image qualitiy
refers to the accuracy with which the anatomical structure that is being imaged is rendered on the radiograph
high quality radiograph
radiograph that faithfully reproduces structure and tissues is identified as a high quality radiograph
most important characteristics of radiographic image quality
spatial resolution
contrast resolutin
noise
detective quantum efficiency (DQE)
Resolution
the ability to image two separate objects and visually distinguish one from the other
spatial resolution
ability of the imaging system to resolve fine details presentin an object; or to differential separate objects of high subject contrast that are adjacent to one another
contrast resolution
ability of a detector to distinguish between differences in image intensity for structures with similar subject constrast
spatial resolution is measured in terms of
spatial frequency, in units of line-pairs per millimeter
if pixel size is 0.1 mm spatial resolution is?
about 5 LP/mm
digital imaging has what compared to film
poorer spatial resolution
better contrast resolution
what does spatial resolution also refer to?
the sharpenss of the image
smaller pixel = __ resoltuion
better
__ matrix and _ pixel = better resolution
larger matrix
smaller pixel
contrast
the difference in densities between various areas on a radiogrpah
high contrast = few shades of gray between black and white
low contrast = demonstrate more shades of gray
what is contrast resolution determined by
the absorption of x ray photons or linear attenuation in body tissues
overall resolution is dependent on?
both spatial resolution (sharpenss) and contrast resolution
what is tradeoff for digital imaging
extremely small details with only slight separation between them cannot be resolved
signal
the portion in a radiographic image of the image forming x rays that represents anatomy
noise
any random background information that is detected but does not contribute to image quality
scatter radiation
undesirable x rays reaching the IR
electronic noise (system noise)
results from undesirable signals from the digital system ifself
often through normal function
quantum mottle noise
results form an insuffience quantity of photons from improperly set exposure factors (mAs and kVps)
produces grainy image
if noisy across the image increase mAs
if noisy only in thicker portions of anatomy increase kVps
signal to noise ratio
measure of signal strength relative to total noise
depends on amount of radiation exposure to the detector (signal) and the detectors detective quantum efficiency
contrast resolution is limited by
noise or SNR
detective quantum efficiency (DQE)
meausres the sensitivity and accuracy by which the image receptor converts the incoming data to the output viewing device
digital radiography imaging systems have DQEs of 0.3-0.7 or 30-70%
IRs with higher DQE are?
more effective at stopping x rays and converting them to electronic signal
increased DQE does what to number of x rays
decreases the number of x rays required to produce a diagnostic image and subsequently decreases patient dose
windowing
adjusting the brightness and contrast of the image displayed at the monitor screen with controls at the console as the image is viewed
window level
changes the overall or average brightness of the image
does not change the number of gray levels (gray scale) in the image
window width
controls the length of the gray scale in the image
increasing window width lengthens the gray scale but does not alter the average birghtness of average density
increasing window level does what
makes radiographic image darker
reduces brightness
increasing window width does what
gives a radiographic image more gray scale
increasing window width reduces contrast