The ultrasound PRF is equal to the ______ pulse repetition frequency of the pulser

voltage

2
New cards

Increased voltage amplitude produced by the pulser increases the ______ and _____ of ultrasound pulses produced by the transducer

amplitude, frequency

3
New cards

If a 6 MHz transducer images to a depth of 10cm, to avoid range ambiguity, the maximum PRF permitted is ______

7.7 kHz

<p>7.7 kHz</p>
4
New cards

Functions performed by the signal processor include:

filtering, detection, compression

5
New cards

What does detection do?

Converts pulses from radio frequency to amplitude form

<p>Converts pulses from radio frequency to amplitude form </p>
6
New cards

What does amplification do?

Increases all amplitudes

<p>Increases all amplitudes </p>
7
New cards

What does compression do?

Decreases dynamic range

8
New cards

What does compensation do?

Corrects for tissue attenuation

9
New cards

What does filtering do?

Reduces noise

10
New cards

If the input voltage to an amplifier is 1mV and the output voltage is 10mV, the voltage amplification ratio is ____, the power ratio is ___, the gain is ____ dB

10,100,20

<p>10,100,20 </p>
11
New cards

An amplifier with a gain of 60 dB has 1 mW of power applied to the input. The output power is:

1 W


<p>1 W </p><p></p>
12
New cards

An amplifier with a gain of 60 dB has 10 mV of voltage applied to the input. The output voltage is what?

10 mV

13
New cards

Time gain compensation is accomplished in the ____

beam former

14
New cards

Compensation takes into account reflector ____

depth

15
New cards

Compensation decreases the dynamic range to a range that the _______ and ______ can handle

display, human vision

16
New cards

If a display has a dynamic range of 20 dB and the smallest voltage it can handle is 200 mV, then the largest voltage it can handle is ___

2 V

17
New cards

Detection converts voltage pulses from _____ form to _____ form

radio frequency, amplitude

18
New cards

True or false: filtering widens bandwidth

False

19
New cards

Filtering is accomplished in the ______

signal processor

20
New cards

The compression or dynamic range control reduces the range of echo amplitudes displayed by reducing the weakest to _____ and/or the strongest ones to the ____ value and assigning the others to increasing values. This produces a ____ contrast images with elimination of weakest and maximizing of strongest echoes.

zero, maximum, higher

21
New cards

An amplifier has a power output of 100mW when the input power is 0.1mW. The amplifier gain is:

30dB

22
New cards

If the beam-former output to the transducer is reduced by 3, 6, or 9 dB, the ultrasound pulse output intensity is reduced by:

50%, 75%, 87.5%

23
New cards

One watt is ___ dB below 100W

20

24
New cards

One watt is ___ dB above 100 mW

10

25
New cards

Id the input power is 1mW and the output is 10,000 mW, the gain is:

40 dB

26
New cards

If an amplifier has a gain of 15 dB, the ratio of output power to input power is ____

32

27
New cards

If the output of a 22dB gain amplifier is connected to the input of a 23 dB gain amplifier, the total gain is _____ dB. The overall power ratio is _____

45, 32000

28
New cards

If a 17 dB electric attenuator is connected to a 15 dB amplifier, the net gain is _____ dB. The net attentuation is ______ dB. For a 1W input, the output is ____W.

-2, 2, .63

29
New cards
<p>For the digital memory shown, enter the number store in each pixel location </p><p>Lower right: </p><p>Middle right: </p><p>Upper right: </p><p>Upper middle: </p><p>Upper left: </p>

For the digital memory shown, enter the number store in each pixel location

Lower right:

Middle right:

Upper right:

Upper middle:

Upper left:

6, 9, 10, 13, 14

30
New cards

The contrast resolution for an instrument that has an exact dynamic range of 43 dB and 32 shades is ___ per shade

1.3

31
New cards

The contrast resolution for an instrument that has a memory and a 45 dB echo dynamic range is ____ dB per shade

0.7

32
New cards

Pixel:

picture element

33
New cards

Preprocessing:

assignment of stored numbers

34
New cards

Digital:

discrete

35
New cards

Bit:

binary digit

36
New cards

Analog:

proportional

37
New cards

Postprocessing:

assignment of displayed brightness

38
New cards

Typical image pixel dimensions are at least _____

1024 × 768

39
New cards

How many bits per pixel are required for each number of shades:

16

32

64

128

256

4

5

6

7

8

40
New cards

______ total memory elements are required for a 100 × 100 pixel, 5 bit digital memory

50,000

41
New cards

Memories of _____ bits per pixel are common in ultrasound today

8-16

42
New cards

Digital memories store what?

Numbers

43
New cards

_____ is commonly controllable by the operator

Postprocessing

44
New cards

In binary numbers, how many symbols are used?

2

45
New cards

The term binary digit is commonly shortened into the single word ____

bit

46
New cards

Each binary digit in a binary number is represented in the memory by a memory element, which, at any time, is in one of _____ states that are ______ or _______

2, on, off

47
New cards

Match the 8-bit binary number hgfedcba to the decimal number represented by a 1 in the column

A 1

B 2

C 4

D 8

E 16

F 32

G 64

H 128

48
New cards

The binary number 10110 represents zero 1, one 2, one 4, and one 16. So the decimal number is 0 + 2 + 4 + 0 + 16 = 22. What decimal number does the binary number 11001 represent?

