by one-half, with input power unchanged, the output power is _____ what it was before
one-half
If gain is 30 dB and output power is reduced by one-half, the new gain is _____ dB
27
If four shades of gray are shown on a display, each twice the brightness of the preceding one, the brightest shade is ___ times the brightness of the dimmest shade
8
The dynamic range is 9dB
In which units are gain and attenuation usually expressed?
dB
True or false TGC makes up for the fact that reflections from deeper reflectors arrive at the transducer with greater amplitude
False
The modes that show 1D real-time images are what?
A and M mode
The mode that shows 2D real-time images is the ____ mode
B
A real-time B mode display may be produced by rapid _____ scanning of a transducer array
electronic
Each complete scan of the sound beam produces an image on the display that is called a ___
frame
For a single focus, the number of lines in each frame is equal to the number of times the transducer is _____ while the frame is produced; that is, while the sound beam is scanned
pulsed
In real-time scanning, the required PRF depends on what?
number of lines per frame, frame rate
Increasing the number of foci may reduce what?
frame rate
To correct for attenuation, TGC must _____ the gain for increasing depth
increase
If a higher frequency is used, resolution is ____, imaging depth ____, and the TGC slope must be ______
improved, decreases, increased
For pixel dimensions 512×512 and 1920×1200, calculate the number of image pixels
262, 144; 2,304,000