scintillation detectors

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/52

encourage image

There's no tags or description

Looks like no tags are added yet.

53 Terms

1
New cards

ionizing rad deposits

energy into scintillator

2
New cards

radidation particle interacting in scintillator results in

one scintillation (flash) of light and, therefore, one p

3
New cards

the greater the energy transferred to scintillator, the greater…

the number of excited electrons, and the greater the number of photons of light emitted

4
New cards

decay time is time required for

emission of 63% of light photons produced in scintillation

shorter the decay time, the better (mostly)

5
New cards

ideal scintillator: probability of radiation interacting w scintillator should be

high

6
New cards

ideal scintillator: a large portion of absorbed rad energy should be

emitted as light

7
New cards

ideal scintillator: linear proportionality…

should exist bw energy deposited and light produced

8
New cards

ideal scintillator: emission of light should occur…

as fast as possible

9
New cards

ideal scintillator: scintillator emission spectrum should…

overlap the photomultiplier tube absorption spectrum

10
New cards

ideal scintillator: material should be ___ to photons emitted by scintillator

transparent

11
New cards

ideal scintillator: absorbs all energy of…

every photon entering detector volume

12
New cards

scintillating material might be

inorganic, organic

13
New cards

scintillating material can be what of three things?

gas, liquid, solid

14
New cards

inorganic scintillators: splitting energy levels

greater number of atoms (heavier element) the greater the splitting of energy levels

15
New cards

(band formation) when many atoms combine to form a solid…

valence shell has split so many times that it forms a continuum of energies, i.e., a band

16
New cards

valence band

highest energy band occupied by electrons

17
New cards

band gap

electrons cannot possess energies here; unless there are impurities or defects

18
New cards

insulator

valence band full; electrons immobile and material doesn’t conduct electricity

19
New cards

conductor

valence band not full; electrons mobile and material conducts electricity

20
New cards

semiconductor

valence band is full; electrons are immobile and material does not conduct electricity

21
New cards

difference bw insulator and semiconductor

the width of the band gap

22
New cards

insulator band gap is

greater than 5 electron volts

23
New cards

semiconductor band gap is

less than 5 electron volts (one eV is typical)

24
New cards

light emitted by insulating material wo an activator is not…

detected by photomultiplier tube as efficiently as light of a longer wavelength (lower energy)

25
New cards

to reduce energy of emitted light…

an activator impurity is added to the crystal

26
New cards

an activator is an impurity such as…

thallium

27
New cards

sodium iodide scintillation detectors

detection of gamma rays/xrays

gamma spectroscopy

exposure rate measurements (uR meter)

28
New cards

NaI crystal activated w

with thallium

29
New cards

NaI has extremely high…

light yield; ca. 12% of energy deposited by gamma ray or xray is emitted as light

30
New cards

NaI has long

decay time (230 ns), limits effectiveness at high count rates

31
New cards

organic scintillators made of

aromatic hydrocarbons

32
New cards

organic crystals: only two have been used to any extent, which are

stilbene, anthracene

33
New cards

plastic scintillators (organic)

fluor (3% vol) is mixed w plastic solvent (97%) that is then polymerized

34
New cards

most commonly employed plastics in plastic scintillators

polystyrene

polyvinyl toluene

35
New cards

advantages of plastic scintillators

manufactured in any size/shape, rugged, light, nonhygroscopic, flat energy response

36
New cards

disadvantages of plastic scintillators

not suitable for gamma/xray, low atomic number means more likely for compton

not super detailed data on energy

37
New cards

scintillators must be protected from

ambient light, sometimes specific types of radiation

38
New cards

if scintillator is hygroscopic, must be housed in

hermetically sealed container

39
New cards

when scintillator and PMT cannot be in physical contact due to hermeticallly sealed,

housing must have transparent window to transmit emitted light

40
New cards

well detector

used to maximize counting efficiency for small samples;

41
New cards

well detectors limited to

small samples; therefore sample count rate might still be low

42
New cards

optical coupling

window of scintillator housing is direct contact w photocathode of PMT

optical coupling grease or gel almost always applied

43
New cards

photomultiplier tube converts

each flash of light from scintillation crystal to electronic pulse

usually round in cross section but might be square if multiple tubes packed close together

44
New cards

PMTs are

fragile, affected by magnetic fields

45
New cards

photocathode

thin metallic deposit inside window of PMT; usually flat and circular

absorption of light photons knocks electrons off photocathode

46
New cards

20-30% light photons incident on

photocathode, resulting in freeing electrons

47
New cards

dynodes

attract and accerlerate electrons

48
New cards

scintillators are essentially

the photon detectors

49
New cards

if PMT in magnetic field

change the paths of electrons, particularly those near photocathode end of the tube

50
New cards

liquid scintillation

used to measure activity (mostly beta/alpha) in liquid samples

very high sample efficiencies for counting (nearly 100%)

51
New cards

liquid scintillation: wipe samples placed in empty vial and

filled with “cocktail” fluid

this converts KE of particles into visible light that is detected by PMT of counter

52
New cards

the higher the energy of the nuclide…

the further it travels in liquid → more light produced

53
New cards

zinc sulfide (ZnS (Ag))

primary application in alpha detection; only available as very fine powder