Scatter Radiation

X-rays that have interacted with an electron of an atom which causes the direction and energy of the x-ray photon to be altered

2
New cards

What does scatter do to contrast?

Decreases it

3
New cards

The 3 things that can happen when an x-ray photon interacts with a patient

  • It can be absorbed

  • It can pass through unaffected

  • It can scatter


4
New cards

2 major controllers of scatter radiation

Beam restriction, Grid

5
New cards

2 factors that affect the amount of scatter radiation produced

kVp, irradiated material

6
New cards

Higher kVp =

Greater the scatter, highly penetrating, less photons being absorbed by the patient, less contrast on image

7
New cards

The amount of scatter created during an interaction is affected by

The volume and the atomic number of the material being irradiated

8
New cards

The greater the volume =

More scatter

9
New cards

Volume increases as…

Field size increases, patient size increases

10
New cards

The higher the atomic mass the greater..

The chance of x-ray interaction due to photoelectric absorption

Less scatter occurs

11
New cards

Bone absorbs more radiation than…

Soft tissue

12
New cards

The larger the field size, the more x-ray photons allowed to interact with the patient which…

Creates more scatter

13
New cards

What should you do with collimation to decrease scatter?

Use the smallest field size possible

14
New cards

The 4 types of beam restricting devices

Aperture Diaphragm

Cones and Cylinders

Collimator

Lead Blocker/Mask

15
New cards

Aperture Diaphragm defintion and characteristics

A flat sheet of metal, usually lead, with a hole cut in the center and attached to the x-ray tube

  • Simplest to use

  • Cheapest to buy

  • Can be made into any size or shape


16
New cards

Disadvantages of Aperture Diaphragm

  • Different sizes and shapes needed to accommodate different film sizes

  • Increased penumbra


17
New cards

Penumbra

Geometric unsharpness around the periphery of the image

18
New cards

What causes penumbra?

Due to x-rays not having a “point source”

19
New cards

Penumbra is reduced when…

Beam restrictor is used, using small focal spot

20
New cards

Decrease penumbra =

Increase detail

21
New cards

What radiation does aperture diaphragms cause an increase in?

Off-focus radiation

22
New cards

Off-Focus Radiation

Photons not created at the focal spot but elsewhere in the tube. Produces ghost-like shadow on the radiograph

23
New cards

Determining the diaphragm field size aids in determining….

The projected image on the image receptor

24
New cards

Formula for Diaphragm Field Size

Projected image size = SID × diameter of diaphragm / distance from focal spot to diaphragm

25
New cards

What do cones affect?

  • Does not reduce penumbra

  • Does not increase recorded detail

  • Improves contrast

  • Only cleans up scatter


26
New cards

What does a cylinder affect?

  • Reduces penumbra

  • Reduces off-focus radiation


27
New cards

Cylinder Disadvantages

  • Difficult to use due to weight

  • Fixed field size


28
New cards

Cone Field Size Equation

Cone Field Size = SID × lower diameter of cone / distance from focal spot to bottom of cone

29
New cards

What is the most common type of beam restrictor used in radiography?

Collimator

30
New cards

Collimator advantages & disadvantages

  • More complex and expensive

  • Allows numerous field sizes

  • Provides a light source to aid in positioning


31
New cards

A collimator should never be opened greater than…

The size of the image receptor

32
New cards

Upper shutters on collimators reduce…

Off-Focus Radiation

33
New cards

Bottom shutters on collimators reduce…

Penumbra

34
New cards

Positive Beam Limitation

Sensing device that determines the size and placement of cassette in bucky tray. Automatically collimates to size of cassette by moving shutters.

35
New cards

US required by law from 1974 to 1993 that all fixed x-ray equipment have…

Positive Beam Limitation Device

36
New cards

Lead Blockers

Sheet of lead-impregnated rubber that can be cut into any size or shape. Absorbs scatter from large body parts and aids in positioning

37
New cards

Where are lead blockers most commonly used?

Lateral L-Spine, Lateral T-Spine

38
New cards

Lead Blocker affects…

  • Cleans up scatter

  • Does not decrease patient dose


39
New cards

Lead Mask

Cut to field size being used, secured at end of collimator, most common in cerebral angiography

40
New cards

Increasing field size also means…

Opening the collimator

Decreasing beam restriction

41
New cards

Increasing field size affects…

  • Increased scatter

  • Increased IR exposure

  • Decreased contrast


42
New cards

Decreasing field size also means….

  • Closing collimation

  • Increasing beam restriction


43
New cards

Decreased field size affects….

  • Decreases beam quantity

  • Decreases scatter

  • Increases contrast

  • Decreases IR exposure


44
New cards

Increase in mAs must accompany a significant reduction in field size to…

Maintain IR exposure