1/71
NOT FINISHED
What is included in image display?
Includes all of the system components necessary to convert the digital data created from the reconstruction process to electrical signals needed by the CT display monitor
Includes the ability to display patient information and scan protocol data
Provides many graphic aides designed to assist in image interpolation
What is the cathode ray tube (CRT)?
Basically, a standard television set with some modifications that improve image resolution
What is the newer flat panel type display monitors called?
TFT LCD (thin-film transistor, liquid crystal display)
LED (light emitting diode)
What advantages do the newer display monitors have over the older type
Produce higher luminescence (brightness)
Higher spatial resolution
What is the HU for bone
+1000
What is the HU for water
0
What is the HU for air
-1000
How many shades of gray can be displayed on a medical monitor
1024 shades of gray
What is the contrast difference that can be appreciated by medical monitors
Can appreciate contrast differences of less than 1%
How many shades of gray can be seen by the human eye
40 shades of gray
What contrast difference can be appreciated by the human eye
Cannot appreciate contrast differences of less than about 10%
What is Artificial Intelligence in Radiology and how does it use radiomics
A CT image contains information that goes beyond what we can appreciate visually so AI uses radiomics to extract quantitative image features
It can also learn image patterns directly
What does radiomics mean
Extracting quantitative image features
What are some examples of radiomics
Size
Volume
Density
Shape
Texture
What is window width?
Determines the quantity of Hounsfield units represented as shades of gray on a specific image
All values higher than the selected range appear WHITE
All values lower than the selected range appear BLACK
Increasing the window width (widening the range) = more numbers are assigned to a shade of gray
What is window level?
Selects which Hounsfield values are displayed as shades of gray
Determines the center pixel value in the gray scale
How do you calculate the range of HU’s
Divide the window width by 2
Add that number to the window level to get the top end of the range
Subtract that number from the window level to get the bottom end of the range
What are window settings?
Manufacturer software will have present values programmed in when new equipment is installed
Radiologists and Technologists work together when a facility gets new equipment to make sure settings are what they want them to be and can make changes if desired
If changes are made, they are saved to an external drive so they can easily be reuploaded if needed (major power outage)
What should the window level be set at relative to the tissue
The window level should be set at a point that is roughly the same value as the average attenuation number of the tissue of interest
Use a wide window width if the goal is to see many different shades of gray
What is the ROI tool
Region is defined by the operator/technologist
The tool is most often circular but can be other shapes
What 2 things is the ROI tool used for
Used to obtain an average Hounsfield unit of an area
Used to acquire the standard deviation
What is standard deviation
The amount of CT number variants within the ROI
If the ROI measurement is 20 and the standard deviation is 0, then every pixel value in the ROI is 20
If the ROI measurement is 20 and the standard deviation is 10, then every pixel in that region is NOT 20
The higher the standard deviation, the greater the variation among pixels within the region
What factors produce high standard deviation
mixed attenuation tissue within the ROI
An ROI that includes streak artifacts
An ROI that is not inside the margins of the object being measured
What are distance measurements?
A tool that is a line the user can put from one area to another
Measurements are given in cm or mm*
Important tool to place a biopsy needle or drainage apparatus
What is image annotation?
Software that allows the operator to annotate specific images with words, phrases, arrows, or other markers
What is a reference image?
Displays the slice lines in corresponding locations on the scout image
What is image magnification?
Just makes the existing image appear larger
Uses only image data
Does not change the spatial resolution
Not the same as making changes to the DFOV
What can magnifying the image help the technologist with
Helps to measure a tiny structure like a kidney stone
Helps to correctly place the ROI in the region you want to measure
What is a histogram
A graphical display showing how frequently a range of CT numbers occurs within the ROI
What is multiplanar reformatting (MPR)
Two-dimensional reformation done to show anatomy in various planes
Coronal
Sagittal
Oblique
What type of data is used to create multiplanar reformats
Image data
What kVp range is used in CT
80-140 kVp
What is the average energy of the x-ray beam
70 keV
What is the SI unit of measurement for absorbed energy
Gray
Gy
What is the SI unit of measurement for deposited energy
Sievert (Sv)
What does quality factors mean
Accounts for the different biological effects produced from different types of ionizing radiation
For diagnostic x-rays used in the CT the quality factor is 1
What does effective dose mean
A measurement that attempts to account for the effects particular to the tissue that has absorbed the radiation dose
What data source does effective dose use to determine risk
Japanese atomic bomb survivors
What is KERMA?