25

49
New cards

The decimal number 13 is made up of one 1, zero 2, one 4, and one 8 (8+4+0+1 = 13). The number therefore is represented by the binary number ____

1101

50
New cards

What is the binary number for the following:

1

5

10

15

20

25

30

40

50

100

0000001

0000101

0001010

0001111

0010100

0011001

0011110

0101000

0110010

1100100

51
New cards

How many binary digits are required in the binary numbers representing the following decimal numbers?

0

1

5

10

25

30

63

64

75

100

1

1

3

4

5

5

6

7

7

7

52
New cards

How many bits are required to represent each decimal number in binary form?

7

15

3

511

1023

63

255

1

127

31

3

4

2

9

10

6

8

1

7

5

53
New cards

How many bits are required to store numbers representing each number of different gray shades?

2

4

8

15

16

25

32

64

65

128

1

2

3

4

4

5

5

6

7

7

54
New cards

The primary formats of image presentations are:

They are used in what types of clinical imaging respectivelly

B mode, M mode, A mode

all, cardiac, opthalmic

55
New cards

____ mode is used for studying the motion of a structure such as a heart valve

M

56
New cards

B scan presents a cross-section through the _____

scan plane

57
New cards

A display that shows various echo strengths as different brightnesses is called a _____ or _____ display

grayscale, B mode

58
New cards

The _______ stores the gray-scale image and allows it to be displayed on a computer monitor

image memory

59
New cards

Which arrays have rectangular image and sector image

Rectangular: linear array

Sector: convex array, phased array, vector array

60
New cards

If the PRF of an instrument is 1kHz and it displays (single focus) 25 fps, there are ____ lines per frame

40

61
New cards

The PRF is ______ is Hz if 30 frames (40 lines each) are displayed per second (single focus)

1200

62
New cards

True or false imaging involving 10cm depth penetration, a single focus, 100 scan lines per frame, and 30 frames per second can be accomplished without range ambiguity.

True

63
New cards

The maximum frame rate permitted for 15cm penetration, 3 foci, and 200 scan lines per frame is _____

8.5

64
New cards

The primary components of a diagnostic ultrasound imaging system are what?

Transducer, beam former, signal processor, image processor, and display

65
New cards

The information that can be obtained from an M mode display includes _____

distance and motion pattern

66
New cards

The TGC control compensates for ____

attenuation

67
New cards

A grayscale display shows _____

a range of echo amplitudes

68
New cards

The dynamic range of an ultrasound system is ______

the ratio of the maximum amplitude to the minimum echo strength that can be displayed

69
New cards

A _______ formats scan line data to image form

scan convertor

70
New cards

Which of the following is not performed by a signal processor?

digital-to-analog conversion

71
New cards

In a digital memory, echo intensity is represented by what?

A number stored in the memory

72
New cards

If there were no attenuation in tissue, ____ would not be needed

TGC

73
New cards

Echo imaging includes ultrasound generation, propagation and reflection in tissues, and reception of returning _____

echoes

74
New cards

The diagnostic ultrasound systems in common clinical use today are the _____ type

pulse-echo

75
New cards

Grayscale instruments show echo amplitude as ____ on the display

brightness

76
New cards

Pulse-echo instruments look for three things:

strength, direction, and arrival time of echoes returning from tissues

77
New cards

An image memory stores image information in the form of ____ numbers

binary

78
New cards

Imaging systems produce a visual ____ from the electric ____ received from the transducer

display, voltages

79
New cards

The transducer is connected to the signal processor through the ______

beam former

80
New cards

The transducer receives voltages from the ____ in pulse-echo systems

beam former

81
New cards

The _______ receives digitized echo voltages from the beam former

signal processor

82
New cards

Increasing gain generally produces the same effect as what?

Decreasing attenuation

83
New cards

The ADC is part of the _______

beam former

84
New cards

Voltage pulses from the pulser are applied through delays to the _____

transducer

85
New cards

Detection is a function of the:

signal processor

86
New cards

If gain is reduced by one-half, with input power unchanged, the output power is _____ what it was before

one-half

87
New cards

If gain is 30 dB and output power is reduced by one-half, the new gain is _____ dB

27

88
New cards

If four shades of gray are shown on a display, each twice the brightness of the preceding one, the brightest shade is ___ times the brightness of the dimmest shade

8

The dynamic range is 9dB

89
New cards

In which units are gain and attenuation usually expressed?

dB

90
New cards

True or false TGC makes up for the fact that reflections from deeper reflectors arrive at the transducer with greater amplitude

False

91
New cards

The modes that show 1D real-time images are what?

A and M mode

92
New cards

The mode that shows 2D real-time images is the ____ mode

B

93
New cards

A real-time B mode display may be produced by rapid _____ scanning of a transducer array

electronic

94
New cards

Each complete scan of the sound beam produces an image on the display that is called a ___

frame

95
New cards

For a single focus, the number of lines in each frame is equal to the number of times the transducer is _____ while the frame is produced; that is, while the sound beam is scanned

pulsed

96
New cards

In real-time scanning, the required PRF depends on what?

number of lines per frame, frame rate

97
New cards

Increasing the number of foci may reduce what?

frame rate

98
New cards

To correct for attenuation, TGC must _____ the gain for increasing depth

increase

99
New cards

If a higher frequency is used, resolution is ____, imaging depth ____, and the TGC slope must be ______

improved, decreases, increased

100
New cards

For pixel dimensions 512×512 and 1920×1200, calculate the number of image pixels

262, 144; 2,304,000