brightfield resolution
300nm
fluoresent (widefield) resolution
250nm
confocal resolution
200nm
2 photon resolution
350 nm
STED resolution
70nm
STORM resolution
20nm
SIM resolution
120nm
light sheet resolution
1um (micrometer)
lattice: 300nm
thin lens equation
1/f = 1/b + 1/v


f = focal length
v = distance of object to lens
b = distance of screen to get a sharp (resolved) image
brightfield resolution equation
R = 1.22L / 1.7 NAobj
L = lambda (wavelength)
NAobj = numberical aperture
widefield (fluoresent) resolution equation
R = 0.61L / NA
confocal/multiphoton resolution equations (lateral, axial)
RESOLUTIONx,y = 0.5L / NA
RESOLUTIONz = 1.4Ln / NA²
n = refractive index
dwell time equation
for confocal, raster scanning
Td = 1 / (line scan rate x pixel format)
= 1 / (scan rate x no. of pixels)
signal to noise ratio equation and definition
SNR = (mean signal value) / (background STDEV)
SNR: strength of true signal to unwanted signal (noise)
mean signal value: average intensity of desired signal
background standard deviation: statistical measure of variability of unwanted noise where no signal is present, selecting background region and calculating STDEV of pixel intensities
what is um (micrometer)
10^-6 m
what is nm
10^-9 m
how much is 100nm in um?
100nm = 1um
which microscopes are 2d and 3d?
2D: brightfield, fluorescent (widefield)
3D: confocal, multiphoton, STED, light sheet
limited 3D: STORM, SIM (better for thin samples)
deconvolution
process to improve resolution and contrast of images
mathematical, need to know PSF and deconvolve signal
optical tissue clearing
another method for making clear images/transparent images
using chemicals that match the refractive index (homogenizing it)
decolorization

jablonski for fluoresent, confocal, light sheet, SIM
normal excitation and emission, emitted wavelength is longer (less energy)

jablonski for multiphoton microscopy
use 2-3 infared photons to excite electron, emit
? pulsed laser?

jablonski for STED (donut)
2 lasers, depletion and excitation
depletion redshifts some photons, so normal green energy is between

jablonski for STORM
because of photoswitching, sometimes don’t emit photons (dark state)
some fluoresent

jablonski for SHG/THG
not excitation of electrons, scattering effect, upper is “virtual state”
energy is equal, no energy loss
based on the structure of the thing
no photobleaching?
which techniques are 2d and 3d?
2d: brightfield, fluoresent widefield
3d: confocal, multiphoton, sted, light sheet
limited 3d: storm, sim
how does brightfield microscopy work? + speed
illumination from underneath
stain different parts (dead cell)
important parts of brightfield microscope

ocular
tube lens
objective lens
stage/sample
condenser: focuses light into cone onto specimen
field diaphragm: controls diameter of light, helps evenness and clarity
led light
how to increase contrast of brightfield microscopy?
staining
DIC
adding normaski prisms, 2 polarizer
light passes through differently, uses polairizer

phase contrast
need phase plate and condenser annulus (opposite circles) above and below
phase shift from amplitude change, can study LIVING CELLS
depends on index of refraction n

how does fluorescent microscopy work? + speed
fast
any fluorophores
use fluorescent dyes to excite, combine pieces into one image
white light
what are extra important parts of fluorescent?
excitation and emission filters, let some light in and get rid of extra light
dichroic mirror: reflects and lets through light
make a cube (different filter sets)

fluorescnet what is the use (typical sample) and type of image produced?
sample, uses camera, produces bright structures on dark background (sequential put togheter)
what are 3 staining methods for fluorescent?
direct binding
easy, but not many avaialbel
fluorescent proteins
genetically modify, live imaging, time consuming
immunostaining (need to fixate)
2 antibodies, primary and secondary
anything can be stained, but also time consuming
stoke shift
difference of absorption and emission maximum absorption graphs

optical filters: short pass, long pass, dichromatic beam splitter, wide band pass, naroow band pass, combination?

short waelengths pass
long wavelengths pass
di reflect and pass
wide range of passing, transmittance short
narrow range of passing
how to do indirect immunostaining
each primary antibody from different species
searching for human → antihuman antibody
different host (rabbit)
2nd is antimouse, host donkey>
each primary antibody from different species
make sure different colours
rainbow wavelenghts (red to blue)
750, 620, 590, 570, 495, 450 nm
red orange yellow green blue
how does confocal microscopy work? (type of image, extra parts, speed)
use pinholes to get ride of out of focus light, otherwise kinda same
by laser and before detector
using signals and a PMT, getting z stacks and putting image together
only take the in focus plane at a time
slow, goes pixel by pixel
can do 3D imaging
uses a laser
typical use of confocal microscopy
cells, thin tissue, resolution 200nm
multiphoton microscopy
SHG
second harmonic gen