puretone testing
quantify auditory access to each frequency needed for speech perception
ear specific information
helps us predict communication struggles (speech banana and articulation index)
used to prescribe hearing aids, cochlear implants or other strategies
below line
where is auditory access on this audiogram? meaning where can the patient hear

audiometer
device used for testing hearing loss (screening and diagnostic purposes)
two channels allow for testing and masking (you can put different sounds in each ear independently)
controls of audiometer
includes stimulus, transducer, and routing
puretone, live voice, recorded speech, noise, or other
what is the stimulus control on an audiometer?
supra-aural headsets, inserts, speaker, bone oscillator
what are the transducers in an audiometer?
a transducer is a device that converts electrical energy from an audiometer into energy to test hearing
right, left, and both
what are the routing controls in an audiometer?
supra-aural earphones/headphones
a type of transducer that is easy to clean and is familiar to patients
cons: collapsed canal, more crossover of sound, standing wave at 8k (distance from earphone to eardrum can cause wave to cancel itself out), sudden drop at 8k…think standing wave

circum-aural earphones/headphones
a type of transducer that is less constricting and more comfortable, typically used for higher frequency testing above 8000 hz, usually for tinnitus patients
cons: sound leaks around headset, less accurate results

insert earphones
a type of transducer with no standing wave, less masking needed, and no collapsed canals
cons: disposable (cannot reuse and costs more money), can’t use with drainage

bone oscillator/conductor
a type of transducer that is placed on the mastoid bone but NOT touching the pinna (placement is very important), can test on forehead if not fitting on bone
pitfalls: may fall off easily, move during testing

speaker
a type of transducer that is in a booth, which can be used with difficult patients/babies, useful for functional benefits of hearing aid/cochlear implant
cons: no ear specific info (can’t tell right vs left)

