kinetic energy is converted to what for x-rays
electromagnetic energy
the function of an x-ray system is to
provide a controlled flow of electrons intense enough to produce an x-ray beam appropriate for imaging
diagnostic units operate btw what kvp
25-150
why is the bottom number of the kVp range of diagnostic units 25
because of mammo
general purpose rooms usually have what imaging
fluro and radio
the fluro tube usually resides
under the table and fixed
medical x-rays are
diagnostic or therapeutic
the x-ray table is made of
carbon fiber
30 degree tilt on a table is
trendelenburg
90 degree tilt on a table is
reverse trendelenburg/ fowler
3 principle components of imaging system
operating console
xray tube
high voltage generator
operating console is used to
establish conditions of the exposure
the console must display
conditions of exposure
there are controls available to control
major and minor kVp adjustments
mA
Exposure time
all except how many circuits connecting the meters and the controls are low voltage to protect hazardous shock
one
a schematic drawing is
not a picture of what it looks like its a picture of what it acts like
what is line compensation used for
to provide consistent voltage to the autotransformer
technical factors must have very _______ voltage
precise
input voltage can vary as much as
5%
vacillation means
theres not constant voltage
Line compensation controls what
the # of turns on the primary side of the autotransformer
what are the 2 main circuits
main and filament
the main circuit does what
modifies incoming voltage to produce x-rays
what does the main circuit consist of
line compensator, autotransformer, kVp controls, High voltage transformer, rectifiers, tube
what does the filament circuit consist of
wire compensator, autotransformer, mA selector, filament transformer,
the kVp meter is connected to the what terminals of the autotransformer
output
the mA controls mA by controlling the
temperature of filament
mA selector is
a bank of varied resistors
filaments operate on what A
3-6 A
the transformer voltage is
fixed
if the transformer voltage is fixed how do we raise current
decrease resistance
what are the three exposure timers
synchronous timers, electronic timers, mA timers, and the AEC
what is the Hz of american power
60
electronic timers allow for how long exposures
1ms
mAs timers are found where
portable machines
mAs timers do not work with
AEC
Three main parts of a high voltage generator
high voltage transformer
filament transformer
rectifiers
high voltage generators are usually immeresed in
oil
the turns ratio for a high voltage generator is
500:1 or 1000:1
the waveform from the secondary side will differ from the waveform on the primary side in what
amplitude
high voltage generators must have what to function
alternatic current
rectifiers do what
convert AC to DC
what is a rectifier
device which allows electrons to flow in only one direction
the 2 types of rectifiers
vacuum tube and solid state
Half wave rectifers usually have how many diodes
2 or none
single phase full wave rectifiers have how mant rectiifers
4 or more
generator types
single phase
three phase
high frequency
Above the 0 line of the graph of rectifiers means the electrons are flowing to what
toward the anode
single phase half wave rectified has how many pulses per second
60
single phase full wave rectified has how many pulses per second
120
the ripple effect is
the measure of efficiency of the system and the lower the # the more efficient
what is the voltage ripple of half wave
100%
what is the voltage ripple of full wave
100%
why are both half wave and full wave 100%
they fall 100% of the way back to 0 on the graph
what is the voltage ripple of three phase 6 pulse
13-14%
what is the voltage ripple of three phase 12 pulse
4%
what is the voltage ripple of high frequency
<1%
three phase power has how many waveforms
three
what makes three phase power more efficient
while one wavelength goes down, the other goes up
high frequency consists of
full wave rectified converted to high frequency
what does high frequency convert 60 Hz to
500-25,000
what chops DC current into smaller square pulses
inverters
what makes high frequency so great
cheaper, smaller, most efficient
the high voltage transformer does what to voltage
steps up
the filament transformer does what to voltage
steps down
the filament transformer does what to current
steps up current
the autotransformer can be
step up or step down
the line compensator
adjusts # of terms on the primary side of the autotransformer to maintain constant voltage
the line compensator essential removes
vasilation
the kVp selection is a
compensator for the secondary side
the resistor bank or mA selector changes resistance to change
Intensity
x-rays are produced in equal intensity in all directions or
produced isotropically
x-rays emitted from any other part other than the window are called
leakage radiation
shielding protects from specfically
leakage radiation
the x-ray tube housing protects from
leakage radiation and shock
the three external components of the x-ray tube are
x-ray tube support
protective housing
glass or metal envelope
proper tube must have leakeage less than or equal to
1mGya/hr (100 mR/hr) at 1 meter
the glass of the glass envelope is
pyrex
metal envelopes are used to prevent
bronzing
the cathode is not always
negative
the cathode is only negative
at the time of exposure
the space charge effect limits
thermionix emission by electrostatics
the saturation current is
the max electrons boiled off at a certain mA and the highest mA is the tube saturation
the anode is only positive at
exposure
the anode serves three functions which are
thermal radiator
conducts electrons
provides mechanical support for target
higher tube currents and shorter exposures times are possible with which anode
rotating
the target is made of
tungesten rhenium
the line focus principle states
a purposeful angle at the target to maximize area of bombardment and minimize effective focal spot size
the effective focal spot size is also known as
projected/apparent focal spot
as the target angle becomes more vertical, the effective spocal spot size
decreases
if the effective focal spot size decreases, the spatial resolution
increases
the anode heel effect is due to
line-focus principle
what three things cause the anode heel effect to manifest
large field size
short SID
large focal spot
the anode heel effect technically occurs by
x-rays attenuating on the anode side due to going through the target to the angled angle
the effective focal spot is smaller towards the
anode
extra focal radiation is
x-rays produced outside the focal spot
extra focal radiation is also called
off focused radiation
what are the 4 problems caused by extra focal/off focused radiation
extends size
increases skin dose
reduces image contrast
exposes unintended tissues
what helps minimize off focus radiation
diaphragm and window
what is the primary cause of tube failure
heat