Kinetic energy released in matter
The amount of energy transferred from the incident x-ray photons to charged particles per unit mass in the medium
What is the difference from x-ray to CT dose geometry
CT gives a more uniform dose
The beam is heavily filtered as it exits the X-ray tube, which means fewer low-energy (soft) photons remain. Because the beam is harder than that used in radiography, a lower percentage of the beam will be absorbed or scattered as it passes through patient tissue
The exposure in CT comes from all directions
What two things cause dose uniformity to decrease in CT
Increased SFOV
Increased patient thickness
What is radiation profile in CT
Variations along the length, or z axis of the patient
What is the AKA for radiation profile
Z-axis dose distribution
How does scatter affect the neighboring slice and what is it called
It attributes to the dose of 1 slice due to the scatter from adjacent slices
“tail”
What is multiple scan average dose
MSAD
Dose calculated from multiple scans
What is the multiple scan average dose used to figure out
Total dose
The central slice radiation dose plus the scatter overlap or tails
What is computed tomography dose index
Allows an estimate of the MSAD to be accomplished with a single scan
Who is the computed tomography dose index reported to
FDA
When does MSAD equal CDTI
When there is no overlap or gap between slices
What 2 things are needed for CTDI testing
CTDI phantoms
CTDI 100
What is pitch
Relation of table speed to slice thickness
Pitch = 1
The table moves a distance equal to the slice thickness
Slices are contiguous with no gaps, providing optimal image quality
Pitch <1
The table moves less than the slice thickness, causing overlapping slices
This improves image resolution but increases radiation dose to the patient
Pitch >1
The table moves more than the slice thickness, creating gaps between slices
This will decrease radiation dose and scan time but may slightly decrease image quality
What is the most common pitch settings used in CT
1-1.5
Most common is 1.2
What is CTDIw
Adjusts for variation across the scan field of view by providing a weighted average of measurements at the center and the peripheral slice locations ( the x and y dimesions of the slice)
How are the CTDIw doses calculated
2/3 CTDI 100 (periphery) + 1/3 CTDI 100 (center)
What is CTDI100
Result when the CTDI is measured using a pencil ionization chamber
What is CTDIvol
A measure of exposure per slice and is independent of scan length
Takes into account the exposure variation in the z direction
It is the preferred expression of radiation dose in CT dosimetry
How are CTDI doses calculated?
CTDIvol= CTDIw/pitch factor
What is DLP
Dose-length product
the DLP more closely reflects the radiation dose for a specific Ct examination, its value is affected by variances in patient anatomy
How are DLP doses calculated?
DLP=CTDIvol x scan length
Which CTDI calculation is the preferred expression for measuring dose in CT
CTDIvol
Know about pediatric patients in CT
Children are much more radiosensitive than adults
What are specific ways to reduce pediatric dose
Use CT only when clinically indicated
Customize the CT examination
Decrease mAs
Why is patient shielding less beneficial in CT
Because of the narrow collimation, radiation to areas outside of that selected scan area is minimal usually attributable to the internal scattering of photons that are unaffected by surface shielding
Is worker shielding still beneficial in CT
Yes it is still important to shield yourself if you must be in the room watching an injection, the patient is vomiting, or for biopsy procedures
What are the 7 strategies for reducing radiation dose
Reduce mAs (quantity of x-ray)
Use ATCM (automated tube current modulation)
Avoid increasing kVp (affects quanity)
Increase the pitch
Limit the use of thin slices
Limit repeat scans
Use iterative and deep learning reconstruction methods
What is ACTM
Automated tube current modulation
A software that allows the computer to estimate the dose a patient needs based off of their scout images
Reduces the amount of radiation b7 35 to 60%
How does increasing the pitch affect reducing radiation dose
From 1 to 1.5 decreases the dose by 33% without loss of information
How does limiting the use of thin slices reduce radiation dose
Using a large number of thin adjacent slices resulting in 30-50% more radiation dose to the patient than using fewer thicker slices to scan the same anatomy
How does the use of iterative and deep learning reconstruction methods reduce radiation dose
Reduces the dose by as much as 50%