Scatter Radiation
X-rays that have interacted with an electron of an atom which causes the direction and energy of the x-ray photon to be altered
What does scatter do to contrast?
Decreases it
The 3 things that can happen when an x-ray photon interacts with a patient
It can be absorbed
It can pass through unaffected
It can scatter
2 major controllers of scatter radiation
Beam restriction, Grid
2 factors that affect the amount of scatter radiation produced
kVp, irradiated material
Higher kVp =
Greater the scatter, highly penetrating, less photons being absorbed by the patient, less contrast on image
The amount of scatter created during an interaction is affected by
The volume and the atomic number of the material being irradiated
The greater the volume =
More scatter
Volume increases as…
Field size increases, patient size increases
The higher the atomic mass the greater..
The chance of x-ray interaction due to photoelectric absorption
Less scatter occurs
Bone absorbs more radiation than…
Soft tissue
The larger the field size, the more x-ray photons allowed to interact with the patient which…
Creates more scatter
What should you do with collimation to decrease scatter?
Use the smallest field size possible
The 4 types of beam restricting devices
Aperture Diaphragm
Cones and Cylinders
Collimator
Lead Blocker/Mask
Aperture Diaphragm defintion and characteristics
A flat sheet of metal, usually lead, with a hole cut in the center and attached to the x-ray tube
Simplest to use
Cheapest to buy
Can be made into any size or shape
Disadvantages of Aperture Diaphragm
Different sizes and shapes needed to accommodate different film sizes
Increased penumbra
Penumbra
Geometric unsharpness around the periphery of the image
What causes penumbra?
Due to x-rays not having a “point source”
Penumbra is reduced when…
Beam restrictor is used, using small focal spot
Decrease penumbra =
Increase detail
What radiation does aperture diaphragms cause an increase in?
Off-focus radiation
Off-Focus Radiation
Photons not created at the focal spot but elsewhere in the tube. Produces ghost-like shadow on the radiograph
Determining the diaphragm field size aids in determining….
The projected image on the image receptor
Formula for Diaphragm Field Size
Projected image size = SID × diameter of diaphragm / distance from focal spot to diaphragm
What do cones affect?
Does not reduce penumbra
Does not increase recorded detail
Improves contrast
Only cleans up scatter
What does a cylinder affect?
Reduces penumbra
Reduces off-focus radiation
Cylinder Disadvantages
Difficult to use due to weight
Fixed field size
Cone Field Size Equation
Cone Field Size = SID × lower diameter of cone / distance from focal spot to bottom of cone
What is the most common type of beam restrictor used in radiography?
Collimator
Collimator advantages & disadvantages
More complex and expensive
Allows numerous field sizes
Provides a light source to aid in positioning
A collimator should never be opened greater than…
The size of the image receptor
Upper shutters on collimators reduce…
Off-Focus Radiation
Bottom shutters on collimators reduce…
Penumbra
Positive Beam Limitation
Sensing device that determines the size and placement of cassette in bucky tray. Automatically collimates to size of cassette by moving shutters.
US required by law from 1974 to 1993 that all fixed x-ray equipment have…
Positive Beam Limitation Device
Lead Blockers
Sheet of lead-impregnated rubber that can be cut into any size or shape. Absorbs scatter from large body parts and aids in positioning
Where are lead blockers most commonly used?
Lateral L-Spine, Lateral T-Spine
Lead Blocker affects…
Cleans up scatter
Does not decrease patient dose
Lead Mask
Cut to field size being used, secured at end of collimator, most common in cerebral angiography
Increasing field size also means…
Opening the collimator
Decreasing beam restriction
Increasing field size affects…
Increased scatter
Increased IR exposure
Decreased contrast
Decreasing field size also means….
Closing collimation
Increasing beam restriction
Decreased field size affects….
Decreases beam quantity
Decreases scatter
Increases contrast
Decreases IR exposure
Increase in mAs must accompany a significant reduction in field size to…
Maintain IR exposure