noise
random variations in detected signal that do not represent the actual distribution of the RP in the body
noise is primarily caused by
random distribution of photons detected by gamma camera
high noise levels make it difficult to
distinguish bw different tissues/abnormalities, which can lead to less accurate diagnoses
what two factors contribute to noise
statistical nature of radioactive decay, electronic interference
what methods can be used to reduce noise
optimizing exam time
adjusting matrix size
using appropriate settings on imaging device
poisson (quantum) noise
gamma photon detection is random; number of counts collected in a pixel follows poisson stats;
the fundamental noise in all nuc med modalities
poisson noise appears
grainy, āsalt and pepperā texture
uniform regions look mottled
noise =
āN
where N is number of detected photons
relative noise =
āN / N
which =
1 / āN
low counts meansā¦
high noise
high counts meansā¦
low noise; smoother, cleaner images
statistical noise % =
(āN / N) x 100%
or
100% / āN
when you have a bigger matrix with smaller pixels, the % noise per pixelā¦
is larger, because there are less counts per pixel
when there is a smaller matrix with bigger pixels, % noise per pixelā¦
is smaller because there are more counts per pixel
also means less resolution
how can time affect noise for dynamic imaging?
not imaging long enough: more noise bc of less counts
imaging too long: could miss important temporal resolution info
a higher statistical noise is often accepted as a tradeoff for
better spatial res in images;
the fix is to ensure there are more counts if a higher matrix is used.
in theory, the number of counts acquired in an image matrix should be ___ when the dimensions of a matrix are ___
counts would be quadrupled when dimensions are doubled
2 ways to increase counts
increase patient dose
increase time per image
electronic noise
occurs inside gamma camera, can be from PMT stat variations, photodiode noise, preamplifier/ADC noise, electronic drift, thermal noise
characteristics of electronic noise
small random fluctuations, usually low compared to poisson, more evident in extremely low count acquisitions
electronic noise is most noticeable inā¦
very low photon flux situations
electronic noise causes instability in
energy measurements, affecting energy windows
scatter noise
gamma photons scatter in patient tissues, collimator holes, detector crystal
detected w incorrect positions and lower energy, producing āblurā
scatter noise characteristics
appears as low frequency bg haze
reduces contrast
introduces spatial uncertainty
contaminates PET, SPECT
acts as āfogā over image
septal penetration noise
high energy photons penetrate the lead septa bw collimator holes
causes of septal penetration noise
inappropriate collimator for higher energy photon emissions
high photon energy passes through septa instead of hole
imperfect collimator design or damage
characteristics of septal penetration noise
blurred regions, āstarā patterns or streak artifacts; reduced contrast, appears as large scale noise
often w high energy isotopes
attenuation related noise
tissues with high attenuation produce more noise
characteristics of attenuation related noise
regions deeper in body have fewer counts (higher poisson noise)
uneven graininess across field
artefactual cold spots in deeper structures
attenuation correction ____ attenuation related noise
reduces, but does not eliminate
in SPECT/PET, reconstruction canā¦
amplify noise
two types of reconstruction
filtered backprojection (FBP)
iterative reconstruction (OSEM)