1/17
Looks like no tags are added yet.
concept of biofeedback
turning muscle activity into visual feedback, auditory feedback
purposes of biofeedback
muscle facilitation or relaxation
electrode
black as ground or common in the middle, red and black on outsides, 2cm apart
placement of electrodes for biofeedback
along muscle fibers - specific to muscle, or across muscle fibers - less specific but helps relax muscle
when you activate a muscle, you also activate
motor units that create voltage
more motor units activated =
more voltage
V1 =
MU activity and noise
V2 =
MU activity and noise
differential voltage measurement
V1-V2 (cancels out the noise)
order of signal processing
raw signal amplification to make larger, rectification to get the absolute value, filtering to get rid of the noise/make smooth, integration of time constant (the feedback we display to the patient)
time constant
how often we give the pt feedback
time constant for muscle facilitation
0.1sec (10x/sec)
time constant for muscle relaxation
0.5sec (2x/sec)
feedback
can be visual, auditory or via NMES activation (use arbitrary goal and use stim to meet the goal if pt cannot)
sensitivty
the ceiling/top of the graph - set at a sensitivity that would offer meaningful feedback to the patient
mode option
continuous or work/rest
application examples for muscle facilitation
TKA, gait, amputees
application examples for muscle relaxation
spasticity/tone, PROM