What is meant by technique selection?
Factors used to produce a radiograph.
What are the four prime factors of radiographic technique?
Distance, mA, time, and kVp.
How many volts are in 1 kV?
1,000 volts.
What does mAs control on the film?
Density.
How can you make a radiograph darker?
Increase mAs.
How can you make a radiograph lighter?
Decrease mAs.
Which prime factor controls filament temperature?
mA, the current through the filament in the tube.
What is the minimum temperature for thermionic emission?
2,200 °C.
How many seconds are in 100 ms?
0.1 second.
What is mA multiplied by exposure time in seconds?
mAs.
How do you find exposure time when mA and mAs are given?
Time = mAs ÷ mA.
How do you find mA when exposure time and mAs are given?
mA = mAs ÷ time.
What does kVp stand for?
Kilovolt peak.
What governs X-ray penetration of a body part?
kVp.
What determines the kV used for a body part?
Thickness.
Which prime factor controls the scale of radiographic contrast?
kVp.
What does SID stand for?
Source-to-image receptor distance.
Which two radiologic exams use a 72-inch SID?
Chest and cross-table lateral cervical spine.
Which two radiologic exams use a 40-inch SID?
Hand and foot examinations.
What is a teleroentgenogram?
A radiograph taken at a 72-inch SID.
How do radiopaque and radiolucent areas differ?
Radiopaque areas stop X-rays and appear white; radiolucent areas transmit X-rays and appear dark.
How do areas of bone appear on a radiograph?
White.
What kV values are recommended for the skull, extremities, abdomen, and PA chest with grid?
Skull: 80; extremities: 55-70; abdomen: 70-80; PA chest with grid: 110.
What are the four qualities of a radiograph?
Density, contrast, recorded detail, and distortion.
What describes radiographic blackening, and what primarily controls it?
Density; mAs is the primary control.
What describes visible density differences between adjacent structures, and what primarily controls it?
Contrast; kVp is the primary control.
What are the two types of distortion?
Shape and size distortion.
What is recorded detail?
How accurately the object is recorded.
What is the range of exposure factors that produces a diagnostic radiograph called?
Exposure latitude.
What two qualities must be balanced for a good-quality radiograph?
Visibility and sharpness.
What is radiation exiting the tube outside the window called?
Leakage radiation.
What is the exit beam, and what information does it carry?
The remnant beam; it carries patient information to the image receptor.
What scale of contrast results from changing from nonscreen to screen exposures?
Short scale / high contrast.
How much does contrast change when switching from a slow-speed to a higher-speed screen?
Little.
How do resolution and recorded detail change when switching to a higher-speed screen?
Both become worse.
Which option produces the greatest resolution?
A cardboard holder.
Which screen produces the highest contrast?
A high-speed screen.
Which screen helps alleviate motion?
A high-speed screen.
How should technique change when converting from high-speed to slow screens?
Increase mAs fourfold to maintain similar film density.
How should technique change when converting from par screens to high-speed screens?
Halve mAs to maintain similar film density.
How should technique change when converting from screen to cardboard technique?
Increase technique to maintain similar film density.
What type of kVp produces more high-energy x-rays?
High kVp.
How many volts equal 50 kVp?
50,000 volts.
Which technique factor has the greatest effect on contrast?
kVp.
Which technique factor has the greatest effect on density?
mAs.
Which technique factor has the greatest effect on scatter production?
kVp.
What improves recorded detail according to the technique guide?
A long SID.
What cannot be selected when using a falling load generator?
mA. This question was marked "omit" on the key.
What SID is recommended for extremities?
40 inches.
What SID is recommended for a chest x-ray?
72 inches.
What SID is recommended for a cross-table lateral cervical spine?
72 inches.
What SID is recommended for a KUB?
40 inches.
What happens to radiation intensity when SID is doubled?
It decreases to one-quarter of the original intensity.
How does patient absorption change as kVp increases?
It decreases.
How does higher kVp affect the scale of contrast?
It produces a long scale of contrast.
How does higher kVp affect the recorded density range?
It produces a broad range with many grays.
How does higher kVp affect scatter production?
It increases scatter production.
How does higher kVp affect patient radiation dose?
It decreases patient dose.
How does higher kVp affect the number of high-energy x-rays produced?
It increases their number.
How does exposure time change as screen speed increases?
It decreases.
What is the 15% rule?
Increasing kV by 15% doubles density.
What two components comprise radiographic contrast?
Film contrast and subject contrast.
How can magnification be minimized?
Use the shortest OID and longest SID.
How much scatter can a typical grid absorb?
90%.
Where is a grid positioned?
Between the patient and image receptor.
What is meant by a high-quality radiograph?
A variation in densities that allows details to be visualized.
What is the advantage of using intensifying screens?
They reduce patient dose and allow lower mAs settings.
What focal spot size produces better recorded detail?
A small focal spot.
What do technique charts generally specify?
kVp, mAs, and patient thickness.
Who creates departmental technique charts?
Radiologists and technologists in the department.
What is fixed in a variable-kVp technique chart?
mAs.
What varies in a fixed-kVp technique chart?
mAs varies according to patient thickness.
What is a disadvantage of a fixed-kVp technique chart?
Increased scatter production.
Why are technique charts used?
To obtain the best image the first time.
What are the primary exposure variables?
mA, exposure time, and kVp.
What is the advantage of using short exposure times?
Reduced patient motion blurring.
What is anatomically programmed radiography?
A computerized system selects exposure settings for various exams.
What factors must the technologist consider with automated exposure systems?
Patient pathology, size, and positioning.
How many seconds are in 35 milliseconds?
0.035 seconds.
What is the current through the filament in the X-ray tube?
mA (milliamperage).
What is the grid factor formula?
G₂ = G₁ × (mAs₂ ÷ mAs₁).
What is the grid factor when no grid is used?
1.
What is the grid factor for a 5:1 grid?
2.
What is the grid factor for a 6:1 grid?
3.
What is the grid factor for an 8:1 grid?
4.
What is the grid factor for a 10:1 or 12:1 grid?
5.
What is the grid factor for a 16:1 grid?
6.
What type of relationship exists between mA and time?
Inverse (I).
What type of relationship exists between mAs and density?
Direct (D).
What type of relationship exists between distance and exposure rate?
Inverse (I).
What type of relationship exists between distance and density?
Inverse (I).
What type of relationship exists between kV and contrast scale?
Direct (D): increasing kV produces longer-scale contrast.
What type of relationship exists between distance and beam width?
Direct (D).
What type of relationship exists between distance and magnification?
Inverse (I).
What type of relationship exists between EFSS and recorded detail?
Inverse (I).
What type of relationship exists between kV and absorbed dose?
Direct (D).
What type of relationship exists between kV and wavelength?
Inverse (I).
Referring to the H&D diagram on Page 1, which film displays greater latitude?
Film B.
Referring to the H&D diagram on Page 1, which film requires less exposure to produce a density of 2.0?
Film A.
Referring to the H&D diagram on Page 1, which film displays the highest scale of contrast?
Film B.