Compressed gas supply; anesthetic vaporizer; breathing circuit; scavenging system (blue tube and collection system)
2
New cards
O2 tank sizes
E tank = 660 L; H tank = 6900 L
3
New cards
Liters left in an E tank
PSI ร 0.3
4
New cards
Liters left in an H tank
PSI ร 3
5
New cards
Minutes of anesthesia left in a tank
Liters left รท O2 flow rate (L/min)
6
New cards
Compressed gas colors
O2 is green in the US (white in Canada); nitrous oxide is blue; CO2 is grey
7
New cards
Pin index safety system
On E tanks; prevents the wrong gas from being attached to the machine
8
New cards
Opening and closing a tank
Open: turn the valve counter-clockwise. Close: turn clockwise
9
New cards
O2 cylinder hazards
O2 is combustible (fire hazard); cylinders are under pressure and can cause corneal lacerations and skin injury
10
New cards
Minimizing O2 cylinder risks
No ignition sources near cylinders; shield during electrocautery or laser procedures; use a wrench to open or change tanks; strap cylinders vertically to the wall; use valve caps and hand carts; wear eye protection
11
New cards
Tank gauge
Shows the pressure in the compressed gas tank (PSI)
12
New cards
Pressure-reducing valve
Reduces tank pressure to 40โ50 PSI; works passively; a line pressure gauge may sit between it and the flowmeter
13
New cards
O2 flowmeter
Measures O2 in L/min. O2 enters at the bottom and exits at the top; further reduces pressure to 12โ15 PSI. Handle the valve gently. If it is off, the patient is not getting O2
14
New cards
O2 flush valve
Delivers pure O2 straight to the breathing circuit, bypassing the flowmeter and vaporizer. High pressure (50 PSI) and high flow (35โ75 L/min), so use with care
15
New cards
Order O2 flows through the machine
Compressed gas tank โ tank gauge โ pressure-reducing valve โ O2 flowmeter โ vaporizer
16
New cards
Vaporizer color coding
Each vaporizer is calibrated to one agent: isoflurane = purple, sevoflurane = yellow, halothane = red
17
New cards
Filling the vaporizer
Use pour-spouts or nozzles; wear gloves and a mask; fill between the upper and lower lines of the indicator window
18
New cards
Range of vaporizer settings
MAC ร 1 = light Stage III; MAC ร 1.5 = moderate; MAC ร 2 = deep
19
New cards
VOC vaporizer
Vaporizer-out-of-circuit: not inside the breathing circuit. O2 from the flowmeter goes through the vaporizer before entering the circuit. These are precision vaporizers
20
New cards
VIC vaporizer
Vaporizer-in-circuit: O2 goes from the flowmeter straight into the circuit, and exhaled gases pass through the vaporizer. These are non-precision vaporizers
21
New cards
WAGs โ sources
The anesthetic machine, gas exhaled by the patient, vaporizer filling, liquid anesthetic spills
22
New cards
WAGs โ hazards
Eye and skin irritation, headaches and dizziness, decreased fertility, cardiovascular and respiratory effects, nausea and loss of appetite
23
New cards
Minimizing WAG exposure
Use a scavenging system; minimize leaks; avoid mask and chamber delivery when possible; avoid spills; wear PPE when filling; cuff the ET tube; use a dosimeter to monitor
24
New cards
Breathing circuit โ functions
Delivers anesthetic mixed with O2 from the fresh gas outlet to the patient and carries expired gases away. Can be rebreathing (circle) or non-rebreathing
25
New cards
Rebreathing circuit โ how it works
The CO2 absorber canister (soda lime) removes CO2 from exhaled gas, so the gas can be recycled and inhaled again with added fresh gas. This allows lower O2 flow rates
26
New cards
Non-rebreathing circuit โ how it works
No CO2 absorber; gas cannot be recycled. High flow rates push waste gas away from the patient
