Main circuit supplies the tube with
properly modified power that will lead to the xray production
Filament circuit supplies the filament of the tube with
properly modified power to create thermionic emission
with modern systems, the incoming line gets supplied with
three-phase power
primary circuit is
low voltage
secondary circuit is
high voltage
the primary and secondary circuit make up the
main circuit
the complete basic circuit is made up of
main breakers, exposure switch, autotransformer, timer circuit, high voltage step up transformer, four diode rectification circuit, filament circuit variable resistance, filament step-down
The circuit must boost the
voltage to the kV range necessary for x-ray production
The circuit must permit the technologist to
adjust the amperage, voltage and length of exposure
The main switch and circuit breakers are usually
enclosed in an electrical power box
the exposure switch permits
current to flow through the circuit
the timer circuit ends the exposure
at a preset time
the exposure switch must be held until
both the audible and visible indicators (buzzer and light) have stopped
holding the exposure switch lets the current flow and activates the
rotating anode
exposure switch half depression
activates anode (rotoring)
exposure switch fully depressed
intitates the exposure
minimum length of the portable exposure switch cord
6ft
switches must be
dead-man type
main breaker is where the
alternating current is coming in
auto transformer
controlled by the kVp selectors on the control console, modifies incoming-line voltage in anticipation of kV that will be produced by step up transformer
some units divide kVp adjustments into
minor (1-2) and major (10) change in units
electronic timers
capable of accurate exposures as short as 0.001 second, microprocessor controlled, contained in most radiographic equipment
AEC timer function by
measuring a preset quantity of radiation and breaking the timer circuit when a dose sufficient to produce the desired exposure is reached
mAs timer
used in capacitor discharge unit to monitor product of mA and time on secondary side of high voltage step up transformer, interrupts circuit when mAs reached, falling load generator, after secondary coil in high volt
all AEC have a minimum reaction time (response time)
determined by the length of time necessary for the AEC to respons to the radiation and terminate exposure, minimum reaction time is shortest time for cells to decide to stop exposure
backup timer
terminates exposure at specified time if AEC fails, automatic termination at 600 mAs when above 50 kVp
autotransformer is controlled by the
kVp selection
timer circuit is wired between
autotransformer and step up transformer
high voltage step up transformer will
boost the voltage to desired kVp set
four diode rectification circuit
changes the incoming alternating current to direct current that the tube needs
the filament circuit
low voltage, modifies incoming power to produce thermionic emission, 3-5 amp and 5-15 volts, mA controller
why does increasing amperage decrease voltage
to balance the two, P = IV
The step-down reduces the V and increases the current (amps) in response to the
rheostat (variable resistor)
filament circuit variable resistance is controlled by the
mA set at the control panel
the filament step down transformer give the filament
high amperage needed to heat the filament
rotor stator
activated by the tech when the two step button is halfway depressed, what causes the anode to rotate during x-ray production
diagnostic xray range of exposure is approximately
10-1200 mA, 0.001-10 seconds, 25-150 kVp
What device in the x-ray circuit controls kVp?
autotransformer
What device in the x-ray circuit controls mA?
step down transformer
What device in the x-ray circuit controls time?
exposure timer
What device in the x-ray circuit controls rectification?
the number of diodes
At what point in the x-ray circuit is the rectification circuit located?
between the step up transformer and the xray tube
To protect the radiographer from electrical shock during x-ray exposures,
a. Operator controls are on the high voltage side of the circuit
b. The operator stands on a grounded rubber mat
c. Release of the switch initiates the exposure sequence
d. The mA and kVp selections are on the low voltage side of the circuit
The mA and kVp selections are on the low voltage side of the circuit
All of the following are radiographer-operated controls EXCEPT
a. kVp selection
b. Timer selection
c. Rotor speed
d. mA selection
Rotor speed