What are the three major factors affecting radiographic quality?
Receptor exposure, spatial resolution, and distortion.
What is receptor exposure?
The degree to which an image receptor is exposed to ionizing radiation.
What is the primary controlling factor of receptor exposure?
mAs.
What does mA control?
The quantity of x-ray photons produced per unit of time.
What does exposure time control?
The duration of x-ray production.
What does mAs stand for?
Milliampere-seconds.
How is mAs calculated?
mA multiplied by exposure time in seconds.
What happens to receptor exposure when mAs doubles?
Receptor exposure doubles.
What is the secondary controlling factor of receptor exposure?
kVp.
What does kVp primarily control?
The energy and penetrating power of the x-ray beam.
What does the 15% rule state?
Increasing kVp by 15% approximately doubles receptor exposure.
How should mAs be adjusted when kVp increases by 15% to maintain receptor exposure?
Decrease mAs by half.
How should mAs be adjusted when kVp decreases by 15% to maintain receptor exposure?
Double the mAs.
What does SID stand for?
Source-to-image receptor distance.
What happens to receptor exposure when SID increases without adjusting technique?
Receptor exposure decreases.
What law describes the relationship between distance and radiation intensity?
Inverse square law.
What happens to radiation intensity when distance is doubled?
It decreases to one-fourth.
What happens to radiation intensity when distance is cut in half?
It increases four times.
What is the primary purpose of a grid?
To absorb scatter radiation before it reaches the image receptor.
What material is commonly used in the absorbing strips of a grid?
Lead.
How does a grid improve image quality?
It reduces scatter radiation and improves contrast.
Does using a physical grid generally require an increase or decrease in exposure factors?
An increase.
What is grid cutoff?
A reduction in receptor exposure caused by unwanted absorption of primary radiation by the grid.
What is a virtual grid?
Scatter correction software that produces an image similar to one obtained using a physical grid.
What is the purpose of collimation?
To restrict the x-ray beam to the area of interest.
How does decreasing collimation field size affect scatter radiation?
It decreases scatter radiation.
How does decreasing collimation field size affect image contrast?
It improves image contrast.
What is an exposure indicator?
A numerical value representative of the radiation exposure received by the image receptor.
What does EI stand for?
Exposure index or exposure indicator.
What does the S-number represent?
A sensitivity number used as an exposure indicator.
What does REX stand for?
Reached exposure value.
What does DEI stand for?
Detector exposure index.
What are the two possible relationships between exposure indicators and receptor exposure?
Direct and inverse relationships.
In a directly proportional system, what happens to EI when receptor exposure increases?
EI increases.
In a directly proportional system, what happens to EI when receptor exposure decreases?
EI decreases.
In an inversely proportional system, what happens to EI when receptor exposure increases?
EI decreases.
In an inversely proportional system, what happens to EI when receptor exposure decreases?
EI increases.
In an inverse EI system with an acceptable range of 400–800, which number represents greater receptor exposure?
400.
Why should a technologist check the exposure indicator?
To help ensure adequate receptor exposure with the least possible patient dose.
What factors affect the exposure indicator?
mAs, kVp, detector area irradiated, and objects exposed.
Can an overexposed digital image still appear acceptable on a monitor?
Yes.
What does insufficient receptor exposure produce?
A noisy image.
What does excessive receptor exposure cause?
Unnecessary radiation exposure to the patient.
Is the exposure indicator a direct measurement of patient dose?
No. It represents exposure received by the image receptor.
What is spatial resolution?
The recorded sharpness of structures on an image.
What are other terms for spatial resolution?
Detail, recorded detail, image sharpness, and definition.
How is spatial resolution measured?
Line pairs per millimeter (lp/mm).
What is a lack of spatial resolution called?
Blur or unsharpness.
What is the greatest deterrent to image sharpness?
Motion.
What are the two types of motion?
Voluntary and involuntary motion.
What is voluntary motion?
Motion that the patient can control.
What is an example of voluntary motion?
Breathing or body movement.
What is involuntary motion?
Motion that the patient cannot consciously control.
What is an example of involuntary motion?
Peristalsis.
How can voluntary motion be minimized?
Patient instructions, cooperation, immobilization, and short exposure times.
How can involuntary motion be minimized?
Short exposure times.
Does motion increase or decrease spatial resolution?
Decrease.
Which focal spot size produces better recorded detail?
Small focal spot.
What is geometric unsharpness?
Blurring of object edges caused by imaging geometry and focal spot size.
What is penumbra?
The unsharp edges of a projected image.
Which focal spot produces less penumbra?
Small focal spot.
How does increasing SID affect spatial resolution?
It improves spatial resolution.
What does OID stand for?
Object-to-image receptor distance.
How does decreasing OID affect spatial resolution?
It improves spatial resolution.
What factors improve spatial resolution?
Small focal spot, short exposure time, faster IR system, longer SID, and shorter OID.
What is distortion?
Misrepresentation of the size or shape of an object on an image.
What are the two types of distortion?
Size distortion and shape distortion.
What is another name for size distortion?
Magnification.
What is magnification?
An increase in the apparent size of an object on an image.
What are the four controlling factors of distortion?
SID, OID, object-IR alignment, and CR alignment.
How does increasing SID affect magnification?
It decreases magnification.
How does decreasing SID affect magnification?
It increases magnification.
How does increasing OID affect magnification?
It increases magnification.
How does decreasing OID affect magnification?
It decreases magnification.
What SID is commonly used for routine skeletal radiography?
40 inches.
What SID is commonly used for chest radiography?
72 inches.
Why is a longer SID used for chest radiography?
To minimize magnification of the heart and other structures.
How should the object be aligned with the IR to minimize distortion?
Parallel to the IR.
What happens when the object is not parallel to the IR?
Shape and size distortion may occur.
What happens to distortion as the angle of inclination between the object and IR increases?
Distortion increases.
What is foreshortening?
Shape distortion in which anatomy appears shorter than its actual length.
What is elongation?
Shape distortion in which anatomy appears longer than its actual length.
What does CR stand for?
Central ray.
Where does the least possible distortion occur within the x-ray beam?
At the central ray.
How is the CR normally aligned to the IR?
Perpendicular, or at 90 degrees.
How does distortion change as beam divergence increases away from the CR?
Distortion increases.
How should digits be aligned with the IR to demonstrate open interphalangeal joints?
Parallel to the IR.
What happens to interphalangeal joint spaces when the digits are not parallel to the IR?
The joint spaces may appear closed.
Why is correct CR centering important?
To minimize distortion and demonstrate anatomy accurately.
What are the three major digital imaging characteristics?
Spatial resolution, contrast resolution, and image signal.
What is a pixel?
A picture element in a digital image.
What is a matrix?
An arrangement of pixels in rows and columns.
How does decreasing pixel size affect spatial resolution?
It improves spatial resolution.
How does increasing matrix size affect pixel size when the field of view remains constant?
Pixel size decreases.
What is pixel pitch?
The distance from the center of one pixel to the center of an adjacent pixel.
How does decreasing pixel pitch affect spatial resolution?
It improves spatial resolution.
What is MTF?
Modulation transfer function, a measure used to evaluate imaging system resolution.
What does contrast describe?
The differences in brightness between light and dark areas of an image.
What is contrast resolution?
The ability of an imaging system to distinguish small differences in intensity.
What is high contrast?
Fewer shades of gray with greater differences between light and dark areas.