attenuators
is a device that reduces the power or amplitude of a signal without significantly distorting its waveform
equipment that measures decibels and frequency
decibels attenuator
measures loudness
-10 dB HL to 120 dB HL (some equipment only goes to 110 dB HL)
frequency attenuator
measures pitch
125-8000 Hz (some equipment has high frequency capability to 12,000 Hz)
calibration
annual assessment
computer based equipment tends to be more stable
portable equipment highly susceptible to errors
correction factors
dB HL to add/subtract per frequency
sticker placed on audiometer
can be different for each transducer
are based on certain frequencies
testing environment
quiet room for screening, diagnostic in sound treated room (not soundproof), is a double walled booth with sound absorbing material between two metal walls and walls have small holes to expose absorption materials
the floor is carpeted with padding and is raised off the ground to reduce vibration noise
double doors with thick soft molding to reduce infiltration of outside noise
fire alarm/sprinklers
lighting specially insulated to reduce outside noise infiltration
ventilation - needs to be very quiet system
no
is a sound treated room soundproof?
so patient can’t see audiologist and the buttons they are pressing
why is there a specially insulated window between patient and tester sides of a testing environment booth?
order of audiologic evaluation
case history, otoscopy, tympanometry (acoustic reflexes, otoacoustic emissions), pure tone testing, SRT, WRS, and bone conduction
pre-evaluation
file review, case history, self rating scales (assess impact and readiness to use hearing aids), conversation (pay attention to patient when conversing), malingerers, assess communicative function, and otoscopy after pre-assessment
testing
seating - eye contact or no? no eye contact so patient doesn’t think to press button but important to see if patient is uncomfortable
instructions - be clear, concise, adjust to patient (stick to 6th grade reading level)
response mode - hand raise, button, game, yes/no, nod
transducer selection and placement - supra-aural headsets, inserts, or bone oscillator
threshold search
finding the softest sound that the patient can hear
includes tone presentation, either an ascending technique or descending technique, combination approach, threshold criteria, and frequency progression
happens in a booth
tone presentation
give one presentation at dB SL to alert (50 dB)
all presentations are 1-2 second duration
pure tone, warble or pulsed tones
vary wait time/presentations (don’t get into a rhythm)
ascending technique
start at 0 dB HL and work up in 10 dB steps
typically used for hard to test patients and children
descending technique
start at alerting tone, work down in 10 dB steps
used more often
combination approach
give alerting tone and work down in 10 dB steps, once patient misses response, go up in 5 dB steps to find threshold
threshold criteria
respond 50% of the time
2/3 presentations elicit response
frequency progression
start at 1k, then 2k, 3k, 4k, 6k, 8k Hz
return to 1k (reliability check) then 250, 500 Hz
more than 20 dB difference between frequencies, add more interoctave testing (750 Hz, 1500 Hz)
calculating pure tone average
add up thresholds at 500, 1k, 2k Hz and divide by 3
15 dBs
how many dBs between the right and left ear signals asymmetry?
500, 1k, 2k, and 4k Hz
what frequencies (Hz) are the most important to test since they most resemble speech sounds?
no response to 1k Hz tone
increase by 20 dB HL, and try again
switch to 500 Hz at some dB HL and try again
reinstruct
false positives
when a patient says they hear the sound even if they don’t
could be because they are eager to please, ringing, nervous, afraid to fail
reinstruct, change waith/off time, change Hz
use another signal (pulse, NBN, warble)
false negatives
this is when patients feel like they have to be sure and want to be 100% sure they heard the noise
could be from being afraid, confused, malingering, ADD/ADHD
reinstruct, change signals, give reassurance of correct responses
pediatric testing
use second tester to assist (define roles and set up communication)
try to get low frequency and high frequency info for each ear (500 Hz, 4000 Hz)
need to move quickly to keep attention
use reinforces like verbal/social praise, food, stickers, games, etc
variations in results
5 to 10 dB HL difference is normal
comparison of serial audiograms can show slow progression vs normal fluctuations
psychoacoustics
auditory access and perception
may affect results: attention to task, state of alertness, motivation, anxiety, barometric pressure, recent noise exposure
pure tone average (PTA)
500 Hz + 1k Hz + 2k Hz / 3
may not reflect slope/configuration of loss
may underestimate the communicative impact of loss
tactile responses
low frequencies are affected (are felt instead of heard), does not affect frequencies above 1000 Hz
transducer contact (bone conductor touching pinna or glasses, earphones touching glasses)
profound hearing loss (bone conduction and air conduction - can look like false air-bone gap)
useful for “teaching” to respond by placing BC in hand to feel
bone conduction testing
threshold search: combination of ascending/descending technique, alerting tone, down 10 dB, up 5 dB, 50% or 2/3 correct responses
unmasked thresholds could be “better ear” response - need to use masking for those
1k, 2k, 3k, 4k, 250 and 500 Hz
what is the frequency progression of bone conduction testing (different from air conduction)
placement options during bone conduction testing
