Audiometric Transducer Calibration: From Absolute Hearing Threshold to Normalized Threshold

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/15

flashcard set

Earn XP

Description and Tags

Vocabulary practice flashcards covering audiometric calibration, absolute perception thresholds (MAF and MAP), ear resonance acoustics (HRTF and REUG), standard calibration couplers, and RETSPL standards.

16 Terms

1
New cards

dB HL (Hearing Level)

A standardized unit of measurement used in clinical audiometry where 0 dB0\,\text{dB} represents the normal hearing threshold regardless of the transducer utilized.

2
New cards

dB SPL (Sound Pressure Level)

A physical measurement quantifying acoustic energy in terms of sound pressure, reflecting the actual pressure delivered at the tympanic membrane to trigger auditory perception.

3
New cards

MAF (Minimum Audible Field)

The absolute binaural hearing threshold measured in an anechoic chamber with a subject seated 1 m1\,\text{m} from a frontal loudspeaker at 0∘0^\circ incidence, quantified by re-emitting threshold levels and measuring them with a microphone placed at the center of the subject's head position in their absence.

4
New cards

ISO 226-2003

The international standard defining the standardized curve for the human absolute threshold of auditory perception in a free field (binaural MAF).

5
New cards

Monaural MAF Conversion (ANSI S3.6-2004)

The standardized formula used to convert a binaural MAF threshold to a monaural MAF threshold when both ears are symmetrical: MAF(monaural)=MAF(binaural)+2 dB\text{MAF}_{(\text{monaural})} = \text{MAF}_{(\text{binaural})} + 2\,\text{dB}.

6
New cards

HRTF (Head Related Transfer Function)

The acoustic filtering effect produced by the anatomical configuration of the torso, neck, head, and pinna, amplifying certain frequencies and attenuating others before sound enters the ear canal.

7
New cards

REUG (Real Ear Unaided Gain)

Also called unaided ear gain or 'gain étymotique,' it is the natural acoustic resonance and amplification provided by the outer ear structures, peaking between 15 dB15\,\text{dB} and 20 dB20\,\text{dB} around 2800 Hz2800\,\text{Hz} in adult humans.

8
New cards

MAP (Minimum Audible Pressure)

The absolute hearing threshold level calculated at the tympanic membrane by adding natural ear amplification (REUG) to the monaural MAF; it is non-standardized due to anatomical variations across individuals.

9
New cards
<p>TDH Headphone</p>

TDH Headphone

A supra-aural audiometric transducer calibrated for clinical hearing testing using an artificial 6 cc6\,\text{cc} coupler.

10
New cards
<p>Insert Earphones (EAR)</p>

Insert Earphones (EAR)

Audiometric transducers connected to disposable foam eartips that insert directly into the external auditory canal, calibrated using a 2 cc2\,\text{cc} coupler.

11
New cards
<p>6cc Coupler (NBS 9A)</p>

6cc Coupler (NBS 9A)

A standardized metal calibration cavity housing a sound level meter at its base, designed to replicate the average combined volume of the human concha and external auditory canal to calibrate supra-aural headphones.

12
New cards
<p>2cc Coupler</p>

2cc Coupler

A standardized metal calibration cavity housing a sound level meter at its base, designed to replicate the residual volume of the ear canal remaining after the insertion of foam insert earphone tips.

13
New cards

RETSPL (Reference Equivalent Threshold Sound Pressure Level)

The sound pressure level in dB SPL\text{dB SPL} produced in a standard coupler or free field corresponding to the normal hearing threshold (0 dB HL0\,\text{dB HL}) at a specified frequency for a particular transducer.

14
New cards

ISO 389 Series

The family of international standards defining RETSPL calibration values across audiometric transducers, including ISO 389-1 for TDH supra-aural headphones, ISO 389-2 for insert earphones, and ISO 389-7 for free-field presentation.

15
New cards

RETSPL Free-Field

Calibration values established for soundfield testing where calibration is performed using a sound level meter at the patient's listening position without a coupler, directly approximating MAF rather than MAP.

16
New cards

Free-Field Tympanic Sound Pressure Calculation

The formula used to compute the sound pressure level reaching the tympanic membrane in a free field: dB SPL (tympan)=dB HL+RETSPL+REUG\text{dB SPL (tympan)} = \text{dB HL} + \text{RETSPL} + \text{REUG}.