RTE 1401 Radiographic Imaging Review

0.0(0)
Studied by 1 person
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/43

flashcard set

Earn XP

Description and Tags

Comprehensive review flashcards covering discovery, SI units, tube design, interactions, exposure factors, grids, fluoroscopy, digital detectors, and dose reduction.

44 Terms

1
New cards

When, where, and by whom were x-rays discovered, and what equipment was being used?

Wilhelm Conrad Roentgen discovered x-rays on November 88, 18951895, in Würzburg, Germany, while experimenting with a Crookes tube (cathode ray tube).

2
New cards

What material fluoresced during the discovery of x-rays, and who was the subject of the first x-ray image?

A barium platinocyanide screen fluoresced even though the tube was covered in black cardboard; the first image produced was of his wife Anna Bertha's hand on December 2222, 18951895.

3
New cards

Who developed the hot-cathode x-ray tube in 19131913, and who was the first documented x-ray fatality in 19041904?

William Coolidge developed the hot-cathode tube in 19131913; Clarence Dally, an assistant to Thomas Edison, was the first documented x-ray death in 19041904.

4
New cards

What are the standard SI units for exposure, air kerma, absorbed dose, and equivalent/effective dose?

Exposure is measured in C/kg\text{C/kg}; air kerma is measured in Gy\text{Gy}; absorbed dose is measured in Gy\text{Gy} (1 Gy=1 J/kg1\,\text{Gy} = 1\,\text{J/kg}); equivalent dose and effective dose are measured in Sv\text{Sv}.

5
New cards

What are the radiation weighting factors (WRW_R) for x-rays, gamma rays, beta particles, and alpha particles?

X-rays, gamma rays, and beta particles have a WR=1W_R = 1; alpha particles have a WR=20W_R = 20.

6
New cards

At what velocity do x-ray photons travel?

X-rays travel at the speed of light, which is 3×108 m/s3 \times 10^8\,\text{m/s}.

7
New cards

What is the difference between atomic number (ZZ) and mass number?

Atomic number (ZZ) is the number of protons and determines element identity; mass number equals the total number of protons plus neutrons.

8
New cards

What is the K-shell binding energy of a tungsten atom?

The K-shell binding energy of tungsten is 69.5 keV69.5\,\text{keV}.

9
New cards

What is the sequence of the electromagnetic spectrum from lowest energy to highest energy?

Radio, Microwave, Infrared, Visible light, Ultraviolet, X-ray, and Gamma.

10
New cards

What material is the cathode filament made of, and to what temperature range must it be heated to produce thermionic emission?

The filament is made of a coil of thoriated tungsten and is heated to incandescence at approximately 2,000–2,400 ∘C2{,}000\text{–}2{,}400\,^\circ\text{C}.

11
New cards

What is the space charge effect, and what is its effect on maximum tube current?

The cloud of negative electrons repels additional electrons from boiling off the filament, limiting maximum tube current to about 1,000–1,200 mA1{,}000\text{–}1{,}200\,\text{mA}.

12
New cards

Why is tungsten preferred as a target material in standard diagnostic x-ray tubes?

Tungsten has a high atomic number (Z=74Z = 74) for efficient x-ray production, a high melting point (3,400 ∘C3{,}400\,^\circ\text{C}), and good thermal conductivity.

13
New cards

What is the regulatory limit on leakage radiation from the protective tube housing?

Leakage radiation must not exceed 1 mGya/hr1\,\text{mGy}_a/\text{hr} at a distance of 1 m1\,\text{m}.

14
New cards

How does the line-focus principle function, and what determines actual focal spot size?

The anode target is beveled so the effective focal spot projected toward the patient is smaller than the actual focal spot where electrons strike; actual focal spot size is determined by filament length.

15
New cards

What is the anode heel effect, and how should patient anatomy be oriented to compensate for it?

Beam intensity is higher on the cathode side and lower on the anode side due to target heel absorption; the thicker anatomical region should always be positioned under the cathode side.

16
New cards

What percentage of electron kinetic energy is converted to heat versus x-rays at the anode target?

Approximately 99%99\% is converted to heat and approximately 1%1\% is converted to x-rays.

17
New cards

How is Bremsstrahlung radiation produced at the target?

An incident electron passes near the target nucleus, slows down, and changes direction; the lost kinetic energy is emitted as a polyenergetic x-ray photon.

18
New cards

What minimum tube potential (kVp\text{kVp}) is required to produce characteristic x-rays in a tungsten target, and why?

A minimum of 70 kVp70\,\text{kVp} is required because the projectile electron must have energy equal to or greater than the tungsten K-shell binding energy of 69.5 keV69.5\,\text{keV}.

19
New cards

What is the regulatory total filtration requirement for diagnostic x-ray tubes operating above 70 kVp70\,\text{kVp}?

Total filtration must be at least 2.5 mm Al2.5\,\text{mm}\,\text{Al} equivalent.

20
New cards

Approximately what proportion of the peak kilovoltage (kVp\text{kVp}) is the average energy of a diagnostic x-ray beam?

The average beam energy is approximately 13\frac{1}{3} of the peak set kVp\text{kVp}.

21
New cards

What is the 15%15\% rule, and how is it used to reduce patient dose while maintaining receptor exposure?

A 15%15\% increase in kVp\text{kVp} doubles IR exposure; to maintain exposure and lower patient dose, increase kVp\text{kVp} by 15%15\% and reduce mAs\text{mAs} by 12\frac{1}{2}.

22
New cards

What is the mathematical formula for the Inverse Square Law?

