1/221
Looks like no tags are added yet.
Reactor-produced radionuclides are neutron-rich and decay by
Beta-minus emission
Cyclotron-produced radionuclides are proton-rich and decay by
Positron emission or electron capture
Production method for most PET isotopes
Cyclotron
Two major fission products used in nuclear medicine
Mo-99 and I-131
Reactor process where a stable nucleus captures a neutron
Neutron activation (n,γ)
Neutron activation product is what element relative to the target
Same element (low specific activity)
Fission product is what element relative to the target
Different element (carrier-free)
Does a generator create new radioactivity
No, it separates daughter from parent
Isotopes made with 0–10 MeV hospital cyclotrons
F-18 and O-15
Isotopes requiring 30+ MeV cyclotrons
Sr-82 and Ge-68
Reaction producing F-18
O-18 (p,n) F-18
Reaction producing In-111
Cd-112 (p,2n) In-111
Parent that decays to Tl-201
Pb-201 (T½ 9.3 h)
Immediate parent that decays to I-123
Xe-123
Half-life of In-111
2.8 days (~67 h)
PET isotope requiring an on-site cyclotron due to its 20 min half-life
C-11
Reactor component that slows neutrons to thermal speed
Moderator
Percent U-235 in natural uranium
0.7%
Fission products mnemonic My I-Xe Story Can Act
Mo-99, I-131, Xe-133, Sr-90, Cs-137, Ac-227
Neutron activation reaction producing Lu-177
Lu-176 (n,γ) Lu-177
Mo-99 half-life
66 h
Tc-99m half-life
6 h
Sr-82 half-life
25 days (25.5 d)
Rb-82 half-life
75 seconds
Equilibrium type of the Mo-99/Tc-99m generator
Transient equilibrium
Equilibrium type of the Sr-82/Rb-82 generator
Secular equilibrium
Parent-to-daughter half-life ratio for transient equilibrium
~10–50 : 1
Parent-to-daughter half-life ratio for secular equilibrium
100 : 1
In transient equilibrium daughter activity compared to parent
Slightly exceeds parent, then tracks it
In secular equilibrium daughter activity compared to parent
Approximately equal
Time for Sr-82/Rb-82 to reach equilibrium
~7.5 min
Time to maximum Tc-99m regrowth after elution
~23–24 hours
Percent of Mo-99 activity present as Tc-99m at 24 h
87%
Column material holding Mo-99 in the generator
Alumina (Al2O3)
Chemical form of the Tc-99m eluate
Sodium pertechnetate (TcO4−, +7)
Theoretical generator yield formula
Mo activity now × Tc/Mo fraction
Actual generator yield formula
Theoretical yield × elution efficiency
Typical elution efficiency
85–95%
Elution efficiency with theoretical 550 mCi and actual 475 mCi
86.4%
Causes of low generator yield
Air leak, channeling, radiolysis
Purity type asking is this the right isotope
Radionuclidic purity
Purity type asking is there an unwanted non-radioactive substance
Chemical purity
Purity type asking is the isotope in the right chemical form
Radiochemical purity
Mo-99 breakthrough limit
0.15 µCi Mo-99 per mCi Tc-99m
Mo-99 breakthrough limit applies at what time
Time of administration
Agency setting the Mo-99 breakthrough limit (10 CFR 35.204)
NRC
How often Mo-99 breakthrough is tested
Every elution
Mo-99 gamma energies detected in the moly assay
740 and 780 keV
Why the lead moly shield works
Blocks 140 keV Tc, passes Mo gammas
Usual Tc-99m eluate expiration
~12 hours
Time to reach the Mo limit formula
t = ln(0.15 ÷ ratio) ÷ 0.105
Mo breakthrough: 18 µCi Mo in 300 mCi Tc
0.06 µCi/mCi, passes
Maximum Mo-99 allowed in 756 mCi Tc-99m
113.4 µCi
Aluminum (Al3+) breakthrough limit
10 µg/mL
Test used for Al3+ breakthrough
Colorimetric spot test (aurin tricarboxylic acid)
Al3+ spot test passes when the eluate spot is
Lighter than the 10 µg/mL standard
Effect of excess Al3+ on a bone scan
Liver uptake
Effect of excess Al3+ on sulfur colloid
Clumping, lung uptake
Agent in sulfur colloid kits that binds excess Al3+
EDTA
Rb-82 is an analog of
Potassium
Rb-82 generator daily QC first step
Elute 50 mL saline and discard
Wait time before re-eluting for Rb-82 QC
~10 minutes
Time the Rb-82 QC sample must decay before Sr measurement
≥1 hour
Sr-82 breakthrough limit
0.02 µCi Sr-82 per mCi Rb-82
Sr-85 breakthrough limit
0.2 µCi Sr-85 per mCi Rb-82
Sr-82 calculation from the decayed sample reading
Reading ÷ [1 + (R × 0.478)]
Sr-85 breakthrough calculation
Sr-82 result × R (Sr-85/Sr-82 ratio)
Rb-82 QC: 50 mCi at EOE, R 1.48, sample 0.8 µCi
Sr-82 0.0094, Sr-85 0.014; passes
Dose calibrator detector type
Ionization chamber
Regulation requiring dose calibrator assay of every dose
10 CFR 35.60
1 mCi equals
37 MBq
1 Ci equals
3.7 × 10^10 Bq
Dose calibrator constancy test frequency
Daily
Dose calibrator linearity test frequency
Quarterly
Dose calibrator accuracy test frequency
Annually
Dose calibrator geometry test frequency
At installation and after repair
Acceptance tolerance for all dose calibrator QC tests
±10%
Sources used for dose calibrator accuracy
Co-57, Cs-137, Co-60
Dose calibrator test checking correct reading across activity levels
Linearity
Dose calibrator test checking volume/container effects
Geometry
Failed linearity or geometry is handled by
Correction factor
Failed accuracy or constancy is handled by
Repair
Retention period for dose calibrator records
3 years
Reducing agent in Tc-99m cold kits
Stannous ion (Sn2+)
Oxidation state change of Tc with stannous reduction
+7 to +4
Kit that uses unreduced Tc and no stannous
Sulfur colloid
Sulfur colloid preparation reagents
Acid + sodium thiosulfate, then buffer
Cause of oversized sulfur colloid particles
Over-heating or high Al3+
Why not to inject air into a kit vial
Oxidizes stannous, causes free pertechnetate
What the kit vial label does NOT require
Patient name
Cause of high free pertechnetate in a kit
Oxidized stannous ion (incomplete reduction)
Mixing technique for MAA kits
Gentle inversion, never vigorous shaking
USP chapter governing radiopharmaceutical compounding
USP
Volume to draw formula
Volume = activity needed ÷ concentration now
Tc-99m precalibration factor at 2 h
1.259
Tc-99m precalibration factor at 4 h
1.587
Tc-99m decay factor at 2 h
0.794
Tc-99m decay factor at 4 h
0.630
Tc-99m decay factor at 6 h
0.5
Precalibration factor formula
1 ÷ decay factor