could be placed on the mastoid bone or forehead
mastoid bone placement
placement during bone conduction testing
pros: less dB to elicit response, more likely test ear
cons: can fall off easily, can move, might touch pinna, patient has to remove glasses to avoid contamination
forehead placement
placement during bone conduction testing
pros: easier to place on children, does not touch pinna, does not interfere with glasses
cons: needs higher dB to elicit response (need to use correction factors), localizes to better ear
masking
intentional use of a stimulus to eliminate participation of the non-test ear, second signal that interferes with the primary (test) signal and causes a shift in threshold (distract better ear)
2 channel audiometer required
asymmetrical
is masking used when there is symmetrical or asymmetrical loss?
masking stimulus
includes speech or noise
speech: babble, nonsense, speech weighted noise
noise: narrow band noise, white noise, pink noise, or fresh noise
effective masking
interaural difference of 40 dB (earphones)
interaural difference of 50 dB (inserts)
just enough to occupy better ear
threshold of better ear (non-test ear) +10 dB HL
(when sound crosses over from one ear to the other, since better ear helps the bad ear)
undermasking
insufficient masking, better ear still contributing
overmasking
excessive masking noise crosses over into test ear
overestimates the hearing loss in the test ear
occlusion effect
is basically when we close our ears and sound is amplified
decrease (improvement) in bone conduction thresholds at 1k or lower frequencies because of tightly fitting earphones
seen only in normal listeners and sensorineural losses
(not conductive losses)
masking dilemma
bilateral conductive hearing loss
any attempt to mask the non-test ear, crosses over into test ear bone conductor thresholds; creates overestimate of hearing loss
speech testing
measures communication skills (quiet vs noise and S/N ratios), site of lesion diagnosis (speech recognition scores), reliability check (consistency/accuracy), and compromised patients (young children, developmental disabilities, and dementia)
intelligibility
audibility ≠ what?
speech awareness threshold (SAT)
also known as speech detection threshold (SDT), can detect speech but can’t understand words
±10 dB from pure tone average
patient can detect speech happening but doesn’t understand
speech recognition threshold (SRT)
also known as speech reception threshold, where patient can hear and understand words
50% accurate word recognition using spondee words
should = pure tone average ±10 dB HL (can be influenced by configuration of hearing loss)
used to determine dB SL for discrimination testing
speech awareness threshold (SAT)
includes any speech material (BaBaBa, name, counting, ABCs, reading passage), tends to be 10 dB below SRT (can be due to a single frequency), use ONLY in patients who cannot give SRT (young children, profound deadness, cognitive disabilities, foreign language student), and ascending/descending technique (50% accuracy)
speech reception threshold (SRT)
uses spondee words (words that are familiar, closed set - knows they’re coming or open set - doesn’t know), transducer (inserts, earphones, speaker, bone conductor), signal (either live voice or recorded materials), and routing can be left ear, right ear, or both ears
spondee words
2 syllable words with equal stress on both syllables (like hot dog, baseball, airplane, etc)
go down 10 dB till patient can’t hear (should be close to pure tone average) which is around the softest they are able to hear
50%
what percent accuracy is needed for accurate word recognition using spondee words on an SRT?
live voice presentation
uses a VU meter
pros: can control rate of presentation, local/native speech patterns, can use non-standardized materials, and can repeat items (not standardized procedure)
cons: hard to keep voice within ±3 dB HL, tendency to overarticulate, and accent/articulation issues
VU meter
equalizes voice so doesn’t go above presentation level
can be a light or needle
O (on the meter) = exactly correct dB presentation
stay within ±3 dB HL for presentation
recorded materials
use calibration tone to set VU meter (dial to set at 0)
pros: keeps presentation level very steady, more standardization (less variability) across patients, no over articulation, and smooth rate of presentation
cons: presentation pace often too fast/slow and some materials may not be available
intelligibility testing
is a test where a word list is presented to a patient and they have to repeat the exact word to get correct
is presented as a percent correct score where word recognition score/word discrimination score
presented as dB SL: SRT + 35 dB HL (or 40)
monosyllabic words with carrier phrase like “say the word…” and are not spondee words
can be open set or closed set
open set
in intelligibility testing where there are phonetically balanced word lists and phonemes are proportionate to typical english language
no idea what word will be presented
closed set
in intelligibility testing used for patients who cannot repeat words and is easier to pick the right word from a set
patient knows what words are coming
open set lists
includes PBK, CID W-22, and NU 6
PBK
is a type of open set list that is normed on kindergarten children
easy word lists (mostly nouns/verbs)
for children and special needs patients
CID W-22
is a type of open set list that is phonetically balanced and contains adult word lists (can use with older children too)
NU 6
is a type of open set list where there is a consonant, nucleus, consonant with harder vocabulary
is used for adults