I1I2=D22D12\frac{I_1}{I_2} = \frac{D_2^2}{D_1^2}

23
New cards

What is the mathematical formula for the Direct Square Law (exposure maintenance formula)?

mAs1mAs2=D12D22\frac{\text{mAs}_1}{\text{mAs}_2} = \frac{D_1^2}{D_2^2}

24
New cards

What occurs during a photoelectric interaction, and what mathematical relationship describes its probability?

An incident photon is completely absorbed by an inner-shell electron, ejecting a photoelectron; its probability is directly proportional to Z3E3\frac{Z^3}{E^3}.

25
New cards

What occurs during a Compton interaction, and what is its impact on imaging and safety?

An incident photon ejects an outer-shell electron, loses energy, and scatters in a new direction; it degrades image contrast with scatter noise and is the primary source of occupational dose.

26
New cards

What are the threshold energy requirements for pair production and photodisintegration?

Pair production requires at least 1.02 MeV1.02\,\text{MeV}; photodisintegration requires greater than 10 MeV10\,\text{MeV}.

27
New cards

What is the order of human tissue attenuation from greatest to least?

Metal/contrast >> bone >> muscle >> fat >> air.

28
New cards

How do additive and destructive pathologies affect technique selection?

Additive pathologies increase tissue thickness, density, or ZZ, requiring an increase in technique (usually kVp\text{kVp}); destructive pathologies decrease tissue thickness, density, or ZZ, requiring a decrease in technique.

29
New cards

How is the Magnification Factor (MF) calculated mathematically?

MF=SIDSOD\text{MF} = \frac{\text{SID}}{\text{SOD}}

30
New cards

What causes elongation versus foreshortening shape distortion?

Elongation is caused by tube or image receptor misalignment; foreshortening is caused by anatomical part misalignment.

31
New cards

What clinical thresholds dictate when an anti-scatter grid should be utilized?

A grid is indicated when anatomical part thickness exceeds 10 cm10\,\text{cm} and tube potential exceeds 60 kVp60\,\text{kVp}.

32
New cards

What are the Grid Conversion Factors (GCF) for no grid, 5:1, 6:1, 8:1, 10–12:1, and 16:1 grids?

No grid =1= 1; 5:1 =2= 2; 6:1 =3= 3; 8:1 =4= 4; 10–12:1 =5= 5; 16:1 =6= 6.

33
New cards

What image pattern results from off-level or off-center grid errors versus off-focus or upside-down grid errors?

Off-level and off-center errors cause grid cutoff across the entire image; off-focus and upside-down errors cause cutoff primarily along the peripheral edges.

34
New cards

What are the source-to-skin distance (SSD) limits for stationary and mobile fluoroscopy units?

Fixed fluoroscopic units require an SSD of at least 38 cm38\,\text{cm} (15 in.15\,\text{in.}); mobile C-arms require an SSD of at least 30 cm30\,\text{cm} (12 in.12\,\text{in.}).

35
New cards

What are the regulatory exposure rate limits for normal and high-level control (HLC) fluoroscopy?

Normal fluoroscopy is limited to 88 mGya/min88\,\text{mGy}_a/\text{min} (10 R/min10\,\text{R/min}); high-level control (HLC) is limited to 176 mGya/min176\,\text{mGy}_a/\text{min} (20 R/min20\,\text{R/min}).

36
New cards

What is the sequence of components in an image intensifier (II) tube from input to output?

Input phosphor (cesium iodide) →\rightarrow Photocathode (antimony–cesium) →\rightarrow Electrostatic focusing lenses →\rightarrow Accelerating anode (∼25,000 V\sim 25{,}000\,\text{V}) →\rightarrow Output phosphor (zinc cadmium sulfide).

37
New cards

How is brightness gain calculated in an image intensifier tube?

Brightness gain=minification gain×flux gain\text{Brightness gain} = \text{minification gain} \times \text{flux gain}, where minification gain=(input diameteroutput diameter)2\text{minification gain} = \left(\frac{\text{input diameter}}{\text{output diameter}}\right)^2.

38
New cards

How does magnification mode in fluoroscopy alter spatial resolution and patient radiation dose?

Magnification mode improves spatial resolution but increases patient dose by approximately (normal field diametermag field diameter)2\left(\frac{\text{normal field diameter}}{\text{mag field diameter}}\right)^2.

39
New cards

How does the fluoroscopy cumulative timer operate?

It tracks cumulative beam-on time and sounds an audible alarm when 5 minutes5\,\text{minutes} of exposure have elapsed to notify the operator.

40
New cards

What materials are used for x-ray detection and conversion in Indirect DR versus Direct DR?

Indirect DR uses a scintillator (cesium iodide or gadolinium oxysulfide) paired with an amorphous silicon photodiode; Direct DR uses amorphous selenium as a photoconductor without a scintillator.

41
New cards

How does the Fuji S# exposure indicator relate to receptor exposure compared to Kodak EI?

The Fuji S# is inversely proportional to exposure (a change of +200+200 indicates half the exposure); Kodak EI is directly proportional (a change of +300+300 indicates double the exposure).

42
New cards

What image characteristics do window level and window width control during post-processing?

Window level controls image brightness (center of the gray scale); window width controls image contrast (the range of displayed gray shades).

43
New cards

In an Automatic Exposure Control (AEC) circuit, which technical parameter is automatically terminated?

The radiographer selects kVp\text{kVp} and mA\text{mA}, and the AEC system automatically terminates the exposure time.

44
New cards

How does mAs adjustment differ between fixed kVp and variable kVp technique systems as anatomical thickness varies?

In fixed kVp\text{kVp} systems, mAs\text{mAs} is doubled for every 4–5 cm4\text{–}5\,\text{cm} increase in part thickness; in variable kVp\text{kVp} systems, mAs\text{mAs} remains constant while tube potential is adjusted by ±2 kVp\pm 2\,\text{kVp} per centimeter.