Equip - Unit 3, Wk 10: Prime Factors

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/59

flashcard set

Earn XP

Description and Tags

Unit 3, Wk 10: Prime Factors

60 Terms

1
New cards

Prime Factors:

principle factors that the radiographer controls

  • mAs, kVp, distance


2
New cards

X-ray Quantity:

Number of x-ray photons in the beam; measured in mR (milliroentgen)

3
New cards

Factors that affect x-ray quantity/Quantitative Factors:

  • milliamperage-second (mAs) - primary controller!

  • kVp

  • distance (IR to the patient)

  • filtration


4
New cards

X-Ray Quality:

Penetrating ability of the x-ray beam

5
New cards

Penetrability:

Distance an x-ray beam travels in matter

  • High-energy (hard x-rays) photons travel farther and are more penetrating


6
New cards

Factors that affect x-ray quality/Qualitative Factors:

  • kilovoltage (kVp)

  • Filtration


7
New cards

Milliamperage (mA):

Number of electrons crossing the tube from the cathode to the anode per second


8
New cards

What is the mA and tube current relationship?

Directly proportional

-as the mA doubles, so does the # of electrons crossing the tube and shooting toward the target


9
New cards

As mAs increases, IR exposure _______

Increases (Directly proportional)

10
New cards

Why is x-ray measured in milli-seconds?

Helps decrease motion distortion because it is very short and fast

11
New cards

What is the unit of measurement for x-ray quantity?

mR

12
New cards

______ is the primary controller of x-ray quantity

mAs

13
New cards

As mAs doubles, x-ray exposure (mR) _______

If either the mA or the exposure time is _________, the mAs will double

  • doubles

  • doubles


14
New cards

Calculating mA:

mAs = s x mA

  • Make sure everything is in the correct units first, may have to convert


15
New cards

Solve:

60 mAs = 300 milliseconds (ms) x _____ mA

  1. convert milliseconds to seconds by moving the decimal 3 places to the left

  2. 60 = 0.3x

  3. Get x alone. 60/0.3 = 0.3x/0.3

  4. 60/0.3 = x

  5. x = 200

  6. 200 mA

  7. double check by plugging it in with the converted seconds

  8. 60mAs = 0.3 seconds x 200mA


16
New cards

What is the minimum mAs change value for a visible change with film/screen?

25-33%

17
New cards

Rule of thumb for mAs changes for repeat exposures:

Make adjustments in doubles or halves

-EX: a repeated image that’s underexposed at 10 mAs should be repeated at 20, 40, or 80, depending on severity of underexposure

18
New cards

All factors remaining constant, which of the following set of exposure conditions would yield an image with the HIGHEST image receptor (IR) exposure?

A. 100 mA, .50 sec, 40” SID
B. 150 mA, .25 sec, 44” SID
C. 300 mA, 125 ms, 40” SID
D. 200 mA, 187 ms, 56” SID

A

  • multiply mA by seconds (mA x time = mAs). In this case option A has the highest mAs


19
New cards

What does kVp control?

Primarily controls quality and secondarily affects the quantity of the x-ray beam emission

20
New cards

What does mA control?

Quantity (current of tube)

21
New cards

Why does kilovoltage also affect the quantity?

Because more interactions occur at the target as kVp (energy) increases, producing more X-rays

22
New cards

X-ray beam penetrability is primarily controlled by

kVp

23
New cards

As x-ray photon energy increases, the ___________ ability of the photon ______

  • penetrating

  • increases


24
New cards

As kVp is doubled the amount of x-ray photons increases approximately _______

four times

25
New cards

True or False:

kVp has a big impact on IR exposure

True

26
New cards

What does kVp change?

the amount of x-ray photons created at the anode

27
New cards

15% Rule:

  • Increase in kVp by 15% will cause doubling in IR exposure

  • Decrease in kVp by 15%, IR exposure will be reduced by one-half


28
New cards

Rule of thumb for kVp changes:

15% rule

29
New cards

True or False:

kVp has an effect on the speed and energy of the electrons in the x-ray tube

True

30
New cards

When kVp goes down, mAs goes ______

up

EX: decreasing kVp by 15% should be accompanied by doubling the mAs

31
New cards

True or False:

When kV is changed, it will affect the contrast of the digital image

False, technology nowadays is advanced to adjust contrast

-This statement is true for film

32
New cards

How to calculate kVp value:

  • Know if you are doubling or halving

  • kVp -/+ (kVp x 15%)


33
New cards

Fixed kVp Systems:

are best suited for use with digital image receptors

  • Use an optimal kVp that is not changed

  • Double or halve mAs for every 5 cm of subject thickness


34
New cards

A radiograph of an ankle is produced at 55 kVp at 2 mAs. What kVp would be required to halve the exposure to the IR?