27
New cards
Choosing a breathing circuit
Based on patient size, cost, control of anesthetic depth, conservation of heat and moisture, and waste gas production. The choice sets the equipment needed and the O2 flow rate
28
New cards
Circuit by patient size
Rebreathing: over 7 kg (15.4 lb); patients 3โ7 kg need pediatric breathing tubes. Non-rebreathing: under 3 kg (6.6 lb)
Breathing tubes (Y tubing or F tube), unidirectional valves, CO2 absorber canister, pressure manometer, reservoir bag, pop-off valve, negative pressure relief valve
31
New cards
Breathing tube styles
Y-tubing: corrugated tubes with a Y connector. Coaxial F-tube: inspiratory (inner) and expiratory (outer) tubes combined
32
New cards
Unidirectional valves
Keep gas flowing one way; they open and close as the patient breathes
33
New cards
Negative pressure relief valve
Also called the air intake valve; lets room air into the circuit if negative pressure is detected
34
New cards
CO2 absorber canister
Contains calcium hydroxide granules that react with CO2 to form calcium carbonate. Replace when depleted: color changes from white to violet, after 6โ8 hours of anesthetic time, or when granules become brittle
35
New cards
Pressure manometer
Shows the pressure inside the breathing circuit (cm H2O, mm Hg, or kPa). Watch it when bagging the patient; it should never go over 20 cm H2O
36
New cards
Pop-off valve (open)
Lets excess gas leave the circuit and go to the scavenger; prevents too much pressure or volume in the circuit
37
New cards
Pop-off valve (closed)
Builds pressure in the circuit so you can manually ventilate or do a machine pressure check. Reopen it immediately afterward
38
New cards
Scavenge system
Carries waste anesthetic gas away from the breathing circuit (controlled by the pop-off valve). Types: active and passive. Parts: scavenge tubing and, for passive, a charcoal canister
39
New cards
Reservoir bag โ functions
Flexible gas storage; shows respiratory rate and depth; confirms ET tube placement; lets you deliver anesthetic gas or pure O2 by manual ventilation. Sizes range from 0.5 L to 30 L
40
New cards
Reservoir bag size
Tidal volume is about 10 mL/kg, and the bag should hold about 6 times that: 60 mL/kg of body weight, rounded up to the next liter
41
New cards
Why manually ventilate (bag) a patient
Minimize atelectasis (bag every 5โ10 minutes during a procedure); force fresh gas into the alveoli for gas delivery and exchange; help remove CO2
42
New cards
How to give a breath
Close the pop-off valve, squeeze the reservoir bag 1โ2 times while watching the manometer (under 20 cm H2O), then reopen the pop-off valve
43
New cards
Pressure (leak) check โ steps
1) Open the O2 tank. 2) Close the pop-off valve. 3) Block the patient end of the Y tubing. 4) Set the flowmeter to 2 L/min. 5) Let pressure rise to 20 cm H2O. 6) Turn flow down to 0.2 L/min. 7) Hold for 20 seconds. 8) Reopen the pop-off valve
44
New cards
Common places for leaks
Reservoir bag, soda lime canister, breathing tubing, connectors
45
New cards
Example: reservoir bag for a 40 lb dog
40 lb รท 2.2 = 18.2 kg. 18.2 ร 60 mL = about 1,091 mL (1.09 L), rounded up to the next liter = 2 L bag
46
New cards
Example: rebreather O2 flow for a 40 lb dog
18.2 kg ร 50โ100 mL/kg/min = about 910โ1,820 mL/min = 0.9โ1.8 L/min
47
New cards
Example: tank time on an E tank at 1234 PSI
1234 ร 0.3 = 370 L. At a flow of 1 L/min, that is about 370 minutes (about 6 hours)
48
New cards
Example: vaporizer settings for isoflurane in a dog (MAC 1.3)
1.3% (ร 1), 1.95% (ร 1.5), 2.6% (ร 2). For a cat (MAC 1.63): 1.63%, about 2.45%, 3.26%