categories of performance
can be scored as poor, excellent, good, or satisfactory
relevance depends on test materials (open vs closed set, difficulty of word list, and how it compares to real life vocab)
booth effect (sound treated room with no distractions, patient focused on testing)
presentation levels
includes sensation level and conversational level
sensation level
typically 35-40 dB above SRT
noted as + ____ dB HL or as dB SL
provides discrimination accuracy at sufficient volume above threshold
may be inadequate depending on hearing loss configuration (cookie cut, ski slope, etc)
conversational level
is presented at 50-55 dB HL
helpful in looking at real world communication skills
may help person see need for hearing aids
masking for speech testing
noise choice could be speech babble, multi-talker babble, speech weighted noise (speech spectrum noise), running discourse, or white noise
dynamic range
is the range of threshold to loudness discomfort level (shorter range for patients with hearing loss), threshold to discomfort
includes MCL and LDL
reduced range is problematic (ex: SRT = 60 dB HL and LDL = 100 dB HL… only gives you 40 dB HL to work with)
MCL
a type of dynamic range that is the most comfortable listening level
patient’s subjective opinion
can be done with pure tones, but primarily used for speech testing
LDL (UCL/ULL)
a type of dynamic range where patient’s subjective opinion of uncomfortable level
pure tones, but primarily speech
dynamic range
the decibel range between SRT (speech reception threshold) and LDL (loudness discomfort level)
recruitment
unusual growth in loudness perception (limited distance from threshold to discomfort)
perception changes quickly with limited increase in dB (MCL may be only slightly above SRT, low % scores due to insufficient gain - trying to listen to speech at very soft levels)
creates a small (reduced) dynamic range - tend to do poorly with hearing aids, may not tolerate increased input levels
testing in noise
transducer: insert, earphones, speaker, or bone
routing: ipsilateral or contralateral (one or both ears, same side or opposite)
signal: speech weighted noise, multi-talker babble/cafeteria noise, running discourse
S/N ratio: percent correct at different S/N provides treatment/amplification info, can be used to show need for assistive devices
useful for aided testing (soundfield/speaker)
quantitative information: percent correct at varying levels
qualitative information: slower response rate, hesitation, guessing, request repetition
quickSIN speech-in-noise-test
the patient has to repeat back 6 sentences binaurally
each sentence is played with background noise of different S/N ratios
there are 5 key words the patient has to repeat correctly in each sentence
score is 25.5 - total score of words correct =
hearing in noise test (HINT)
is an open set speech test that is presented in quiet and noise, presentation levels for sentences changes using ascending/descending technique
noise (speech weighted noise) stays stable
uses formula to calculate RTS (reception threshold for sentences)
has normative data
can alter speech and/or noise presentation to simulate different listening situations
can evaluate aided (with cochlear implant) and unaided (natural hearing)
CID everyday sentences
is an open set speech test where 10 sentences are presented, 50 key words within sentences
percent correct of key words = percent correct score
speech perception in noise (SPIN)
is an open set speech test that assesses key word (last word in sentence)
high predictability vs low predictability lists
synthetic sentence identification
is a type of closed set speech test that has 10 nonsense sentences, present competing noise using connected discourse
pick correct sentence from list
must be able to read
word identification by picture identification (WIPI)
is a type of closed set speech test where there are 6 color pictures on a page
words sound similar
normed on 4-5 year olds

nu-chips
is a closed set speech test where there are 4 black/white pictures per page
normed on 3-4 year olds

california consonant test (CCT)
is a closed set speech test where there are 100 test items, pick from 4 choices
highly sensitive to high frequency losses
identifies consonant confusions
ling six sound test
a type of speech test where each sound representative of a frequency range, can help estimate threshold levels, can be used to check aided function, and can show reception or perception

body parts
a type of speech test that can give estimates of speech perception
can use parent, mr. potato head, or self to point to ear/hair, nose/toes, teeth/knee
same/different
a type of speech test with different levels: monosyllabic vs bi-syllabic, spondee words, monosyllabic with diverse vowels, and monosyllabic with diverse consonants
sound effects recognition test (SERT)
technically not a speech test but can give information about frequency perception and pattern perception
taped sounds, pick from four pictures
non-english patients
recorded materials in native language: many tests available in spanish and other
live voice in native language: how accurate is the pronunciations of tester?
use of interpreter/translator: breaks standardization but can give some qualitative information
other considerations
number of test items
fewer items = more variability in results
single word tests designed for 50-word presentation (many audiologists use only 25 words, increases variability between tests)
non-standardized materials: can be misleading to parents, patients, and therapists
lower
are stops and vowels higher or lower frequency?