A. 63 kVp
B. 47 kVp
C. 28 kVp
D. 25 kVp

B

  • 15% of 55 = 8.25

  • 55-8.25 = 46.75

  • round up, answer is 47 kVp


35
New cards

What to do to the technical factors to decrease dose?

Increase kVp, decrease mAs

36
New cards

The product of tube current and exposure time is equal to ______

x-ray beam quantity

37
New cards

Memorize

knowt flashcard image
38
New cards

What to do to technical factors to increase dose?

increase mAs, increase dose

39
New cards

Distance and x-ray emission relationship

Inversely proportional (greater the distance, weaker the x-ray/intensity)

40
New cards

kV controls the heat at ____

anode

41
New cards

mAs controls the heat at ______ and _______

filament and anode

42
New cards

Easier version of the Inverse Square Law:

knowt flashcard image
43
New cards

Inverse square relationship between intensity and distance:

As distance increases, radiation intensity and IR exposure decrease

<p>As distance increases, radiation intensity and IR exposure decrease</p>
44
New cards

When you double the distance, you _____ the intensity

1/4

<p>1/4</p>
45
New cards

Changes in distance do not cause changes in mAs.
The radiographer must adjust the _____ to compensate for distance changes

mAs

46
New cards

Exposure Maintenance Formula:

knowt flashcard image
47
New cards

Exposure Maintenance Example


<p></p>
48
New cards
<p>With digital, these x-rays appear the same but have different EI values.</p><p>In order to maintain the same appearance, what correction was made in relation to the SID?</p>

With digital, these x-rays appear the same but have different EI values.

In order to maintain the same appearance, what correction was made in relation to the SID?

Distance was doubled, so mAs was increased 4x to maintain exposure

EX: In Figure A, 60@2 In Figure B, 60@8

49
New cards

Exposure Index (EI):

Actual Exposure

50
New cards

Target EI (EIt):

Ideal Exposure

51
New cards

Deviation Index (DI):

% Deviation from target (how far you were off)

  • DI value of 2 or more, you may need a repeat


52
New cards

Carestream/Siemens Optimal EI Range:

1500-1800

  • Higher EI = overexposed

  • Lower EI = underexposed


53
New cards

If you double the mAs, your EI will _______

double

-directly proportient

54
New cards

You have taken a digital radiograph at 80@20 that yields an EI# of 85. The recommended target EI# for this projection is 170. When you repeat the image, you will achieve the desired results by:

doubling mAs to 40

  • 85 × 2 = 170


55
New cards

Fuji, Konica, and Phillips Optimal S number Range

150-250

  • Higher S number = underexposed

  • Lower S number = overexposed


56
New cards

Setting the appropriate prime factors of ____, _____, and _______ will ensure effective exposure to IR

  • mAs

  • kVp

  • distance


57
New cards

Use the fixed kVp appropriate for that exam (which is based on the desired contrast), then mAs is adjusted to provide the appropriate exposure to the IR (this is what AEC does...)

!!

58
New cards

All of these are true EXCEPT:

A. Brightness and contrast are controlled primarily through post-processing parameters
B. Exposure index numbers are inconsistent between manufacturers
C. Digital detectors will not compensate for minor errors in technique selection
D. Fixed kVp systems are best suited for use with digital image receptors

C

-Digital detectors WILL compensate for minor errors in technique selection

59
New cards

Which equation do we use for this?

An x-ray exposure of 200 mR is recorded at a distance of 40 inches. If the same technical factors are used, what will the exposure be if the distance is increased to 72 inches?

Inverse square law

60
New cards

In order to increase image receptor (IR) exposure by 40%, the radiographer would

change the mAs from 15 to 21 mAs

<p>change the mAs from 15 to 21 mAs</p>